ATP 1603 M — product leaflet
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum ATP 1603 M magnetically levitated turbomolecular vacuum pump; dimensional/technical data sheet (3 pages, dated 01.07.2025). Notes available accessories/options: installation kit, cooling water valve, purge gas valve, handheld remote controller, inlet protection screen, heater unit, valve cable and coil, various inlet flange types, various user interfaces; on-board control options Profibus or Remote.
Section-by-Section Technical Breakdown
Positions the ATP 1603 M as a quiet, magnetically bearing turbomolecular pump suited to vibration-sensitive semiconductor and analytical applications. It highlights compatibility with various backing pumps, high compression for light gases, and no wearing parts. A list of accessories and options is provided.
- Low noise and vibration levels for vibration-sensitive applications in semiconductor manufacturing and analytical instrumentation
- Magnetic bearing technology — no wearing parts, maintenance-free, long maintenance intervals
- Compatible with a variety of backing pumps; excellent compression ratio for light gases
- Accessories/options: installation kit, cooling water valve, purge gas valve, handheld remote controller, inlet protection screen, heater unit, valve cable and coil, various inlet flange types, various user interfaces
Core vacuum performance data: pumping speeds for N₂, Ar, He and H₂, ISO 5302 compression ratios, ultimate pressure, forevacuum and inlet pressure limits, gas throughput and rotational speed. The rotor speed is variable from 15000 rpm up to nominal speed.
- Pumping speed: N₂ 1250 l/s, Ar 1180 l/s, He 1220 l/s, H₂ 990 l/s
- Compression ratio (ISO 5302): N₂ >1·10⁷, Ar >1·10⁸, He 2·10⁴, H₂ 3.5·10³
- Ultimate pressure < 1·10⁻⁸ mbar
- Max. forevacuum: N₂ 0.8 mbar, Ar 1.5 mbar; max. inlet vacuum: N₂ 0.4 mbar, Ar 0.027 mbar
- Gas throughput at final rotational speed (Ar): 600 sccm
- Rotational speed 37000 rpm, variable 15000 rpm to nominal speed
Electrical supply requirements, run-up time, power consumption, baking limit, noise/vibration levels, dimensions and flange connections, magnetic field limit, and on-board control options. Recommended backing pumps are ADP/ADS series (dry, magnetically levitated).
- Controller power supply 200–240 V, 50/60 Hz single-phase; max start-up power 800 W; nominal (stand-by) power 120 W
- Run-up time to 90% of final rotational speed (pressure exhaust < 0.1 mbar): < 5.3 min
- Max. baking temperature 120 °C
- Recommended backing pump: ADP/ADS
- Sound pressure level (ISO 2151, 1 m, nominal speed) < 48 dB(A); vibration level at nominal speed < 0.01 μm
- Dimensions Ø 259 x 386 mm (height * depending on flange variant); gas inlet DN 200 ISO-K, outlet DN 40 ISO-KF; N₂ purge flange DN 16 ISO-KF
- Stainless steel inlet screen as factory setting
- Permissible radial magnetic field at 10 cm: < 12 mT
- On-board control options: Profibus, Remote; dimensional drawing and pumping speed curves shown
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Pumping speed for N₂ is 1250 l/s; He 1220 l/s; Ar 1180 l/s; H₂ 990 l/s p.1603 ↗ | Select pump sizing based on the process gas; He and H₂ throughput sizing must account for lower H₂ speed despite high compression capability | ATP 1603 M |
| Compression ratio (ISO 5302) for H₂ is 3.5·10³ and He 2·10⁴, vs >1·10⁷ for N₂ and >1·10⁸ for Ar p.1603 ↗ | For hydrogen processes, the forevacuum pressure dominates the ultimate pressure; choose a high-quality backing pump (e.g. ADP/ADS) to reach low ultimate pressures with light gases | ATP 1603 M |
| Ultimate pressure < 1·10⁻⁸ mbar p.1603 ↗ | Suitable for high/ultra-high vacuum applications; system base pressure will be limited by foreline pressure for light gases | ATP 1603 M |
| Max. forevacuum pressure: 0.8 mbar (N₂), 1.5 mbar (Ar); max. inlet vacuum 0.4 mbar (N₂), 0.027 mbar (Ar) p.1603 ↗ | Backing pump must maintain foreline pressure below these limits or the pump will not sustain vacuum performance | ATP 1603 M |
| Gas throughput at final rotational speed is 600 sccm for Ar p.1603 ↗ | Process gas flows above ~600 sccm Ar would exceed the pump's capacity; relevant for sizing gas delivery in sputter/deposition processes | ATP 1603 M |
| Rotational speed is 37000 rpm nominal and is variable from 15000 rpm to nominal speed p.1603 ↗ | Speed can be reduced (e.g. to reduce vibration, heat, or to tune compression) via control interface | ATP 1603 M |
| Run-up time to 90% of final rotational speed (exhaust pressure < 0.1 mbar) is < 5.3 min p.1603 ↗ | Plan process start-up delays accordingly; ensure backing pressure < 0.1 mbar at start for the quoted run-up time | ATP 1603 M |
| Controller power supply: 200–240 V, single-phase, 50/60 Hz; max start-up power 800 W; nominal (stand-by) power 120 W p.1603 ↗ | Specify single-phase mains and a controller sized for the 800 W start-up surge, not the 120 W steady-state power | ATP 1603 M |
| Maximum baking temperature is 120 °C p.1603 ↗ | Bakeout of the vacuum system including this pump must not exceed 120 °C (lower than many all-metal chamber bakes; check flange variant) | ATP 1603 M |
| Sound pressure level < 48 dB(A) (ISO 2151, 1 m, nominal speed) and vibration < 0.01 μm p.1603 ↗ | Can be installed close to vibration-sensitive instrumentation (electron microscopes, metrology) without heavy vibration isolation | ATP 1603 M |
| Dimensions Ø 259 x 386 mm (height depends on flange variant); inlet DN 200 ISO-K, outlet DN 40 ISO-KF, N₂ purge flange DN 16 ISO-KF p.1603 ↗ | Chamber flange design must match DN 200 ISO-K; outlet allows a standard DN 40 ISO-KF foreline; height varies with chosen inlet flange option | ATP 1603 M |
| Permissible radial magnetic field at 10 cm distance: < 12 mT p.1603 ↗ | Avoid installation near strong magnetic sources (magnets, MRI, magnetic stirring) or shield the pump; the magnetic bearing system is sensitive to external fields | ATP 1603 M |
| Stainless steel inlet screen is the factory setting p.1603 ↗ | Particles protection is provided by default; for particle-free UHV applications the screen may need removal — check application | ATP 1603 M |
| Magnetic bearing technology — no wearing parts, no lubricant, long maintenance intervals p.1603 ↗ | Eliminates oil contamination risk and oil-change maintenance; suitable for oil-free/clean processes and for integration with oil-free ADP/ADS backing pumps | ATP 1603 M |
| Recommended backing pumps are the ADP/ADS series (magnetically levitated dry pumps) p.1603 ↗ | For oil-free, vibration-sensitive applications, pair with Pfeiffer ADP/ADS; other backing pumps may also be compatible but ADP/ADS is the documented recommendation | ATP 1603 M |
| On-board control options: Profibus or Remote; accessories include handheld remote controller, valve cable and coil, cooling water valve, purge gas valve, heater unit p.1603 ↗ | Integration into PLC/ fab control is via Profibus; purge/cooling valve accessories enable automated gas purge and cooling control via valve cables | ATP 1603 M |