HiScroll 46 — product leaflet
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum HiScroll 46 scroll vacuum pump datasheet. Variants/options listed: single-phase VSD, single-phase VSD with integrated pressure gauge, three-phase version, C-version (corrosive gas), plus accessories: tip seal replacement kit, tip seal and bearing replacement kit, gas-ballast valve kit, bellow kit, cover kit, electronics kit, exhaust silencer, inlet filter, automated vacuum safety valve, OmniControl control device, EtherCAT and Profinet converters.
Section-by-Section Technical Breakdown
Highlights the HiScroll 46 as a high-performance, quiet, compact scroll pump with an IPM synchronous motor producing less heat than conventional drives. Lists available accessories, spare parts, and version options including corrosive-gas and network-converter variants.
- Fastest pump-down time in its class; highly energy-efficient interior permanent magnet (IPM) synchronous motor generating less heat than conventional drives
- Sound pressure level less than 47 dB(A) in single-phase version
- Smallest and lightest pump in its performance class, suited for laboratory applications
- Accessories/spare parts: tip seal replacement kit, tip seal and bearing replacement kit, gas-ballast valve kit, bellow kit, cover kit, electronics kit, exhaust silencer, inlet filter
- Options: automated vacuum safety valve, OmniControl control device, EtherCAT converter, Profinet converter
- Versions: single-phase VSD, single-phase VSD with integrated pressure gauge, three-phase, C-version for corrosive gas
Key performance figures for the HiScroll 46 including pumping speed, ultimate pressures at various gas-ballast settings, power consumption, dimensions, and connection sizes. Pumping speed curves are given for air at 20 °C with ±10% tolerance at 50 Hz and 60 Hz.
- Pumping speed: 43 m³/h (50 Hz and 60 Hz curves shown, tolerance ±10%)
- Ultimate pressure (gas-ballast closed): 0.007 hPa (mbar)
- Ultimate pressure (gas-ballast open stage 1): ≤0.03 hPa (mbar)
- Ultimate pressure (gas-ballast open stage 2): ≤0.2 hPa (mbar)
- Power consumption at ultimate pressure: 295 W
- Motor speed 720–1800 min⁻¹; nominal 1800 min⁻¹
- Sound pressure level ≤47 dB(A) (ISO 3744, accuracy grade 2)
- Weight approx. 32 kg; dimensions 463 x 299 x 334 mm (L x W x H)
- Gas inlet DN 40 ISO-KF; outlet DN 25 ISO-KF
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Pumping speed is 43 m³/h for air at 20 °C, with a ±10% tolerance on the speed curves p.2 ↗ | Size the backing/fore-vacuum application with a 10% margin; curves exist for both 50 Hz and 60 Hz supply | HiScroll 46 |
| Ultimate pressure with gas-ballast valve closed is 0.007 hPa (mbar) p.2 ↗ | Defines best achievable base vacuum; keep gas-ballast closed for finest ultimate pressure | HiScroll 46 |
| Gas-ballast open at stage 1 degrades ultimate pressure to ≤0.03 hPa; stage 2 to ≤0.2 hPa p.2 ↗ | Operating with gas ballast for condensate handling costs roughly 4x–30x in ultimate pressure; select stage accordingly | HiScroll 46 with gas-ballast valve kit |
| Power consumption at ultimate pressure is 295 W p.2 ↗ | Low standby power draw; useful for sizing lab power budgets and energy cost estimates | HiScroll 46 |
| Motor speed range is 720–1800 min⁻¹ with nominal speed 1800 min⁻¹ p.2 ↗ | Speed can be modulated down (VSD variants) for process control or reduced throughput | HiScroll 46 |
| Sound pressure level ≤47 dB(A) per ISO 3744 accuracy grade 2 (single-phase version <47 dB(A)) p.1 ↗ | Suitable for noise-sensitive lab environments without additional acoustic enclosure | HiScroll 46 |
| Weight is approx. 32 kg; dimensions 463 x 299 x 334 mm p.2 ↗ | Compact footprint for lab benches; plan handling/support for ~32 kg | HiScroll 46 |
| Gas inlet is DN 40 ISO-KF, outlet DN 25 ISO-KF p.2 ↗ | Specify matching KF clamps, centering rings and a DN 25 foreline; inlet is larger than outlet | HiScroll 46 |
| IPM synchronous motor generates less heat than conventional drives p.1 ↗ | Reduced thermal load on the pumped process and enclosure; beneficial for heat-sensitive applications | HiScroll 46 |
| A C-version is available for corrosive gas applications p.1 ↗ | For corrosive processes, order the C-version rather than the standard pump to avoid premature wear | HiScroll 46 C-Version |
| Fieldbus integration is available via EtherCAT or Profinet converters, and the OmniControl control device can be added p.1 ↗ | Plan remote monitoring/control wiring if process automation is required | HiScroll 46 with accessories |
| Maintenance kits exist for tip seal alone and tip seal plus bearing replacement p.1 ↗ | Stock appropriate kits based on duty; tip seals are the primary wear item on scroll pumps | HiScroll 46 |
| Automated vacuum safety valve, inlet filter, and exhaust silencer are available as accessories p.1 ↗ | Use the safety valve to protect the chamber on power failure; filter/silencer for particulate and noise control | HiScroll 46 |
| Three supply variants exist: single-phase VSD, single-phase VSD with integrated pressure gauge, and three-phase p.1 ↗ | Match variant ordering to available site power; integrated gauge variant saves a separate measurement port | HiScroll 46 |