HiCube Neo RGA (QMG 250 F1) — product leaflet
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum HiCube Neo RGA compact residual gas analysis systems with integrated turbomolecular pump and QMG 250 mass spectrometer. Six variants: F-series with Faraday detector only (QMG 250 F1/F2/F3, mass ranges 1–100/1–200/1–300 u) and M-series with C-SEM + Faraday detectors (QMG 250 M1/M2/M3, same mass ranges). Options include chamber heating, calibration gas, shut-off valve, transport box, mobile cart, various backing pumps/gauges/filaments, and Pfeiffer MassSpec software.
Section-by-Section Technical Breakdown
The HiCube Neo RGA is a compact residual gas analysis system combining an integrated turbomolecular pump unit with a mass spectrometer. Safety is provided by an integrated DigiLine full-range transmitter monitoring total pressure to protect the filament. The system is portable and offers an extended accessory range.
- Integrated turbomolecular vacuum pump unit and mass spectrometer in one compact unit
- Integrated DigiLine full-range transmitter monitors total pressure and protects the filament
- Spare parts: electronics, ion source, filament, analyzer, gauge, pumps/pump service kits, inlet valve
- Options: mass ranges, filament materials, various backing pumps and gauges, chamber heating, PC with Pfeiffer MassSpec software, shut-off valve, calibration gas option, transport box, mobile cart
The F-series covers three mass ranges (1–100, 1–200, 1–300 u) with Faraday cup detection only. Detection limits range from 4·10⁻¹³ to 7·10⁻¹³ hPa and sensitivity decreases with higher mass range. All variants share identical mechanical, electrical and interface specifications.
- Detection limit (Faraday): 4·10⁻¹³ (F1), 5·10⁻¹³ (F2), 7·10⁻¹³ hPa/mbar
- Ar sensitivity (Faraday): 5·10⁻⁴ (F1), 4·10⁻⁴ (F2), 3·10⁻⁴ A/hPa
- Max. operating pressure (Faraday): 5·10⁻⁴ hPa
- Rod system: 6 mm diameter, 125 mm length
- Analyzer operating temperature 200 °C; electronics 5–50 °C; bake-out 300 °C (electronics removed)
- Dimensions 446 x 583 x 434 mm; weight approx. 22–27 kg (depending on options)
- Connecting flange DN 16 ISO-KF / DN 40 CF-F (depending on inlet valve)
- Peak neighbor contribution < 10 / < 20 / < 50 ppm; resolution 0.5–2.5 u at 10% peak height; dwell time 1 ms–16 s/u
- Supply 100–240 VAC (50/60 Hz); Ethernet interface; power max. 240 W (400 W with heating); peak-ratio reproducibility ± 0.5%
The M-series adds a channel electron multiplier (C-SEM) to the Faraday detector, extending detection limits to the 10⁻¹⁵ hPa range. The C-SEM limits the analyzer operating temperature to 150 °C and the max. operating pressure to 5·10⁻⁵ hPa. All other specifications match the F-series.
- Detection limit Faraday/C-SEM: 4·10⁻¹³/3·10⁻¹⁵ (M1), 5·10⁻¹³/4·10⁻¹⁵ (M2), 7·10⁻¹³/5·10⁻¹⁵ hPa
- Max. operating pressure Faraday/C-SEM: 5·10⁻⁴ / 5·10⁻⁵ hPa
- Analyzer operating temperature 200 °C, but max. 150 °C when operating with the SEM
- Same rod system (6 mm x 125 mm), flange options (DN 16 ISO-KF / DN 40 CF-F), bake-out 300 °C (electronics removed)
- Identical electrical data: 100–240 VAC, Ethernet, 240 W (400 W with heating), ± 0.5% peak-ratio reproducibility
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| M-series (C-SEM) detection limit reaches 3–5·10⁻¹⁵ hPa versus 4–7·10⁻¹³ hPa for Faraday-only F-series p.2 ↗ | Choose M-series when trace-gas detection two orders of magnitude lower is required; Faraday suffices for routine RGA work | QMG 250 M1/M2/M3 vs F1/F2/F3 |
| Analyzer max. operating temperature with SEM in use is limited to 150 °C instead of 200 °C p.3 ↗ | Chamber heating setpoints above 150 °C are incompatible with SEM operation; select F-series or reduce heating | QMG 250 M1/M2/M3 |
| Max. operating pressure with C-SEM is 5·10⁻⁵ hPa vs 5·10⁻⁴ hPa for Faraday p.3 ↗ | The DigiLine full-range transmitter must interlock filament/SEM protection before 5·10⁻⁵ hPa in SEM mode | QMG 250 M1/M2/M3 |
| Bake-out temperature is 300 °C with electronics removed p.2 ↗ | Plan bake-out procedures with electronics demounting; do not exceed 200 °C (analyzer) with electronics installed | whole series |
| Electronics operating temperature range is 5–50 °C p.2 ↗ | Ensure ambient/enclosure cooling keeps electronics within 5–50 °C, especially with chamber heating active | whole series |
| Power consumption is max. 240 W without heating and max. 400 W with heating p.3 ↗ | Size supply circuit and heat dissipation for 400 W per unit when chamber heating option is fitted | whole series |
| Supply voltage 100–240 VAC, 50/60 Hz p.2 ↗ | Universal mains input; no voltage variant selection needed | whole series |
| Connecting flange is DN 16 ISO-KF or DN 40 CF-F depending on inlet valve p.2 ↗ | Confirm inlet valve option before specifying chamber flange hardware | whole series |
| Mass ranges are 1–100 u (F1/M1), 1–200 u (F2/M2), 1–300 u (F3/M3) p.2 ↗ | Select variant by highest gas mass to be analyzed; higher ranges trade detection limit and sensitivity | whole series |
| Ar sensitivity decreases from 5·10⁻⁴ (1–100 u) to 3·10⁻⁴ A/hPa (1–300 u) p.2 ↗ | For maximum sensitivity applications prefer the 1–100 u variant | whole series |
| Resolution is 0.5–2.5 u at 10% peak height; dwell time adjustable 1 ms–16 s/u p.2 ↗ | Tune resolution/dwell time per application between fast surveys and high-resolution scans | whole series |
| Contribution to neighboring mass < 10 / < 20 / < 50 ppm by mass range variant p.2 ↗ | For low-level adjacent-peak separation work, prefer the 1–100 u variant | F1/M1 <10 ppm, F2/M2 <20 ppm, F3/M3 <50 ppm |
| Peak-ratio reproducibility is ± 0.5% p.3 ↗ | Suitable for quantitative gas-comparison measurements within ± 0.5% repeatability | whole series |
| Rod system: 6 mm diameter, 125 mm length quadrupole p.2 ↗ | Common mechanical design across variants simplifies spare analyzer stocking | whole series |
| Dimensions 446 x 583 x 434 mm; weight approx. 22–27 kg depending on options p.2 ↗ | Compact portable unit; verify bench/cart load capacity and transport box compatibility for heavier optioned units | whole series |
| Interface is Ethernet; software is Pfeiffer MassSpec (PC available as option) p.1 ↗ | Plan network integration via Ethernet; order the PC option or provide a compatible host | whole series |
| An integrated DigiLine full-range transmitter protects the filament by monitoring total pressure p.1 ↗ | No separate pressure-protection gauge needed, but ensure the DigiLine interlock is wired/configured on startup | whole series |