Induztr
Engineering Digest Document Viewer Related Models

ATH 2300 MT — product leaflet

Technical Reference Summary

Covered Models & Document Scope

Covers two Pfeiffer Vacuum turbomolecular pump models with magnetic bearings: ATH 2300 MT (with integrated heating system for demanding chemical/corrosive conditions) and ATH 2303 M. Flange variants DN 200 / DN 250 affect height dimension. Accessories/options listed: installation kit, cooling water valve, purge gas valve, handheld remote controller, inlet protection screen, heater unit, valve cable and coil, various inlet flanges, various user interfaces, temperature management system for corrosive processes, air cooling system.

Section-by-Section Technical Breakdown

pp. 1
Features and Accessories

Introduces the ATH 2300 MT / 2303 M magnetically levitated turbomolecular pumps. Highlights maintenance-free operation, low contamination in the MT version, frequency-controlled DC motors, and suitability for semiconductor and analytical applications. Lists available accessories and options.

  • Magnetic bearing technology: no wearing parts, maintenance-free
  • MT version: reduced particle contamination / residue build-up, integrated heating system for chemical conditions
  • Compatible with a variety of backing pumps
  • Low noise and vibration; suited to vibration-sensitive applications
  • Options: installation kit, cooling water valve, purge gas valve, handheld remote controller, inlet protection screen, heater unit, valve cable/coil, inlet flanges, user interfaces, temperature management system, air cooling system
pp. 2-3
Pumping Speed, Compression Ratio and Vacuum Performance

Pumping speeds for N2, Ar, He, H2 are identical for both models (1550–2150 l/s N2). Compression ratios to ISO 5302 are >1·10⁸ for N2/Ar but differ for H2 (MT >2·10³ vs M >1·10³). Ultimate pressure <6·10⁻⁹ mbar; max forevacuum and max inlet vacuum limits are tabulated per gas.

  • Pumping speed N2 1550–2150 l/s; Ar 1450–2000 l/s; He 1650–1800 l/s; H2 1100 l/s (both models)
  • Compression ratio ISO 5302: N2/Ar >1·10⁸; He >3·10⁴; H2 >2·10³ (MT) / >1·10³ (M)
  • Ultimate pressure <6·10⁻⁹ mbar
  • Max. forevacuum: N2 2.8 mbar (MT)/2.9 mbar (M); Ar 3.5 mbar; He 2 mbar; H2 0.75 mbar
  • Max. inlet vacuum: N2 0.015 mbar (MT)/0.25 mbar (M); Ar 0.006 mbar (MT)/0.025 mbar (M)
  • Gas throughput at final speed N2: 1000 sccm (MT) / 4000 sccm (M); Ar: 500 / 1500 sccm; He and H2: >1000 / >4000 sccm
pp. 3
Rotational Speed, Power Supply and Cooling

Both pumps run at 31000 rpm nominal, variable from 15000 rpm up to nominal, with <8 min run-up to 90% speed at exhaust pressure <0.1 mbar. Controller requires 200–240 V single-phase 50/60 Hz, with up to 1000 W at start-up. Water cooling is recommended (60 l/h at 15–40 °C); MT supports heating to 75 °C and both allow 120 °C baking.

  • Rotational speed 31000 rpm; variable 15000 rpm to nominal speed
  • Run-up to 90% of final speed <8 min with exhaust pressure <0.1 mbar
  • Controller supply 200–240 V, 50/60 Hz single-phase; max start-up power 1000 W; nominal (stand-by) power 130–380 W
  • Max. heating temperature 75 °C (MT only); max. baking temperature 120 °C (both)
  • Recommended cooling water flow 60 l/h, temperature 15–40 °C
  • Recommended purge flow rate 50 sccm
  • Recommended backing pumps: ADP/ADS
pp. 3
Mechanical Data, Connections and Control Interfaces

Dimensions, weight, flange connections and acoustic/vibration data are identical for both models, except minor fore/inlet vacuum differences. Gas inlet DN 200 or DN 250 (height varies 354.1–398.1 mm with flange variant); forevacuum outlet DN 40 ISO-KF; purge flange DN 16 ISO-KF. Control via Magpower handheld remote; Profibus and Remote on-board options.

  • Dimensions Ø312 x 354.1–398.1 mm (height depends on flange variant); pump weight ~60 kg; controller ~8.7 kg
  • Gas inlet DN 200 / DN 250; outlet DN 40 ISO-KF; N2 purge flange DN 16 ISO-KF
  • Stainless steel inlet screen is factory standard
  • Sound pressure level <47 dB(A) at 1 m (ISO 2151) at nominal speed
  • Vibration level <0.01 μm at nominal speed
  • Permissible radial magnetic field at 10 cm: <12 mT
  • On-board control options: Profibus, Remote; remote controller type Magpower

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Pumping speed for N2 is 1550–2150 l/s and ultimate pressure <6·10⁻⁹ mbar for both ATH 2300 MT and ATH 2303 M. p.2 ↗ Suitable for high-vacuum process and analytical applications requiring UHV-range base pressure without lubricant contamination. Both models
H2 compression ratio differs: >2·10³ (MT) vs >1·10³ (M); max forevacuum H2 is 0.75 mbar for both. p.2 ↗ For hydrogen-heavy processes the MT gives 2× compression margin; backing pump sizing must hold foreline below 0.75 mbar for H2 operation. Both models (H2 compression differs per model)
Gas throughput at final rotational speed for N2: 1000 sccm (ATH 2300 MT) vs 4000 sccm (ATH 2303 M); Ar: 500 vs 1500 sccm. p.2 ↗ ATH 2303 M tolerates ~4× higher gas loads — choose it for high gas-flow processes; MT for clean/low-flow or chemically demanding duty. Model-specific
Max. inlet vacuum (critical backing limit) N2 is 0.015 mbar for MT but 0.25 mbar for M; Ar 0.006 vs 0.025 mbar. p.2 ↗ During roughing/over-pressure events the 2303 M accepts an order of magnitude higher inlet pressure before damage — relevant for cyclic load-lock processes. Model-specific
Nominal rotational speed 31000 rpm, variable down to 15000 rpm; run-up to 90% speed in <8 min at exhaust pressure <0.1 mbar. p.3 ↗ Speed can be throttled for process control; startup sequence must keep exhaust <0.1 mbar to meet the <8 min run-up time. Both models
Controller power supply 200–240 V, 50/60 Hz single-phase; max start-up power 1000 W; nominal/stand-by power 130–380 W. p.3 ↗ Single-phase supply and ≤1 kW inrush simplify installation; size breakers/fusing for 1000 W peak during start-up. Both models
Recommended cooling water flow rate 60 l/h at 15–40 °C; air cooling system available as an option. p.1 ↗ Provide a 60 l/h water loop within 15–40 °C for standard operation; specify the air-cooling option where water is unavailable. Both models
Max. heating temperature 75 °C applies only to ATH 2300 MT (n/a for 2303 M); max. baking temperature 120 °C for both. p.3 ↗ Only the MT variant can be heated for corrosive/condensable processes; bakeout of either pump must not exceed 120 °C. MT: heating; both: baking
Recommended purge flow rate is 50 sccm via DN 16 ISO-KF purge flange. p.3 ↗ Set purge gas supply to 50 sccm to protect bearings from process contamination in corrosive/particle applications. Both models
Recommended backing pumps are the ADP/ADS series. p.3 ↗ Pair with Pfeiffer ADP/ADS dry backing pumps for a validated pump-stack; forevacuum limits (e.g., 2.8–2.9 mbar N2, 3.5 mbar Ar) must be respected. Both models
Gas inlet DN 200 or DN 250, forevacuum outlet DN 40 ISO-KF; pump height 354.1–398.1 mm depending on flange variant, Ø312 mm, weight ~60 kg. p.3 ↗ Verify flange variant when checking chamber-interface and envelope clearances; mounting structure must support 60 kg pump plus 8.7 kg controller. Both models
Sound pressure <47 dB(A) at 1 m (ISO 2151) and vibration <0.01 μm at nominal speed. p.3 ↗ Suitable mounting near vibration-sensitive metrology/semiconductor tools without elaborate isolation. Both models
Permissible radial magnetic field at 10 cm distance is <12 mT. p.3 ↗ Keep strong magnetic sources (magnets, inductors, magnetic-bearing tools) at least ~10 cm away or verify field <12 mT at that distance. Both models
On-board control options are Profibus and Remote; remote controller type is Magpower (handheld unit available as accessory). p.3 ↗ For factory integration select Profibus variant and plan bus cabling; Magpower handheld is used for local setup/diagnostics. Both models
Magnetic bearing technology means no wearing parts and maintenance-free operation; stainless steel inlet screen is factory-fitted. p.1 ↗ No scheduled lubrication/oil changes; the factory inlet screen should normally be retained unless a specific flange/particle trade-off requires removal (accessory inlet protection screens available). Both models