OmniGrade — product leaflet
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum OmniGrade leak detection / outgassing measurement system. Modular system available in 2-chamber and 3-chamber configurations; configured with either PrismaPro or HiQuad Neo quadrupole mass spectrometers.
Section-by-Section Technical Breakdown
The OmniGrade is a high-performance modular vacuum measuring system with fully automated measuring process and low detection limits. Testing follows generic standards GSA 07 1221 and GSA 07 2221, and the modular design allows individual configuration for specific process requirements.
- Fully automated measuring process with low detection limits
- Testing according to Generic Standards GSA 07 1221 and GSA 07 2221
- Modular design enabling individual configuration for specific process requirements
The OmniGrade achieves an ultimate pressure below 5·10⁻¹⁰ hPa using a HiPace 300 M turbopump with dry forevacuum pump. Total and partial pressure are managed with a Pirani/Bayard-Alpert HPT 200 gauge and PrismaPro or HiQuad Neo quadrupole mass spectrometers. Two sample-size configurations are offered: 2-chamber (max 700×700×700 mm³) and 3-chamber (max 150×150×150 mm³).
- Ultimate pressure: < 5·10⁻¹⁰ hPa
- Sample size: max 150×150×150 mm³ (3-chamber system); max 700×700×700 mm³ (2-chamber system)
- Vacuum pumps: HiPace 300 M turbopump and dry forevacuum pump
- Total pressure management: Pirani/Bayard-Alpert HPT 200
- Partial pressure management: quadrupole mass spectrometers PrismaPro or HiQuad Neo
- Heating: heating jackets up to 180 °C
- Cooling: temperature control system with silicone-free thermofluid
- Calibration: UHV all-metal leak valve and calibration gas
- Pumping speed determination: with calibrated N₂ leak
- User interface: PLC with 19" full HD touch display
- Overall dimensions (L×W×H): 1.1 × 1.9 × 2.3 m
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Ultimate pressure of the OmniGrade system is below 5·10⁻¹⁰ hPa. p.2 ↗ | Suitable for UHV-grade outgassing/leak rate measurements; process chamber bakeout and cleanliness must support this pressure level. | whole series |
| 3-chamber system accepts samples up to 150×150×150 mm³; 2-chamber system up to 700×700×700 mm³. p.2 ↗ | Chamber configuration must be selected based on maximum sample size; larger samples require the 2-chamber variant. | 3-chamber and 2-chamber variants |
| High vacuum is generated by a HiPace 300 M turbopump combined with a dry forevacuum pump. p.2 ↗ | No oil mist contamination of measurement volume; fore-line infrastructure need not be planned since dry backing pump is included. | whole series |
| Total pressure is measured with a Pirani/Bayard-Alpert HPT 200 combination gauge. p.2 ↗ | Wide pressure measurement span from rough vacuum (Pirani) down to UHV (hot cathode) without gauge changeover. | whole series |
| Partial pressure analysis is performed by PrismaPro or HiQuad Neo quadrupole mass spectrometers. p.2 ↗ | Two RGA options exist; selection affects mass range and configuration choices at ordering. | whole series |
| System heating uses heating jackets up to 180 °C. p.2 ↗ | Bakeout to 180 °C is possible for degassing; materials of samples and fixtures must tolerate this temperature. | whole series |
| Cooling is provided by a temperature control system using silicone-free thermofluid. p.2 ↗ | Thermofluid supply/integration must be planned; silicone-free fluid avoids contamination of the UHV system. | whole series |
| Calibration is performed with a UHV all-metal leak valve and calibration gas. p.2 ↗ | On-site calibration of detection limits is possible without external reference leak removal/insertion. | whole series |
| Pumping speed is determined using a calibrated N₂ leak. p.2 ↗ | Effective pumping speed can be verified in operation using the integrated calibrated N₂ reference leak. | whole series |
| User interface is a PLC with a 19" full HD touch display. p.2 ↗ | Automated measurement sequences are run from the built-in PLC; no external PC is required for operation. | whole series |
| System footprint dimensions are 1.1 × 1.9 × 2.3 m (L×W×H). p.2 ↗ | Installation planning must reserve floor space of ~2.1 m² and ceiling height above 2.3 m plus service access. | whole series |
| Testing is performed according to generic standards GSA 07 1221 and GSA 07 2221. p.1 ↗ | Measurement results can be aligned with these standards; verify standard compliance requirements for your process qualification. | whole series |