Induztr
Engineering Digest Document Viewer Related Models

c197-e003

Technical Reference Summary

Covered Models & Document Scope

Nexera™ IC ion chromatography system: main configuration with anion suppressor (ICDS-Ai) or cation suppressor (ICDS-Ci), non-suppressor configuration, and application-specific configurations (organic acid analysis, combustion ion chromatography). Document is a product leaflet, not a full datasheet; no complete ordering codes or part numbers are printed.

Technical Data

Parameter Value Condition
Injection volume 50 µL p.7 ↗ cation analysis, ICDS-Ci suppressor
Injection volume 50 µL p.7 ↗ cation analysis without ICDS-Ci suppressor
Detector baseline noise 0.150 µS/cm p.7 ↗ cation analysis without ICDS-Ci suppressor
Detector baseline noise 0.0060 µS/cm p.7 ↗ cation analysis with ICDS-Ci suppressor
Detector baseline noise 0.050 µS/cm p.9 ↗ non-suppressor configuration, Li 25 ppb
Detected ion concentration 0.01 mg/L p.7 ↗ Na, NH4, K, Mg, Ca with ICDS-Ci suppressor
Detected ion concentration 25 ppb p.9 ↗ Li, non-suppressor configuration
Solvent delivery pump type serial double-plunger, metal-free p.4 ↗ built into IC-150
Degassing online degassing unit, small internal volume p.4 ↗ built into IC-150
Anion suppressor ICDS-Ai electrodialytic, reversed-flow eluent channel structure (patent No. 6897772) p.7 ↗ anion configuration
Cation suppressor ICDS-Ci electrodialytic p.7 ↗ cation configuration
Optional detection UV integration p.8 ↗ selective detection option
Simultaneous analysis anion and cation p.8 ↗ pump port 2 (cation)

Related models