HiLobe 1002 — product leaflet
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum HiLobe 1002 and HiLobe 1302 vacuum boosters (Roots-type). Same envelope and flanges, differing mainly in rotational speed range (1002: 1500–1600 rpm; 1302: 1500–7500 rpm) and max differential pressure. Options/accessories listed include Profinet converter, barrier gas kit (plastic/stainless steel), gear chamber evacuation, flushing device, temperature/pressure monitoring, extension cable package, OmniControl, horizontal version, drive electronics variants (box, switch-cabinet mounting plate, none), and cooling options (convection, active-air, water).
Section-by-Section Technical Breakdown
Highlights condition monitoring via integrated sensors, IE5 hermetically sealed canned-motor design without shaft seals, variable speed drive, and extended maintenance intervals. Lists available accessories and drive/cooling options.
- Integrated sensor condition monitoring for high up-time
- IE5 hermetically sealed synchronous motor, canned motor design, no shaft seals
- Variable speed drive and optimized rotor geometries
- Options: Profinet converter, barrier gas kit (plastic or stainless steel), gear chamber evacuation, flushing device (spray lance)
- Temperature and pressure monitoring (inlet and exhaust side), extension cable package, OmniControl, horizontal version
- Drive electronics: box version, mounting plate for switch cabinet integration, or without drive electronics
- Cooling options: convection, active-air, water cooling
Pumping speed curves at 50/60 Hz plus core performance and electrical specifications. Both models share nominal pumping speed 260–1050 m³/h, DN100 flanges, and RC 5500 electronic drive unit; they differ in speed range and allowable differential pressure.
- Nominal pumping speed 260–1050 m³/h (both models)
- Rotational speed: HiLobe 1002 1500–1600 rpm; HiLobe 1302 1500–7500 rpm
- Max differential pressure at max speed: 21 hPa (1002), 13 hPa (1302); at min speed 41 hPa (both)
- Flanges: DN 100 ISO-F | DN 100 PN 16 (inlet and exhaust, both models)
- Input voltage 380–480 V AC (±10%), 50/60 Hz; three-phase motor with PTC protection
- Rated power at full speed: 2 kW (1002), 2.5 kW (1302)
- Electronic drive unit RC 5500; recommended mains fuse 16 A
- I/O interfaces: PV.can and RS485
Acoustic, environmental, fluid, and mechanical specifications. Identical for both models: D2 operating fluid, IP54/type 12 enclosure, identical dimensions and weight.
- Sound pressure level (EN ISO 2151) at 1 hPa inlet pressure: 56 dB(A) (1002), 60 dB(A) (1302)
- Operating fluid D2, filling 0.65 l
- Protection degree IP54; type 12
- Integral leak rate 1E-06 Pa·m³/s
- Dimensions 678 x 282 x 240 mm (L x W x H); weight approx. 105 kg
- Ambient temperature 5–40 °C; shipping temperature -10–40 °C
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| HiLobe 1002 runs at 1500–1600 rpm; HiLobe 1302 at 1500–7500 rpm. p.2 ↗ | The 1302 offers far higher speed for throughput; the 1002 is speed-limited for lower speed/longer life applications. | HiLobe 1002 vs 1302 |
| Max differential pressure at max rotational speed: 21 hPa (1002) and 13 hPa (1302). p.2 ↗ | Downstream backing pump sizing must keep differential pressure below these limits at full speed, especially for the 1302. | Both models (values differ per model) |
| Max differential pressure at minimum rotational speed is 41 hPa for both models. p.2 ↗ | Speed reduction can be used to tolerate higher differential pressures during startup/crossover. | Both models |
| Nominal pumping speed 260–1050 m³/h for both models. p.2 ↗ | Both cover the same speed range for pump-down design; actual speed depends on inlet pressure per the 50/60 Hz curves. | Both models |
| Inlet/exhaust flanges are DN 100 ISO-F or DN 100 PN 16. p.2 ↗ | Choose ISO-F for vacuum-grade clamped connections or PN16 for pressure-rated flanged piping; both ports identical. | Both models |
| Supply voltage 380–480 V AC ±10%, 50/60 Hz, three-phase motor with PTC protection. p.2 ↗ | Requires a suitable three-phase supply within tolerance; PTC trip must be wired into the drive/control for motor protection. | Both models |
| Rated power 2 kW (1002) / 2.5 kW (1302); recommended mains fuse 16 A. p.2 ↗ | Size supply and breaker per 16 A fuse recommendation; drive electronics is the RC 5500. | Both models |
| I/O interfaces are PV.can and RS485; Profinet converter available as option. p.1 ↗ | For Profinet integration an external converter accessory must be ordered; native control via PV.can/RS485. | Both models |
| Hermetically sealed IE5 synchronous canned motor with no shaft seals. p.1 ↗ | Eliminates shaft-seal leaks in the vacuum envelope, supporting the 1E-06 Pa·m³/s integral leak rate; no shaft seal maintenance. | Whole series |
| Integral leak rate 1E-06 Pa·m³/s. p.3 ↗ | Suitable for high-vacuum systems; validate system leak-up against this pump contribution. | Both models |
| Operating fluid D2, filling volume 0.65 l. p.3 ↗ | Stock D2 operating fluid and plan lubricant changes around 0.65 l fill per gearbox. | Both models |
| Ambient temperature 5–40 °C; shipping/storage -10–40 °C. p.2 ↗ | Ensure installation environment stays within 5–40 °C; freeze protection needed below -10 °C during transport. | Both models |
| Protection degree IP54, type 12; default cooling is convection (active-air and water cooling optional). p.1 ↗ | IP54 enclosure allows industrial panel-adjacent installation; for high heat loads select active-air or water cooling option. | Both models |
| Sound pressure level at 1 hPa inlet pressure: 56 dB(A) (1002) / 60 dB(A) (1302) per EN ISO 2151. p.3 ↗ | Low noise; both acceptable in laboratory/industrial spaces without special acoustic measures. | Both models |
| Dimensions 678 x 282 x 240 mm, weight approx. 105 kg (both models). p.3 ↗ | Compact footprint but ~105 kg requires lifting equipment or frame handling for installation. | Both models |
| Accessories include barrier gas kit (plastic or stainless steel), gear chamber evacuation, flushing device (spray lance), and temperature/pressure monitoring on inlet and exhaust. p.1 ↗ | Barrier gas and monitoring options should be ordered for processes with condensable media or requiring process diagnostics. | Whole series |