ASM 390 — product leaflet

Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum ASM 390 and ASM 392 helium/hydrogen leak detectors. Both models share the same test methods, power data and dimensions; they differ mainly in helium pumping speed (10 l/s vs 25 l/s) and weight (125 kg vs 130 kg). Document also lists accessories/options: sniffer probe (5 or 10 m hose), helium spray gun, remote control, bottle holder, dust filter.

Section-by-Section Technical Breakdown

pp. 1
Product Highlights

Positioning of the ASM 390/392 as high-performance leak detectors for demanding applications in clean environments. Emphasis on fast start-up, high pumping speed, compact ergonomic design, and user-friendly operation.

  • Fast start-up, high pumping speed, suited for demanding applications in clean environments
  • Compact and ergonomic design with easy access to all testing areas
  • Large rotatable color touch display
  • Integrated tool box and storage space
pp. 1
Accessories, Spare Parts and Options

Lists available accessories and options for the ASM 390/392 leak detectors.

  • Sniffer probe with 5 m or 10 m hose length
  • Helium spray gun
  • Remote control
  • Bottle holder
  • Dust filter
pp. 2
Technical Data

Side-by-side specification table for ASM 390 and ASM 392 covering performance, electrical, environmental and interface parameters. Both models are identical in most respects; the key difference is helium pumping speed.

  • Test methods: spraying (vacuum) and sniffing tests for both models
  • Detectable gases: ⁴He, ³He, H₂
  • Min. detectable leak rate (spraying, helium): 1·10⁻¹² mbar l/s / 1·10⁻¹³ Pa m³/s (both models)
  • Min. detectable leak rate (sniffing, helium): 1·10⁻⁸ mbar l/s / 1·10⁻⁹ Pa m³/s (both models)
  • Helium pumping speed: ASM 390 = 10 l/s, ASM 392 = 25 l/s
  • Power supply: 100-240 V, 50/60 Hz; power consumption 800 W, max. 1600 W
  • Max. test inlet pressure: 20 mbar
  • Start-up time at 20 °C: <2 min; response time <1 s
  • Backing pump: multi-stage Roots vacuum pump; roughing capacity 35 m³/h
  • Operating temperature: 10-35 °C
  • Interfaces: I/O, RS-232, Ethernet (PC, not industrial)
  • Sound pressure level: ≤55 dB(A)
pp. 2-3
Dimensions, Weight and Inlet Connection

Mechanical data including weight, external dimensions and inlet port for both models, plus a dimensional drawing reference.

  • Weight approx.: ASM 390 = 125 kg, ASM 392 = 130 kg
  • Dimensions (L x W x H): 1072 x 455 x 1025 mm (both models)
  • Inlet port: DN 40 ISO-KF (both models)

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Minimum detectable leak rate for helium in spraying (vacuum) test is 1·10⁻¹² mbar l/s (1·10⁻¹³ Pa m³/s) on both ASM 390 and ASM 392. p.2 ↗ Suitable for ultra-tight, high-integrity leak testing applications; selection between models need not be driven by ultimate sensitivity. ASM 390 and ASM 392
Minimum detectable leak rate for helium in sniffing test is 1·10⁻⁸ mbar l/s (1·10⁻⁹ Pa m³/s). p.2 ↗ Sets the practical limit for external sniffer-mode testing; leaks below this rate will not be detected in sniffing applications. ASM 390 and ASM 392
Helium pumping speed: 10 l/s for ASM 390, 25 l/s for ASM 392. p.2 ↗ The ASM 392 recovers faster from large helium loads and suits high-throughput spraying tests; ASM 390 is the lighter, lower-speed option. ASM 390 vs ASM 392 model selection
Detectable tracer gases are ⁴He, ³He and H₂. p.2 ↗ Hydrogen testing is possible (e.g., sensitive method with H₂/H₂ mixtures), not limited to helium only. ASM 390 and ASM 392
Maximum test inlet pressure is 20 mbar. p.2 ↗ Test parts must be evacuated below 20 mbar before connecting to the inlet, or connected via appropriate throttling/roughing, to avoid overpressurizing the mass spectrometer stage. ASM 390 and ASM 392
Start-up time is <2 min at 20 °C and response time is <1 s. p.2 ↗ Enables rapid cycle times in production leak testing; short warm-up allows late start-up in shift-based workflows. ASM 390 and ASM 392
Backing is provided by a multi-stage Roots vacuum pump with 35 m³/h roughing capacity. p.2 ↗ Dry (oil-free) backing simplifies maintenance and avoids oil contamination; 35 m³/h roughing supports gross-leak evacuation of test parts. ASM 390 and ASM 392
Power supply is 100-240 V AC, 50/60 Hz; power consumption 800 W nominal, max. 1600 W. p.2 ↗ Runs from standard single-phase mains worldwide; circuit/fuse sizing should account for the 1600 W maximum draw, not just nominal 800 W. ASM 390 and ASM 392
Operating temperature range is 10-35 °C. p.2 ↗ Installation environment must be climate-controlled; ambient below 10 °C at start-up may lengthen warm-up beyond the <2 min spec. ASM 390 and ASM 392
Communication interfaces are I/O, RS-232 and Ethernet (PC-type, not industrial). p.2 ↗ For PLC integration use the I/O interface; Ethernet is for PC communication only and is not rated for industrial fieldbus use. ASM 390 and ASM 392
Sound pressure level is ≤55 dB(A). p.2 ↗ Suitable for cleanroom/laboratory environments without additional acoustic enclosures. ASM 390 and ASM 392
Inlet port is DN 40 ISO-KF on both models. p.3 ↗ Test fixtures and connection lines must be specified with DN 40 ISO-KF clamps, centering rings and O-rings for full conductance. ASM 390 and ASM 392
Weights are approx. 125 kg (ASM 390) and 130 kg (ASM 392); dimensions 1072 x 455 x 1025 mm (L x W x H) for both. p.3 ↗ Requires rated transport equipment (crane/lift) and a fixed floor position; the narrow 455 mm width suits placement in compact production cells. ASM 390 and ASM 392
Sniffer probes are available with 5 m or 10 m hose length. p.1 ↗ Hose length affects sniffing-mode response time; choose 5 m where shorter response is needed, 10 m for larger test volumes. Optional sniffer probe accessory
A dust filter accessory is available for the inlet. p.1 ↗ Recommended for applications with particle-contaminated test parts to protect the internal pumping system. ASM 390 and ASM 392 (accessory)