SplitFlow 350 — product leaflet
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum SplitFlow 350 turbomolecular vacuum pump datasheet/brochure. The pump has three gas inlet stages/ports designated H0, H1, H2 (plus 'FV' fore-vacuum stage referenced in compression ratios); pumping speeds are specified for 'extern' and 'intern' configurations. Covers accessories/options: power supply kit and display unit, mounting kit, venting accessories, safety valve, power failure venting unit, electrical adapter, mains cable, connecting cable.
Section-by-Section Technical Breakdown
Introduces the SplitFlow 350 turbomolecular pump highlighting high gas throughput, reduced power consumption, an integrated sensor system for maximum safety, and hybrid bearings for robustness and high temperature resistance. Continuous lubricant filtration is claimed to lower cost of ownership.
- High gas throughput with reduced power consumption
- Integrated sensor system for maximum safety
- High temperature resistance, robust hybrid bearing
- Cost-efficient through continuous lubricant filtration
Pumping speed data (air at 20 °C, tolerance ±10%) for He and N₂ at stages H0, H1, H2 for external and internal configurations. Ultimate pressure is 1·10⁻⁷ hPa; compression ratios are given for stage pairs FV/H0, H1/H0 and H2/H1.
- He pumping speed: H0 extern 350 l/s / intern 450 l/s; H1 extern 250 l/s / intern 280 l/s; H2 extern 45 l/s / intern 50 l/s
- N₂ pumping speed: H0 extern 350 l/s / intern 450 l/s; H1 extern 300 l/s / intern 370 l/s; H2 extern 40 l/s / intern 45 l/s
- Ultimate pressure 1·10⁻⁷ hPa
- Compression ratio FV/H0: He 1·10⁸, N₂ 1·10⁹; H1/H0: He 50, N₂ 1.5·10⁴; H2/H1: He 1·10², N₂ 1.5·10⁵
Gas throughput at full rotational speed is specified per stage for He and N₂, showing the H2 stage handles the highest throughput (22 hPa·l/s He, 8 hPa·l/s N₂).
- He: H0 7.3 hPa·l/s, H1 10 hPa·l/s, H2 22 hPa·l/s
- N₂: H0 2 hPa·l/s, H1 3 hPa·l/s, H2 8 hPa·l/s
Operating and installation data: rotational speed (variable 60–100%), max fore-vacuum pressure, I/O interfaces, port sizes, run-up time, protection degree, supply voltage, bearing type, mounting orientation, noise level, weight and outlet flange.
- Rotational speed 66000 rpm ±2%, variable 60–100%
- Max fore-vacuum pressure for N₂: 10 hPa
- I/O interfaces: RS485, Remote
- Port sizes: H0 D = 136.5 mm, H1 601 · 130.4 mm, H2 31.5 · 135.5 mm (as printed)
- Run-up time 6 min; protection degree IP44, Type 12
- Input voltage 24 V DC ±10%; hybrid bearing; arbitrary mounting orientation
- Sound pressure level (EN ISO 2151) 50 dB(A); weight approx. 7.6 kg
- Flange (out): DN 16 ISO-KF, G 3/8"
Lists available accessories and options for the SplitFlow 350, covering power supply, venting, safety and electrical connection components.
- Power supply kit and display unit
- Mounting kit
- Venting accessories and safety valve
- Power failure venting unit
- Electrical adapter, mains cable, connecting cable
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Rotational speed is 66000 rpm (±2%) and can be varied between 60% and 100% of full speed. p.3 ↗ | Speed reduction can be used to tune throughput/power; performance data in the datasheet assumes full speed. | SplitFlow 350 |
| Ultimate pressure is 1·10⁻⁷ hPa. p.2 ↗ | Sets the base-pressure limit of the vacuum system; suitable for UHV-adjacent applications only with appropriate backing. | SplitFlow 350 |
| Pumping speed for He at H0 is 350 l/s (extern) and 450 l/s (intern); N₂ at H0 is also 350/450 l/s. p.2 ↗ | External vs internal configuration significantly changes effective pumping speed (up to ~28%), so placement matters when sizing the pump. | SplitFlow 350, H0 stage |
| H2 stage pumping speed is only 45 l/s (He extern) / 50 l/s (intern) but delivers highest gas throughput (22 hPa·l/s He). p.2 ↗ | Use the H2 port for high gas-load duties, H0 for maximum speed/ultimate pressure. | SplitFlow 350, H2 stage |
| Maximum fore-vacuum pressure for N₂ is 10 hPa. p.3 ↗ | The backing pump must maintain ≤10 hPa fore-line pressure; determines required backing pump size and crossover. | SplitFlow 350 |
| Compression ratio FV/H0 is 1·10⁸ for He and 1·10⁹ for N₂. p.2 ↗ | Very high compression means fore-vacuum backstreaming/pressure transients are strongly suppressed at the inlet. | SplitFlow 350, FV/H0 |
| Compression ratio H1/H0 is 50 for He but 1.5·10⁴ for N₂. p.2 ↗ | Light gases (He/H₂) migrate far more readily between stages; differential pumping between H1 and H0 is much weaker for He. | SplitFlow 350, H1/H0 stages |
| Compression ratio H2/H1 is 1·10² for He and 1.5·10⁵ for N₂. p.3 ↗ | For He, H2 offers limited isolation from H1; relevant when H2 is used for differential pumping of light gases. | SplitFlow 350, H2/H1 stages |
| Run-up time to full speed is 6 minutes. p.3 ↗ | System pump-down sequencing must account for 6 min before full pumping performance is available. | SplitFlow 350 |
| Input voltage is 24 V DC ±10%. p.3 ↗ | Specify a regulated 21.6–26.4 V DC supply (e.g. the offered power supply kit); standard industrial 24 V DC buses are compatible. | SplitFlow 350 |
| Protection degree is IP44 / Type 12. p.3 ↗ | Suitable for splash-protected industrial environments but not wash-down or outdoor exposure. | SplitFlow 350 |
| Sound pressure level is 50 dB(A) per EN ISO 2151. p.3 ↗ | Quiet enough for laboratory placement without acoustic enclosures. | SplitFlow 350 |
| Weight is approximately 7.6 kg and mounting orientation is arbitrary. p.3 ↗ | Simplifies integration; no orientation restrictions and light mounting structures suffice (use mounting kit). | SplitFlow 350 |
| Hybrid bearing design with continuous lubricant filtration. p.1 ↗ | Combines robustness/temperature resistance with reduced maintenance costs; relevant for long-term duty planning. | SplitFlow 350 |
| Outlet flange is DN 16 ISO-KF with G 3/8" connection. p.3 ↗ | Downstream fore-line plumbing must match DN 16 ISO-KF / G 3/8" fittings. | SplitFlow 350 |
| I/O interfaces are RS485 and Remote. p.3 ↗ | Enables serial monitoring/control integration (e.g. display unit) plus hardwired remote control. | SplitFlow 350 |
| Integrated sensor system for maximum safety and a power failure venting unit are available. p.1 ↗ | Specify the power failure venting option where sudden power loss could damage the process or pump. | SplitFlow 350 |
| Pumping speed data tolerance is ±10% (air at 20 °C). p.2 ↗ | Apply a conservative margin when sizing systems based on the stated l/s values. | SplitFlow 350, all speed data |