HiQuad QMA 1-300 — instruction manual qmg 800
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum QMG 800 HIQUAD NEO quadrupole mass spectrometer system operating instructions (EN, 68 pages). Covers article numbers PT Q25 112 xxx, PT Q25 121 xxx, PT Q25 132 xxx, PT Q25 312 xxx (QMG 800 HiQuad Neo incl. SEM 217 and EP 822) and PT 168 011 (input/output module IO 820). System comprises control unit QMS 800 (system chassis SC 800 + quadrupole controller QC 800, plug-in modules IS 816, HV 801, IO 820), HF generator QMH 800-x, analyzer QMA 4x0 with SEM 217 and EP 822 preamplifier. Related documents: BG 6016 (QMH 800-x), BG 6017 (QMG 800 communication), BG 6018 (QMA 4x0).
Section-by-Section Technical Breakdown
Describes the system architecture: control unit QMS 800 (chassis SC 800 with power pack, system bus, ventilation; QC 800 controller with LAN/USB interfaces), plug-in modules IS 816 (ion source supply, reversible polarity, short-circuit proof), HV 801 (high voltage for SEM 217, positive ions), IO 820 (analog/digital I/O plus total pressure gauge connection), QMH 800-x HF generator, QMA 4x0 analyzer, and EP 822 electrometer preamplifier. Provides detailed pin assignments for ANALOG I/O, DIGITAL I/O, gauge connectors, CTRL, AUX I/O, QMA and LAN connectors, plus electrode potential tables and wiring diagrams for Faraday and 90° SEM operating modes.
- IO 820: 5 differential analog inputs, 4 analog outputs, 4 digital inputs (24 V logic), 16 optocoupler-isolated digital outputs; SERIAL GAUGE connector for DigiLine and TP GAUGE for ActiveLine gauges (signal 0-+10 V DC, supply +24 V DC)
- CTRL connector (HD-D-Sub 26-pin): TTL inputs RUN IN / IN 0-2 / SYNC IN with internal pull-up 5.6 kΩ to +5 V; differential analog inputs EXT IN 1/2 max ±10 V; analog outputs ELM OUT / AO MON 0-1 ±10 V 12-bit, SCO mass-number output 0-+10 V 16-bit, all max 0.1 mA via 200 Ω
- Detector range code output: 1 V at 1×10⁻¹² A per V, up to 8 V at 1×10⁻⁵ A (0 V in ion counter mode); intensity signal linear ±10 V/decade or logarithmic 1-10 decades
- AUX I/O pin 1 EXT PROT: 24 V filament protection input — a potential-free contact to GND (pull-up 5.6 kΩ to +24 V) must hold the pin, otherwise emission is switched off
- Ion reference IONREF (V1) range −150 to +150 V, ±2.5 mA, resolution 20 mV, gain error 1.6%; V0 may be set to max ±200 V to QMA GND for special applications; external voltage source for V0 must be limited to ≤2 mA; degas potential on V1 is +550 V
- Other electrodes: CATH (V2) 0 to −125 V; WEHNELT (V9) 0 to −60 V; EXTRACT/DEF-I/DEF-O/RES (V5-V8) ±100 µA, resolution 58.82 mV; FOCUS/F-AXIS ±3 mA
- In Faraday operating mode, V6 and V7 are connected to QMA GND; SPEC SRC ON signal is +24 V if activated, max 200 mA
- Cable set: IS 816–QMA 4x0 and QC 800–QMH 800-x and LAN 2 QC–QMH cables in 1.5/3/10 m lengths (PT 168 511-513/521-523/541-543 -T); HF connection QMH–QMA 0.7 m (PT 168 560 -T); FA connection 0.7 m (BG541962 -T); EP–QMH cable 0.85 m permanently mounted
- 90° SEM mode additionally requires HV connection HV 801–QMA (HV−, PT 168 521-523 -T) plus HV jump plug (SHV type) and EP jump plug (TNC type)
Covers installation of the overall system: SC 800 as 19" rack module or tabletop unit (plastic feet, locking pins), inserting/replacing plug-in modules in standard positions, mounting EP 822 directly on the analyzer, and installing QMH 800-x and QMA 4x0. Includes cable-making instructions for ANALOG I/O and DIGITAL I/O connections. Warns of lethal voltages and contamination effects on measurements.
- Wear clean, lint-free, powder-free laboratory gloves during high/ultra-high vacuum assembly — fingerprints increase desorption and falsify measurements
- Always connect the mains cable first (before all other cables) to ensure continuous protective earthing
- Protect the analyzer electrode system from external voltages, galvanic connections, flashover, plasma and particle beams — even small external voltages damage electronics and cause unreliable results
- Provide mechanical touch protection and compulsory supply separation (e.g. door contact) if the analyzer can be touched when the vacuum system is open
Describes installation of PV MassSpec software, switching on the system, commissioning via Hardware Configuration or OPC connection with IP address entry, and use of the PV MassSpec search tool to change the device IP address, subnet mask and gateway. Connection to the PC is via Ethernet (LAN 2, RJ-45).
- QMG 800 can be added to PV MassSpec via Hardware Configuration or via OPC; the device IP address must be entered
- PV MassSpec search tool allows changing device IP address, subnet mask, and gateway
- LAN 2 pinout (RJ-45): pins 1/2 TD+/TD−, pins 3/6 RD+/RD−, pins 4,5,7,8 not used
Covers cleaning of the control unit and system chassis, cleaning the fan (accessible via fan cover on SC 800), and changing the mains fuse (fuse holder on the SC 800 back panel). Disconnect power before all cleaning work.
- Always disconnect the power supply before cleaning the unit
- Fan cleaning is performed via the fan cover; mains fuse replacement via the fuse holder on the SC 800 back panel
A malfunction table (Tbl. 9) lists faults; shipping sections require decontamination of components exposed to toxic process media before shipment and disposal. Recycling/disposal follows general disposal regulations.
- Decontaminate parts contaminated by toxic process media before maintenance, shipping or disposal
- Use original transport packaging, transport locks and protective caps for every transport
Lists spare parts and accessories for QMS 800 (Tbl. 10), spare parts for QMA 4x0 (Tbl. 11), system components (Tbl. 12), and cables/jump plugs (Section 14.3), including the connection cables and cross-cable note per test report.
- Cable lengths available: 1.5 m, 3 m, 10 m (PT 168 511/521/541 -T, 512/522/542 -T, 513/523/543 -T)
- HF cable QMH–QMA: PT 168 560 -T (0.7 m); FA cable: BG541962 -T (0.7 m)
- Cross cables 5 and 6 if required according to the test report
Section 15 tabulates technical data for all modules: SC 800 (limits of use, dimensions), QC 800 (operating modes/parameters, mass flow rate modes, mass scale resolution, measurement speeds, detector types, measuring ranges/resolution, analog and digital FIR filters with filter time constant), IS 816 (filament supply, emission), HV 801 (high voltage part), IO 820 (analog/logic/RS-485 interfaces, gauge connector), QMH/QMA types, and EP 822 (amplifier data, offsets, dimensions). Rating plate: QMS 800 supply 100–240 V AC, 500 W, made in USA.
- Mains supply QMS 800: 100–240 V AC, 500 W (per rating plate example)
- QC 800 offers analog and digital (FIR) filters with documented filter time constants; measurement speeds, mass scale resolution and detector types are tabulated
- EP 822 offers linear (±10 V/decade) and logarithmic (1–10 decades) intensity output configurations; EP 422 with slide-locking plug can be used as alternative (contact PV Service)
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| QMS 800 control unit operates from 100–240 V AC mains, power consumption 500 W (rating plate example, manufactured 2024/03). p.8 ↗ | Standard single-phase supply worldwide; size the supply and fuse for 500 W worst case. | QMS 800 / SC 800 |
| Voltages above 30 V AC or 60 V DC are considered dangerous contact voltages per EN 61010; the QMA analyzer electrode system carries dangerous voltage during operation (electrodes up to +550 V degas on V1, HV 801 for SEM). p.10 ↗ | Analyzer and vacuum chamber must be permanently protective-earthed; provide mechanical touch protection and compulsory supply separation (door contact) when the system is opened. | whole QMG 800 system |
| After switching off the device, wait 60 seconds before disconnecting all cables (power cable last). p.11 ↗ | Internal high-voltage capacitors need this discharge time; violating it risks electric shock during maintenance. | QMS 800 control unit |
| Mains and extension cables must meet double isolation per IEC 61010 / IEC 60950 and be 3-pin with protective earth; connect mains cable before all other cables. p.11 ↗ | Ensures continuous protective earthing; non-compliant cabling creates life-threatening housing voltages. | whole system |
| QMS 800 must only be switched on with the QMA cable properly connected. p.13 ↗ | Powering up without the analyzer cable connected risks damage to the ion source supply and high voltage stages. | QMS 800 / QMA 4x0 |
| IO 820 provides 5 differential analog inputs, 4 analog outputs, 4 digital inputs and 16 optocoupler-isolated digital outputs; TP GAUGE port supplies 0–+10 V DC signal input with +24 V DC supply for ActiveLine gauges. p.18 ↗ | Process interlocks and analog monitoring can be wired directly to the QMS 800 without external DAQ; total pressure gauges (ActiveLine/DigiLine) connect natively. | IO 820 (PT 168 011) |
| CTRL TTL inputs (RUN IN, IN 0–2, SYNC IN) have internal pull-ups of 5.6 kΩ to +5 V and are low-active; RUN IN starts an external measuring cycle. p.24 ↗ | External trigger sources only need an open-collector/potential-free contact pulling to GND; no external pull-up supply needed. | QC 800 CTRL connector |
| Detector range code output encodes gain range as 1–8 V (8 V = 1×10⁻⁵ A down to 1 V = 1×10⁻¹² A); in ion counter mode output is 0 V. Intensity signal is linear ±10 V/decade or logarithmic ±10 V FS with 1–10 decades. p.25 ↗ | Downstream DAQ can decode the active electrometer range and intensity directly; account for the code when logging data automatically. | QC 800 analog monitoring outputs |
| All analog outputs (ELM OUT, AO MON 0/1, SCO, EP OUT) are limited to max 0.1 mA, via 200 Ω series resistors; SCO mass-number output is 0 to +10 V, 16-bit; AO MON 0/1 are ±10 V, 12-bit. p.25 ↗ | Load inputs must be high-impedance; do not drive low-impedance loads or the signal will be distorted. | QC 800 CTRL connector |
| Differential external analog inputs EXT IN 1/2 accept max ±10 V; EXT IN 1 always has low-pass filter/amplifier, EXT IN 2 optionally with or without. p.25 ↗ | External signals (e.g. process parameters) can be acquired in sync with the mass scan; stay within ±10 V to avoid damage. | QC 800 |
| AUX I/O filament protection input EXT PROT (24 V logic, internal pull-up 5.6 kΩ to +24 V): if Extern_Protection is enabled, the pin must be held to GND via a potential-free contact, otherwise emission is switched off. p.26 ↗ | Wire an interlock (e.g. chamber pressure or door contact) to protect the filament from burning out at too-high pressure. | IS 816 ion source supply |
| Electrode potentials: IONREF (V1) −150 to +150 V at ±2.5 mA (resolution 20 mV, gain error 1.6%); CATH (V2) 0 to −125 V at −10 mA; WEHNELT (V9) 0 to −60 V at −500 µA; EXTRACT/DEF-I/DEF-O/RES (V5–V8) ±100 µA with 58.82 mV resolution; degas potential on V1 is +550 V (relative to V0). p.28 ↗ | Defines the electrical operating envelope of the ion source; V1 can reach +550 V in degas — ensure isolation and grounding are rated for this. | IS 816 with QMA 4x0 |
| External voltage source applied to V0 must be limited to ≤2 mA; V0 is normally at QMA GND but may be set to max ±200 V to QMA GND for special applications. p.27 ↗ | When biasing the analyzer (e.g. for in-situ experiments) the current limit protects the electrode system; standard setups keep V0 grounded. | QMA 4x0 analyzer / IS 816 |
| SPEC SRC ON signal at the QMA connector outputs +24 V when activated, max 200 mA. p.27 ↗ | Can be used to power or signal ion-source status in customer interlocking, but the 200 mA budget must not be exceeded. | QMA connection / IS 816 |
| Standard cable kit lengths: 1.5 / 3 / 10 m for IS–QMA (PT 168 511-513 -T), QC–QMH (PT 168 521-523 -T) and LAN 2 QC–QMH (PT 168 541-543 -T); HF cable QMH–QMA 0.7 m (PT 168 560 -T); FA cable 0.7 m (BG541962 -T); EP–QMH cable 0.85 m fixed on EP 822. p.30 ↗ | Physical layout planning of rack/chamber distances must fit these cable lengths; HF/FA cables must not be extended. | whole system |
| For 90° SEM operating mode, additional HV cable HV 801–QMA (HV−) and HV jump plug (SHV type) and EP jump plug (TNC type) are required; EP 822 (2) is used with the SEM 217. p.31 ↗ | SEM detection is only possible with HV 801 module fitted and the correct jumpering; Faraday mode needs none of these. | QMG 800 variants with SEM 217 / HV 801 |
| In Faraday operating mode, deflection electrodes V6 and V7 are connected to QMA GND. p.28 ↗ | No deflection voltages are needed in Faraday mode; wiring per basic Faraday scheme (Fig. 29) suffices. | QMA 4x0 Faraday mode |
| EP 822 electrometer preamplifier is mounted directly on the analyzer to minimize interference, has low noise/drift/vibration sensitivity, and fast recovery after overmodulation; EP 422 with slide-locking plug can be used as an alternative (via Pfeiffer Vacuum Service). p.19 ↗ | Short signal path preserves low-level ion-current measurement integrity (down to ~1×10⁻¹² A range per the range code); do not relocate the preamp. | EP 822 |
| IO 820 provides an RS-485 interface (Tbl. 31) and the QC 800 offers LAN 2 (RJ-45) and USB interfaces to the PC; LAN 1 is for internal use only; CP connector is not used. p.17 ↗ | System integration is via Ethernet/OPC-UA (PV MassSpec, OPC); LAN 1 and CP must not be used for external connections. | QC 800 / IO 820 |
| Improper use expressly includes use with corrosive or explosive media, outdoor use, use outside limits of use, and measurements whose results determine the safety of persons or large values. p.13 ↗ | Process compatibility must be verified before deployment; safety-critical decisions must not rely solely on this instrument. | whole QMG 800 system |
| QC 800 provides analog and digital (FIR) filters with documented filter time constants, plus tabulated measurement speeds, mass scale resolution, detector types and measuring ranges/resolution (Tables 16–24). p.58 ↗ | Speed/resolution/noise trade-offs can be configured per application; consult these tables in the full manual to select scan parameters. | QC 800 |