ASI 35 — product leaflet
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum ASI 35 helium/hydrogen leak detector (spraying and sniffing tests). Accessories/options listed: cables, sniffer module, sniffer probe with different hose lengths, communication interfaces. Inlet port options DN 25 ISO-KF or DN 16 ISO-KF.
Section-by-Section Technical Breakdown
The ASI 35 is presented as flexible, accurate, and robust. Its modular design has a small footprint for easy integration into leak detection systems, with high sensitivity for short cycle times and high production throughput, plus long maintenance intervals.
- Flexible: modular design, small footprint, easy integration into any leak detection system
- Accurate: high sensitivity, repeatable measurements, suitable for short cycle time and high production throughput
- Robust: high reliability, long maintenance intervals
Lists available accessories and options for the ASI 35. Includes sniffer hardware and industrial-standard communication interfaces.
- Cable
- Sniffer module
- Sniffer probe with different hose length
- Communication interface compatible with industrial standards
Core specifications of the ASI 35 covering test methods, sensitivity, gas compatibility, interfaces, power, dimensions and weights. The unit consists of three physical modules: vacuum module, electronic module, and control panel.
- Test methods: spraying and sniffing tests
- Detectable gases: ⁴He; ³He; H₂
- Min. detectable leak rate (Helium, spraying): 5·10⁻¹² mbar l/s / 5·10⁻¹³ Pa m³/s
- Min. detectable leak rate (Helium, sniffing): 1.5·10⁻⁸ mbar l/s / 1.5·10⁻⁹ Pa m³/s
- Helium pumping speed: 6 l/s
- Max. test inlet pressure: 18 mbar
- Start-up time (20 °C): <3 min
- Operating temperature: 10–45 °C
- Communication interfaces: I/O, RS-232, Profibus, Profinet
- Inlet port: DN 25 ISO-KF, DN 16 ISO-KF
- Dimensions (L×W×H): vacuum module 279×264×235 mm; electronic module 216×317×111 mm; control panel 241×54×133 mm
- Weight approx.: vacuum module 15 kg / electronic module 5 kg / control panel 1.3 kg
- Power supply: 90–240 V (50/60 Hz); max. power consumption 300 W
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Minimum detectable leak rate for helium in spraying (vacuum) test: 5·10⁻¹² mbar l/s (5·10⁻¹³ Pa m³/s) p.2 ↗ | Suitable for high-sensitivity vacuum leak testing; spec must not be confused with the much higher sniffing detection limit | ASI 35 |
| Minimum detectable leak rate for helium in sniffing test: 1.5·10⁻⁸ mbar l/s (1.5·10⁻⁹ Pa m³/s) p.2 ↗ | For atmospheric sniffing applications, design test criteria around this ~4 orders of magnitude worse limit than vacuum mode | ASI 35 (sniffing operation) |
| Detectable tracer gases: ⁴He, ³He and H₂ p.2 ↗ | Hydrogen test possible where helium cost/supply is an issue; verify gas-specific sensitivities before certifying | ASI 35 |
| Helium pumping speed is 6 l/s p.2 ↗ | Determines response/recovery time of vacuum test cycles and pump-down of small test volumes | ASI 35 |
| Maximum test inlet pressure is 18 mbar p.2 ↗ | Test parts must be evacuated below 18 mbar before opening the leak detector inlet; use a backing/roughing pump above this pressure | ASI 35 |
| Start-up time at 20 °C is less than 3 minutes p.2 ↗ | Enables short shift startup in production environments; allow ~3 min before first valid measurement | ASI 35 |
| Operating temperature range is 10–45 °C p.2 ↗ | Installation site must be climate-controlled; outside this range accuracy/reliability are not guaranteed | ASI 35 |
| Communication interfaces available: I/O, RS-232, Profibus, Profinet p.2 ↗ | Direct PLC integration possible in automated production lines without additional gateways for Profibus/Profinet environments | ASI 35 |
| Inlet ports available: DN 25 ISO-KF and DN 16 ISO-KF p.2 ↗ | Select KF flange hardware/adapters accordingly; DN 16 reduces conductance versus DN 25 | ASI 35 |
| Power supply is 90–240 V, 50/60 Hz with max. 300 W power consumption p.2 ↗ | Single-phase worldwide supply compatibility; size supply/breaker for at least 300 W per unit | ASI 35 |
| System splits into vacuum module (279×264×235 mm, 15 kg), electronic module (216×317×111 mm, 5 kg) and control panel (241×54×133 mm, 1.3 kg) p.2 ↗ | Modular layout allows distributed installation; mounting plates and enclosures must be dimensioned per module, total mass ~21.3 kg | ASI 35 |
| Optional sniffer probe is available with different hose lengths p.1 ↗ | Hose length selection affects sniffing response time and the stated 1.5·10⁻⁸ mbar l/s sniffing limit; choose shortest practical length | ASI 35 with sniffer module/probe option |