Induztr
Engineering Digest Document Viewer Related Models

A 1204 H — product leaflet

Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum multi-stage Roots vacuum pumps, datasheet A 124-1204 H, covering three models: A 124 H, A 604 H, A 1204 H. Dry, energy-efficient pumps for semiconductor and coating applications; SEMI S2 and UL 61010 certified. Document includes a performance data table and dimensional drawings.

Section-by-Section Technical Breakdown

pp. 1
Product Overview

The A 124-1204 H series comprises dry, multi-stage Roots vacuum pumps targeted at semiconductor and coating applications. Key selling points are high performance, dry and energy-efficient operation, and adjustable settings for process flexibility. The pumps are SEMI S2 and UL 61010 certified.

  • Multi-stage Roots principle, dry operation (no lubricant in pumping chamber)
  • Energy-efficient design
  • Adjustable settings to match individual process requirements
  • Certified to SEMI S2 and UL 61010
  • Intended for semiconductor and coating applications
pp. 2
Technical Data, Pumping Speed and Dimensional Drawing

A comparison table gives pumping speed, ultimate pressure, power consumption, cooling water flow, noise level, weight, dimensions and flange sizes for the A 124 H, A 604 H and A 1204 H, mostly at 50/60 Hz. Dimensional drawings accompany the data. Inlet flanges grow with pump size (DN 50 to DN 160) while outlet is DN 40 on all models.

  • Nominal pumping speed: A 124 H 95/110 m³/h; A 604 H 480/560 m³/h; A 1204 H 1050/1150 m³/h (50/60 Hz)
  • Ultimate pressure: A 124 H 5·10⁻²/2·10⁻² hPa; A 604 H and A 1204 H 3·10⁻³/2·10⁻³ hPa (50/60 Hz)
  • Power consumption at ultimate pressure: A 124 H 1.3/1.5 kW; A 604 H and A 1204 H 2.1/2.3 kW (50/60 Hz)
  • Cooling water flow: A 124 H 2.3–3 l/min; A 604 H and A 1204 H 2.5–4 l/min
  • Sound pressure level: A 124 H <65 dB(A); A 604 H <68 dB(A); A 1204 H <69 dB(A)
  • Weight approx.: A 124 H 240 kg; A 604 H 370 kg; A 1204 H 530 kg
  • Dimensions L×W×H: 961×388×579 mm (A 124 H); 961×388×776 mm (A 604 H); 961×388×911 mm (A 1204 H)
  • Gas inlet/outlet: DN 50/DN 40 (A 124 H); DN 100/DN 40 (A 604 H); DN 160/DN 40 (A 1204 H)

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
All three models are certified to SEMI S2 and UL 61010. p.1 ↗ Suitable for semiconductor fab tool integration and electrical safety compliance without additional certification effort. whole series (A 124 H, A 604 H, A 1204 H)
The pumps operate dry (no oil in the pumping chamber). p.1 ↗ No oil mist exhaust or oil contamination of the process; no oil changes for the pump chamber, simplifying maintenance in clean processes. whole series
A 124 H nominal pumping speed is 95 m³/h at 50 Hz and 110 m³/h at 60 Hz. p.2 ↗ Site line frequency affects delivered speed by ~15%; use 60 Hz value if the pump is run on a 60 Hz supply or via frequency converter. A 124 H
A 604 H nominal pumping speed is 480/560 m³/h (50/60 Hz). p.2 ↗ Mid-size model selection for throughputs between the A 124 H and A 1204 H; speed advantage at 60 Hz is ~80 m³/h. A 604 H
A 1204 H nominal pumping speed is 1050/1150 m³/h (50/60 Hz). p.2 ↗ Largest model in the series; reserve floor space for its 911 mm height and 530 kg weight during layout planning. A 1204 H
Ultimate pressure of A 604 H and A 1204 H is 3·10⁻³ hPa at 50 Hz and 2·10⁻³ hPa at 60 Hz, while A 124 H only reaches 5·10⁻²/2·10⁻² hPa. p.2 ↗ For processes requiring base pressures in the 10⁻³ hPa range, the A 124 H is not suitable; choose A 604 H or A 1204 H. A 124 H vs. A 604 H / A 1204 H
Power consumption at ultimate pressure: 1.3/1.5 kW (A 124 H) and 2.1/2.3 kW (A 604 H, A 1204 H) at 50/60 Hz. p.2 ↗ Supply sizing can be kept modest; even the largest pump draws under 2.3 kW at ultimate pressure, supporting the energy-efficient claim. whole series
Cooling water flow requirement: 2.3–3 l/min (A 124 H), 2.5–4 l/min (A 604 H and A 1204 H). p.2 ↗ Water cooling is mandatory; provision cooling water circuits sized to at least 4 l/min per pump and plan plumbing connections accordingly. whole series
Sound pressure levels: <65 dB(A) for A 124 H, <68 dB(A) for A 604 H, <69 dB(A) for A 1204 H. p.2 ↗ All models fall below typical 70–80 dB(A) workplace noise thresholds; no special acoustic enclosure generally required. whole series
Weights: approx. 240 kg (A 124 H), 370 kg (A 604 H), 530 kg (A 1204 H). p.2 ↗ Lifting equipment rated for ≥550 kg needed for the A 1204 H installation; verify floor/rack load capacity. whole series
Footprint is identical for all models: 961 mm length × 388 mm width; height varies (579/776/911 mm). p.2 ↗ A common mounting footprint allows model interchange in existing installations, with only vertical clearance varying between models. whole series
Gas inlet/outlet connections: DN 50/DN 40 (A 124 H), DN 100/DN 40 (A 604 H), DN 160/DN 40 (A 1204 H). p.2 ↗ Vacuum plumbing and flange hardware must match inlet DN per model; the common DN 40 outlet simplifies foreline design across the series. whole series (per model)
Settings are adjustable to match individual process requirements. p.1 ↗ Operating parameters (e.g. speed) can be tuned to the process rather than fixed at nominal values; consult the manual for actual adjustment ranges. whole series