Induztr
Engineering Digest Document Viewer Related Models

ATH 500 M — product leaflet

Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum ATH 500 M and ATH 500 MT turbomolecular vacuum pumps data brochure (3 pages, dated 01.07.2025). MT variant adds an integrated heating system (max 65 °C) for corrosive/chemical process conditions. Covers accessories, options and key technical data.

Section-by-Section Technical Breakdown

pp. 1
Overview and Product Features

Introduces the ATH 500 M/MT as compact turbomolecular pumps with integrated drive unit, installable in any orientation. Low noise and vibration suit semiconductor and analytical applications. MT variant offers integrated heating for demanding chemical conditions; both use frequency-controlled DC motors.

  • Compact: small footprint, integrated drive unit, installation in any orientation
  • Quiet: low noise/vibration for vibration-sensitive semiconductor and analytical use
  • Flexible: compatible with a variety of backing pumps; MT with integrated heating; frequency-controlled DC motors
  • Accessories/options: installation kit, cooling water valve, purge gas valve, air cooling kit, handheld remote controller, 48 VDC power supply, inlet protection screen, valve cable and coil, various inlet flanges, various user interfaces, temperature management system for corrosive processes
pp. 2-3
Technical Data, Pumping Performance and Dimensions

Dimensional drawing and full technical data table for ATH 500 M vs MT: pumping speeds, compression ratios (ISO 5302), pressures, gas throughputs, electrical and cooling data, acoustics/vibration, dimensions, weights, flange and interface details.

  • Pumping speed N₂: 350–550 l/s (M) / 550 l/s (MT); H₂: 170–190 / 190 l/s
  • Compression ratio (ISO 5302): N₂ 2·10⁷, Ar 8·10⁶, He 1·10⁴, H₂ 2·10²
  • Ultimate pressure < 1·10⁻⁸ mbar; max forevacuum N₂ 2.6 mbar
  • Rotational speed 50000 rpm, variable 15000 rpm to nominal; run-up to 90% < 2 min (exhaust < 0.1 mbar)
  • 48 VDC supply; start-up power max 560 W; stand-by 100 W; baking max 120 °C; MT heating max 65 °C
  • Cooling water 60 l/h at 15–25 °C; recommended backing pump ACP40
  • Noise < 42 dB(A) at 1 m; vibration < 0.01 μm; Ø180 x 159.3–209.3 mm; 17 kg (M) / 18 kg (MT)
  • Interfaces: DN 100/DN 160 (M) / DN 160 (MT) inlet, DN 25 (M) / DN 40 ISO-KF (MT) outlet; purge 1/8 G (M) / 1/4 VCR (MT); on-board control: Profibus/Remote (M), Remote (MT)

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Pumping speed N₂ is 350–550 l/s (ATH 500 M) and 550 l/s (ATH 500 MT); He 310–390 / 390 l/s; H₂ 170–190 / 190 l/s p.2 ↗ Size backing pump and process time budgets per gas type; MT gives the upper end across all gases ATH 500 M / MT
Compression ratios (ISO 5302): N₂ 2·10⁷, Ar 8·10⁶, He 1·10⁴, H₂ 2·10² p.2 ↗ Hydrogen compression is low — foreline pressure dominates ultimate pressure in H₂-rich processes both variants
Ultimate pressure < 1·10⁻⁸ mbar p.2 ↗ Suitable for high/ultra-high vacuum applications both variants
Max. forevacuum N₂ 2.6 mbar (Ar 3.3 mbar, He 1.0 mbar, H₂ 0.25 mbar) p.2 ↗ Backing pump must maintain these ceilings; H₂ foreline must stay ≤ 0.25 mbar both variants
Max. inlet vacuum (tolerable inlet pressure) at N₂: 1.0 mbar for M but only 0.04 mbar for MT; He: 10 mbar (M) vs > 0.1 mbar (MT) p.2 ↗ The heated MT variant tolerates much lower inlet pressure at start — do not expose MT to high inlet pressures comparable to the M variant M vs MT specifically
Gas throughput at final speed N₂: 4000 sccm (M) vs 500 sccm (MT); He/H₂ > 10000 vs > 1000 sccm p.3 ↗ M variant has ~8–10× higher gas throughput capacity; MT is for near-base-pressure corrosive service M vs MT
Rotational speed 50000 rpm, variable from 15000 rpm to nominal speed p.3 ↗ Speed can be reduced for process control or vibration minimization both variants
Run-up time to 90% final speed < 2 min with exhaust pressure < 0.1 mbar p.3 ↗ Venting/backing pressure during start-up must be < 0.1 mbar to meet this spec both variants
Controller power supply 48 VDC; max start-up consumption 560 W; nominal (stand-by) power 100 W p.3 ↗ Size the 48 VDC PSU (offered as accessory) for ≥ 560 W start-up peaks both variants
MT max heating temperature 65 °C; both variants max baking temperature 120 °C p.3 ↗ Bakeout to 120 °C permissible; MT heating limited to 65 °C for corrosive gas operation MT heating / both for baking
Recommended cooling water flow 60 l/h at 15–25 °C p.3 ↗ Provide chilled water circuit in this range; air cooling kit available as option both variants
Recommended backing pump: ACP40 p.3 ↗ Reference sizing for fore-vacuum selection both variants
Sound pressure < 42 dB(A) (ISO 2151, 1 m) and vibration < 0.01 μm at nominal speed p.3 ↗ Can be mounted in vibration-sensitive semiconductor/analytical setups without heavy isolation both variants
Dimensions Ø180 x 159.3–209.3 mm (height depends on flange variant); weight ~17 kg (M) / 18 kg (MT) p.2 ↗ Compact envelope allows installation in any orientation; account for flange-dependent height both variants
Gas inlet DN 100/DN 160 (M) or DN 160 (MT); outlet DN 25 ISO-KF (M) / DN 40 ISO-KF (MT); N₂ purge flange 1/8 G (ISO 228) for M, 1/4 VCR for MT p.3 ↗ Plumbing and purge line fittings must match variant; MT VCR purge suits corrosive service M vs MT
Recommended purge flow rate 50 sccm p.3 ↗ Set purge gas flow to 50 sccm for corrosive operation both variants
Permissible magnetic field (radial) at 10 cm: < 12 mT p.3 ↗ Keep magnetic sources (e.g. other pumps, magnets) beyond field limits to protect motor/bearing operation both variants
On-board control: Profibus and Remote (M); Remote only (MT); stainless steel inlet screen is factory setting on both p.3 ↗ Choose M if Profibus integration is required; MT limited to remote control interface M vs MT

Related models