OmniDetect — product leaflet
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum OmniDetect leak test module — fully automated integral vacuum ('inside-out') leak testing chamber. Variants: chamber size DN300 or DN500; backing pump oil-lubricated DuoVane or dry HiScroll; optional turbomolecular pump HiPace 80 up to HiPace 300 (customizable to HiPace 700); integrated leak tester ASI 35; test pressure up to 50 bar (customizable to 200 bar).
Section-by-Section Technical Breakdown
OmniDetect is a leak test module positioned as high performance with low detection limits and a fully automated measuring process. It is presented as a tailor-made system solution covering conception, production and service, with a modular design allowing individual configuration for specific process requirements.
- High performance with low detection limits and fully automated measuring process
- Tailor-made system solution from conception through production to service
- Modular design for individual configuration to specific process requirements
Core specification table covering the test method, tracer gases, detection limits, cycle times, chamber sizes, pump configuration, and automation/interface options. A dimensional drawing of the module is included.
- Integral vacuum test ('inside-out') method
- Tracer gases: He, He/N₂ mixture or forming gas
- Min. detection limit 1·10⁻⁷ mbar·l/s
- Ultimate pressure < 5·10⁻⁶ mbar
- Min. cycle time 2 min (DN300), 3 min (DN500)
- Test pressure (absolute): variable, up to 50 bar
- Chamber: cubic DN300 or cubic DN500
- Leak tester: ASI 35; backing pump: oil-lubricated DuoVane or dry HiScroll
- Optional turbomolecular pump HiPace 80 up to HiPace 300
- Integrated He test leak for calibration
- PLC with full HD touch display; interfaces USB, Ethernet, OPC-UA
- Part handling not included
Operating duty, machine availability and physical dimensions of the module, with footnotes on extended detection limits, higher pressures/pumps and transport height.
- Operating time: 7 days × 24 h, or paragraph-wise operation in R&D
- Machine availability > 97 %
- Dimensions (L×W×H): 1800 × 1100 × 2400 mm
- 1·10⁻⁸ mbar·l/s detection limit possible for some configurations with long evacuation times
- Test pressure customizable up to 200 bar with additional design expanses
- Turbomolecular pump customizable up to HiPace 700 with housing change
- Signal lamp demountable; transport height 2.0 m
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Minimum detection limit is 1·10⁻⁷ mbar·l/s, extendable to 1·10⁻⁸ mbar·l/s for some configurations with long evacuation times. p.2 ↗ | Spec a long evacuation phase if sub-1e-7 leak rates must be resolved; verify configuration feasibility with the manufacturer for 1e-8 applications. | whole series, config-dependent for 1e-8 |
| Minimum cycle time is 2 min for DN300 and 3 min for DN500 chambers. p.2 ↗ | Determines maximum throughput; actual cycle time depends on pump configuration, required detection limit, and part evacuation/filling times. | DN300 = 2 min, DN500 = 3 min |
| Test pressure (absolute) is variable up to 50 bar, customizable to 200 bar with additional design effort. p.2 ↗ | Pressure-tested parts beyond 50 bar require a custom design order; chamber and vessel design must be validated accordingly. | whole series; >50 bar is a custom option |
| Test method is integral vacuum ('inside-out') with tracer gases He, He/N₂ mixture or forming gas. p.2 ↗ | Parts must be pressurized with the tracer gas inside the evacuated chamber; forming gas compatibility allows lower-cost tracer mixtures. | whole series |
| Ultimate pressure of the system is < 5·10⁻⁶ mbar. p.2 ↗ | Sets the achievable background level before testing; relevant for achieving the lowest detection limits. | whole series |
| Leak tester is the ASI 35 with an integrated He test leak for calibration. p.2 ↗ | No external calibration leak needed; automated in-situ calibration supports unmanned 24/7 operation. | whole series |
| Backing pump choice is oil-lubricated DuoVane or dry HiScroll; optional turbomolecular pump HiPace 80–300, customizable up to HiPace 700 with housing change. p.2 ↗ | Dry pump option suits oil-sensitive applications; turbo pump selection trades footprint/housing changes against pump-down speed and detection limit. | whole series; HiPace up to 700 is a custom option |
| Footprint is 1800 × 1100 × 2400 mm (L×W×H); signal lamp is demountable, reducing transport height to 2.0 m. p.3 ↗ | Plan installation floor space accordingly; for shipping through <2.4 m door heights, remove the signal lamp. | whole series |
| Machine availability is specified as > 97 % with continuous operation 7 days × 24 h, or paragraph-wise operation for R&D. p.3 ↗ | Suitable for production-line duty; downtime budget must remain under ~3 %. | whole series |
| Control is a PLC with full HD touch display; interfaces are USB, Ethernet and OPC-UA. p.2 ↗ | OPC-UA enables integration into Industry 4.0/MES environments; plan network connectivity and USB access for data export. | whole series |
| Part handling is not included in the module. p.2 ↗ | Load/unload automation or manual handling must be sourced and integrated separately into the cell layout. | whole series |