Pfeiffer Vacuum ASM 2000 helium leak detector for pharmaceutical package integrity testing (vials, syringes, bottles, other non-porous packages), turnkey with customized test chambers. Covers accessories/options: helium spray gun, sniffer probe, exhaust filter, dry chiller module – 90, customized tooling.

Section-by-Section Technical Breakdown

pp. 1
Product Highlights

The ASM 2000 is a versatile turnkey helium leak detection solution for sterile packaging validation with customizable test chambers. It achieves helium leak detection down to ≥0.01 µm and delivers quantifiable, repeatable results per USP <1207>.

  • Turnkey solution with customized test chambers adaptable to vials, syringes, bottles and other non-porous packages
  • Helium leak detection down to ≥0.01 µm, designed for validating sterile packaging
  • Deterministic, quantifiable and repeatable results according to USP <1207> standard
  • Accessories and options: helium spray gun, sniffer probe, exhaust filter, dry chiller module – 90, customized tooling
pp. 2
Test Methods and Supply Requirements

Supports vacuum and sniffing leak detection with three measurement ranges: quantitative fine leak helium, qualitative gross leak helium, and qualitative massive leak air. Operation requires CDA, helium and nitrogen supplies at 65–145 psig, plus 90–250 V AC mains power.

  • Test methods: vacuum and sniffing leak detection
  • Quantitative fine leak helium range: <1·10⁻⁹ – 1·10⁻⁵ mbar·l/s
  • Qualitative gross leak helium range: 1·10⁻⁵ – 1·10⁻³ mbar·l/s
  • Qualitative massive leak air range: >1·10⁻³ mbar·l/s
  • Minimum detectable helium leakage (sniffing test): >1·10⁻⁵ mbar·l/s
  • CDA supply required, quality 1.3.1 per ISO 8573-1, pressure 65–145 psig / 4.5–10 bar rel.
  • Helium supply required at 65–145 psig / 4.5–10 bar rel.
  • Nitrogen supply required at 65–145 psig / 4.5–10 bar rel.
  • Power supply 90–250 V AC, typical power consumption 3750 W
pp. 3
Operating Environment, Software and Hardware Specifications

Specifications for installation environment (temperature, humidity), the 10" touchscreen user interface, GMP/21 CFR Part 11 compliant software on Windows 10, connectivity, and physical dimensions/weight.

  • Operating temperature 15–25 °C; humidity 30–80 %
  • User interface: 10" multi-touch full HD color screen
  • Software: 21 CFR Part 11 compliance, PDF GMP test and calibration reports, local or domain (LDAP) authentication
  • Operating system: Windows 10
  • Network: 1x LAN (RJ45); interfaces: 2x USB 3.0 (external), 1x HDMI
  • Sound pressure level <53 dB(A)
  • Weight approx. 140 kg; dimensions 913 x 558 x 1197 mm (L x W x H)

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Quantitative fine leak helium measurement range is <1·10⁻⁹ to 1·10⁻⁵ mbar·l/s p.2 ↗ Enables deterministic, quantifiable leak-rate measurement for fine-leak pharmaceutical package integrity testing per USP <1207> ASM 2000
Minimum detectable helium leakage in sniffing mode is >1·10⁻⁵ mbar·l/s (qualitative) p.2 ↗ Sniffing mode is less sensitive than vacuum mode; use vacuum method when quantifying very small leaks ASM 2000
Gross leak (helium, qualitative): 1·10⁻⁵–1·10⁻³ mbar·l/s; massive leak (air, qualitative): >1·10⁻³ mbar·l/s p.2 ↗ Three-tier test strategy: quantify fine leaks, classify gross and massive leaks qualitatively ASM 2000
CDA supply required with quality 1.3.1 per ISO 8573-1 at 65–145 psig (4.5–10 bar rel.) p.2 ↗ Facility compressed-air system must meet ISO 8573-1 class 1.3.1 purity and pressure before installation ASM 2000
Helium and nitrogen supplies are both required at 65–145 psig (4.5–10 bar rel.) p.2 ↗ Plan for three gas lines (CDA, He, N2) at matching pressure range in the installation area ASM 2000
Power supply 90–250 V AC; typical power consumption 3750 W p.2 ↗ Size electrical supply for ~3.75 kW load; wide voltage input simplifies global deployment ASM 2000
Operating temperature 15–25 °C and humidity 30–80 % p.3 ↗ Requires climate-controlled room; ambient outside these limits may affect measurement validity ASM 2000
Unit weighs approx. 140 kg with dimensions 913 x 558 x 1197 mm (L x W x H) p.3 ↗ Plan handling equipment and floor/bench loading for installation ASM 2000
Software is 21 CFR Part 11 compliant with PDF GMP test and calibration reports and local or LDAP domain authentication p.3 ↗ Suitable for GMP-regulated pharma environments; integrate with corporate LDAP for user management ASM 2000
Runs on Windows 10 with 1x LAN (RJ45), 2x external USB 3.0, 1x HDMI, and a 10" multi-touch full HD screen p.3 ↗ Network and peripheral integration (data export, external display) via standard IT interfaces ASM 2000