NJ-SERVO Instruction Manual
Technical Reference Summary
Covered Models & Document Scope
Covers the Shimadzu NJ-SERVO series electric motor driven actuators NJ-5kNV, NJ-10kNV, NJ-20kNV and NJ-30kNV, together with the associated Servo Amplifier Unit, 4830 controller, cables and optional attachments (swivel head, trunnion bracket). The document is the operating instructions for the actuator as a partly completed machine to be incorporated into a customer-supplied reaction force floor or frame.
Section-by-Section Technical Breakdown
Describes the servo-motor-driven actuator principle, the closed-loop control system and the standard system configuration with 4830 controller, Servo Amplifier Unit, load cell, jig and Windows software. The actuator needs no hydraulic source or oil, uses a center trunnion and an integrated spline detent/bearing, and applies repeated loads via a load cell and jig.
- Servo motor drive with closed-loop feedback from stroke detector and load cell; movement continues until detector output equals input.
- No hydraulic power source or hydraulic oil required; compact installation compared with hydraulic actuators.
- Center trunnion in the middle allows various load tests with trunnion bracket or other brackets.
- Integrated detent and bearing mechanism (ball spline) gives a simple loading mechanism.
- Typical system: actuator main unit, Servo Amplifier Unit, 4830 controller, Windows software, load cell, swivel head, trunnion bracket.
Lists the standard components and accessories supplied with the NJ-SERVO system, including actuator main unit, Servo Amplifier Unit, detection cables, motor/encoder/fan cables, grounding leads, controller, amplifiers, CAL cables, connectors and spare fuses. Part numbers are given for the controller, amplifiers, CAL cables, servo amplifier, connectors and fuse.
- 4830 Controller Main Unit P/N S347-39536-40.
- DC Amplifier AMP30-DC P/N S347-39503-41 for test force detector; AC Amplifier AMP30-AC P/N S347-39503-42 for stroke detector.
- CAL Cable CBL-CAL-DC P/N S347-39534-40 (DC amplifier) and CBL-CAL-AC P/N S347-39534-41 (AC amplifier).
- Servo Amplifier AMP30-SV P/N S347-39504-40.
- Digital I/O connector DX40-28P P/N S070-02291-32 with hood DX-28-CV1 P/N S070-02291-52; Alarm I/O connector XM2A-0901 P/N S070-02175-04 with hood XM2S-0913 P/N S070-02174-14.
- Spare fuse 215-005, 2 pcs, P/N S072-02034-01.
- Fan cable supplied only for 10 kN, 20 kN and 30 kN models.
Explains the mechanical structure: servo motor, ball screw, piston rod, ball spline and bearing, center trunnion, load cell and displacement gauge. The servo motor rotation is converted to horizontal piston movement by a silent ball screw; a spline shaft serves as detent and bearing, and a differential-transformer stroke sensor is built in.
- Servo motor receives control signal from Servo Amplifier Unit and outputs torque.
- Ball screw nut connected to piston converts rotation to horizontal test load; silent ball screw for smooth motion.
- Piston rod is a spline shaft with detent mechanism; tip has screw with positioning spigot joint for load cell and jig.
- Ball spline bearing at the head supports load reaction force and transverse load.
- Differential transformer type stroke sensor built into actuator for accurate piston position control.
- Center trunnion position in the middle of the actuator.
Describes the Servo Amplifier Unit configurations for the 5 kN model versus the 10 kN, 20 kN and 30 kN models. The unit contains the servo amplifier SA1 and, for larger models, fan cooling.
- Separate Servo Amplifier Unit designs for 5 kN and for 10/20/30 kN models.
- Servo amplifier SA1 inside the Servo Amplifier Unit displays alarm and status codes.
- Fan cable is used only on 10 kN, 20 kN and 30 kN models.
Contains the technical specifications, external dimensions and amplitude characteristics of the NJ-SERVO series. These data define the rated test force, stroke and performance limits used for test condition planning.
- Specifications table for the NJ-SERVO series.
- External dimension drawings for installation planning.
- Amplitude characteristics diagrams define usable stroke/frequency limits.
- Test conditions must remain within the characteristics diagram and system specifications.
Covers operating environment, transport, fastening the actuator, bolts used, installation cautions, attachment of load cell/jig, optional swivel head and trunnion bracket, power supply capacity per model and instrument wiring between units. Installation, adjustment and re-installation must be performed by a Shimadzu field engineer.
- Operating environment requirements and transport procedure.
- Fastening the instrument with prescribed bolts and prescribed torque.
- Load cell and jig attachment; optional swivel head and trunnion bracket.
- Power supply capacity specified separately for NJ-5kNV, NJ-10kNV, NJ-20kNV and NJ-30kNV.
- Wiring connections between units, test force detector cable, displacement gauge cable, motor power cable, motor encoder cable and controller-to-servo-amplifier cable.
- ROM and software information.
Describes startup and stop operating procedures for the actuator system. Operation is performed through the 4830 controller and Windows software.
- Startup procedure.
- Stop procedure.
Provides operating cautions for conditioning operations, tests with micro displacements, and continuous operation at high speeds and high accelerations. Conditioning operations require a slow full stroke of the piston before and after use to ensure lubrication.
- Perform conditioning operations (slow full piston stroke) before and after use.
- Tests with micro displacements require special caution.
- Continuous high-speed and high-acceleration operation raises ball screw and regeneration resistance temperatures and must be limited.
Lists symptoms, causes and corrective measures for controller/servo amplifier power-on failure, servo amplifier malfunction, alarms, abnormal measured values, loss of control, distorted test waveform, absence of test force and controller malfunction. Before corrective measures, the system as a whole must be turned OFF.
- Turn OFF the system as a whole before taking corrective measures; otherwise the piston may move suddenly.
- If a problem is not described, contact a field engineer.
- Read the instruction manuals for both the controller and the Servo Amplifier Unit.
- Alarm cancellation may require turning the emergency stop switch to the right.
- Heating alarms cannot be cancelled until temperature drops.
- Controller restart procedure after program freeze: emergency stop, main power OFF, front power OFF, cancel emergency stop, main power ON, front power ON.
Describes alarm handling flow, regeneration resistance and ball screw heating alarms with temperature limits and guidelines, and servo amplifier alarms inside the Servo Amplifier Unit. Alarm messages are displayed on the controller screen; depending on the alarm the power unit or waveform stops.
- Alarm message displayed on controller screen; power unit or waveform stops depending on alarm details.
- External alarm or transducer alarm can be configured to stop power supply to actuator or stop waveform.
- Ball screw upper temperature limit 80 °C; regeneration resistance upper limit 100 °C.
- When temperature limit is reached, actuator issues temperature/heating alarm and stops automatically; piston can only be moved by jog dial until temperature drops.
- 'Over Temp' blinks at bottom right of controller screen; 'Heating ALM' displayed at top left in software window.
- Servo amplifier alarm codes and status codes are displayed on servo amplifier SA1; record them and contact Shimadzu representative or field engineer.
Covers maintenance, grease application, replacement guides for consumables, maintenance parts and consumables, testing machine inspections (routine maintenance, whole machine, actuator, overall system), handling bolts and storage. Daily inspection and periodic inspection/calibration are required to maintain performance and accurate measurement data.
- Use grease specified by Shimadzu.
- Perform slow full stroke of piston via conditioning operations before and after use for lubrication.
- Replacement cycles for periodic replacement parts are rough estimates and may be shorter depending on environment and frequency.
- Periodic inspection/calibration should be requested from a Shimadzu representative.
- Inspection scope includes routine maintenance, testing machine as a whole, actuator and overall system.
- Storage requires attention to ambient temperature, humidity, dust and rust.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Ball screw upper temperature limit is 80 °C and regeneration resistance upper limit is 100 °C; reaching these temperatures triggers a temperature/heating alarm and automatic stop. p.68 ↗ | Test conditions must be planned so that continuous high-speed/high-acceleration operation does not exceed these temperatures; after alarm, wait for cooling before restarting. | NJ-SERVO series |
| For NJ10kN, test force -7.5 kN ± 2.5 kN with stroke control 2 Hz ±40 mm gives approximately 1 hour before alarm at ambient approximately 20 °C; test force -5 kN ± 5 kN with 0.8 Hz ±100 mm issues no alarm at approximately 20 °C. p.68 ↗ | These guidelines help select sustainable amplitude/frequency combinations for endurance tests on NJ10kN. | NJ10kN |
| For NJ30kN, test force -15 kN ± 15 kN with stroke control 3 Hz ±16 mm gives approximately 1 hour before alarm at approximately 20 °C; test force -7.5 kN ± 7.5 kN with 4.2 Hz ±8 mm issues no alarm at approximately 20 °C. p.68 ↗ | These guidelines help select sustainable amplitude/frequency combinations for endurance tests on NJ30kN. | NJ30kN |
| High voltage may remain in the servo amplifier after power OFF; do not touch power supply terminals until power has been OFF for at least five minutes, and confirm the CHARGE lamp is unlit. p.9 ↗ | Maintenance and wiring inspection must wait at least five minutes after power removal and until the CHARGE lamp extinguishes. | NJ-SERVO series with Servo Amplifier Unit |
| When attaching or removing the product, the main breaker for the power source must be turned OFF. p.7 ↗ | Prevents dangerous piston movement during attachment/removal. | NJ-SERVO series |
| When attaching the product, use the prescribed bolts and tighten them with the prescribed torque. p.7 ↗ | Fastening must follow the bolt and torque data in the installation section to ensure safe reaction force transmission. | NJ-SERVO series |
| When hoisting the actuator, use the eye bolts provided for the product; transport with a crane and chain block. p.6 ↗ | Improper hoisting points can cause falling and injury or instrument failure. | NJ-SERVO series |
| Avoid applying excessive eccentric load or bending load to the piston in excess of permitted values. p.8 ↗ | Load cell and jig alignment must be checked to stay within permitted eccentric/bending limits. | NJ-SERVO series |
| The actuator is a partly completed machine and cannot be CE marked alone; the user must evaluate the final equipment against EC directives. p.18 ↗ | System integrators must perform their own conformity assessment and add protective measures (fence, interlock) for residual risks. | NJ-SERVO series |
| EMC immunity tested to IEC 61000-4-2 air 8 kV / contact 4 kV, IEC 61000-4-3 10 V/m and 3 V/m, IEC 61000-4-4 ±2.4 kV to AC power line and ground, IEC 61000-4-5 ±1 kV line-to-line and ±2 kV line-to-ground, IEC 61000-4-6 10 V. p.14 ↗ | Install away from strong electromagnetic noise sources and use a separate power source for such devices. | NJ-SERVO series |
| The product is Class A equipment (EN55011:2009+A1:2010 Group 1) and is not designed for residential environments. p.14 ↗ | In residential or nearby environments, maintain appropriate distance from disturbed devices. | NJ-SERVO series |
| Replacement parts are available for seven years after product discontinuation, but may cease earlier depending on supplier conditions and required quantity. p.19 ↗ | Long-term test programs should plan spare parts procurement within this window. | NJ-SERVO series |
| Fan cable is supplied only for 10 kN, 20 kN and 30 kN models. p.28 ↗ | 5 kN model does not use the fan cable; wiring and spare parts lists differ by model. | NJ-10kNV, NJ-20kNV, NJ-30kNV |
| The 4830 controller uses DC Amplifier AMP30-DC for the test force detector and AC Amplifier AMP30-AC for the stroke detector, with matching CAL cables CBL-CAL-DC and CBL-CAL-AC. p.29 ↗ | CAL cable must be compatible with the sensor; incompatible CAL cable causes abnormal measured values. | NJ-SERVO system with 4830 controller |
| Before restarting a test after a heating alarm, revise the test conditions and allow the temperature to drop sufficiently; the piston can only be moved by jog dial until then. p.68 ↗ | Do not attempt normal test restart while 'Over Temp' is blinking. | NJ-SERVO series |
| When the servo is turned ON again after an alarm, protect the sample and ensure safety in the vicinity of the actuator main unit. p.67 ↗ | Restart can cause sudden piston movement; personnel must be clear of the test space. | NJ-SERVO series |
| Use only grease specified by Shimadzu and confirm greasing before use; perform a slow full stroke of the piston via conditioning operations before and after use. p.9 ↗ | Correct lubrication prevents wear and maintains measurement accuracy. | NJ-SERVO series |
| Before taking corrective measures when a problem occurs, turn OFF the system as a whole. p.63 ↗ | Prevents sudden piston movement and injury during troubleshooting. | NJ-SERVO series |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| The controller or Servo Amplifier Unit will not turn ON. p.63 ↗ | The power cable to the controller is not connected or is coming loose. | Connect the power cable properly. | Section 7.1 |
| The controller or Servo Amplifier Unit will not turn ON. p.63 ↗ | The power switch on the back of the controller is not ON. | Turn ON the power switch. | Section 7.1 |
| The controller or Servo Amplifier Unit will not turn ON. p.63 ↗ | The fuse in the controller has blown. | Replace the fuse in the controller. For the replacement procedures, refer to the instruction manual for the controller being used. | Section 7.1 |
| The Servo Amplifier Unit is not operating properly. p.63 ↗ | The Servo Amplifier Unit cables are not connected properly or are coming loose. | Connect the servo amplifier cables securely. | Section 7.1 |
| The Servo Amplifier Unit is not operating properly. p.63 ↗ | The breaker for the Servo Amplifier Unit is not ON. | Turn ON the breaker for the Servo Amplifier Unit. | Section 7.1 |
| The Servo Amplifier Unit is not operating properly. p.63 ↗ | An alarm has been issued. | Remedy the cause of the alarm, and then cancel the alarm. | Section 7.1 |
| The alarm cannot be canceled. p.64 ↗ | Emergency stop switch pressed as is. | With the emergency stop switch pressed as is, turn the switch to the right to cancel the alarm. | Section 7.1 |
| The alarm cannot be canceled. p.64 ↗ | Regeneration resistance heating alarm or ball screw heating alarm has been issued; alarm cannot be canceled until temperature drops. | Cancel the alarm after the temperature drops. Before restarting the test, revise the test conditions. | Section 7.1; see 7.2 |
| The alarm cannot be canceled. p.64 ↗ | Alarm issued by the servo amplifier inside the Servo Amplifier Unit. | Check details in Servo Amplifier Alarm Inside the Servo Amplifier Unit in section 7.2, and then contact a field engineer. | Section 7.1 |
| A system memory error is displayed. p.64 ↗ | System memory error. | Contact a field engineer. | Section 7.1 |
| An alarm is issued. p.64 ↗ | The motor is overloaded. | Revise the test conditions, and use the instrument within its rated limits. | Section 7.1 |
| An alarm is issued. p.64 ↗ | The emergency stop switch has been pressed. | Reset the emergency stop switch. | Section 7.1 |
| An alarm is issued. p.64 ↗ | A servo amplifier alarm in the Servo Amplifier Unit has been issued. | Check if there is a problem with the wiring. If you cannot remedy the problem, contact a field engineer. | Section 7.1 |
| An alarm is issued. p.64 ↗ | The 4830 controller has issued an alarm. | Refer to the instruction manual for the 4830 controller. | Section 7.1 |
| The measured values are abnormal. p.64 ↗ | The sensor cable is not connected or is coming loose. | Connect the sensor cable securely. | Section 7.1 |
| The measured values are abnormal. p.64 ↗ | The CAL cable is incompatible with the sensor. | Connect the sensor to a compatible CAL cable and then restart the system. | Section 7.1 |
| The instrument cannot be controlled. p.65 ↗ | The sensor cable is not connected or is coming loose. | Connect the sensor cable securely. | Section 7.1 |
| The instrument cannot be controlled. p.65 ↗ | The sensor polarity is reversed. | Adjust the sensor polarity. | Section 7.1 |
| The instrument cannot be controlled. p.65 ↗ | The servo amplifier cable has come loose or is coming loose. | Connect the servo amplifier cables securely. | Section 7.1 |
| The instrument cannot be controlled. p.65 ↗ | The test conditions exceed the system specifications. | Revise the test conditions. | Section 7.1 |
| The instrument cannot be controlled. p.65 ↗ | The ball screw, thrust bearing, or other parts are damaged. | Contact a field engineer. | Section 7.1 |
| The test waveform is distorted. p.65 ↗ | The gain value is inappropriate. | Perform tuning. | Section 7.1 |
| The test waveform is distorted. p.65 ↗ | The brackets or test jigs are not attached securely and are rattling. | Attach them securely so that they do not rattle. | Section 7.1 |
| The test waveform is distorted. p.65 ↗ | The sample is not attached securely and is rattling. | Attach it securely so that it does not rattle. | Section 7.1 |
| The test waveform is distorted. p.65 ↗ | The test conditions exceed the system specifications. | Revise the test conditions. Check the characteristics diagram, and use the instrument within its specified limits. | Section 7.1 |
| The test waveform is distorted. p.65 ↗ | The grounding is inappropriate. | Connect the grounding wire appropriately. | Section 7.1 |
| The test waveform is distorted. p.65 ↗ | The ball screw, thrust bearing, or other parts are damaged. | Contact a field engineer. | Section 7.1 |
| No test force is applied. p.65 ↗ | The test conditions exceed the system specifications. | Revise the test conditions. | Section 7.1 |
| No test force is applied. p.65 ↗ | The ball screw, thrust bearing, or other parts are damaged. | Contact a field engineer. | Section 7.1 |
| The controller is not operating properly. p.66 ↗ | The operation lock is active. | Cancel the operation lock. | Section 7.1 |
| The controller is not operating properly. p.66 ↗ | The program has frozen. | Restart the controller: (1) Press the emergency stop switch, and then confirm that the testing machine is stopped. (2) Turn OFF the main power switch on the back of the controller, and then turn OFF the front power switch. (3) Cancel the emergency stop, turn ON the main power switch, and then turn ON the front power switch. | Section 7.1 |
| Over Temp p.68 ↗ | Ball screw or regeneration resistance temperature reached upper limit (80 °C ball screw, 100 °C regeneration resistance) due to continuous high-speed/high-acceleration operation. | Revise the test conditions and allow temperature to drop sufficiently before restarting; lower frequency or operate intermittently. | Section 7.2; displayed blinking at bottom right of controller screen |
| Heating ALM p.68 ↗ | Ball screw or regeneration resistance heating alarm due to continuous high-speed/high-acceleration operation. | Revise the test conditions and allow temperature to drop sufficiently before restarting; lower frequency or operate intermittently. | Section 7.2; displayed at top left in software window |