Near Infrared Photometric Accuracy of Agilent Cary 7000 UV-Vis-NIR Spectrophotometer
Technical Reference Summary
Covered Models & Document Scope
Technical overview evaluating photometric (absorbance) accuracy in the near infrared (NIR) region of the Agilent Cary 7000 universal measurement spectrophotometer (UMS) using three metal-on-quartz filter reference materials; no model/size variants or ordering codes are covered.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Absorbance accuracy requirement below 1 Abs | ±0.01 % p.2 ↗ | USP/Ph. Eur. guideline |
| Absorbance accuracy requirement above 1 Abs | ±1 % p.2 ↗ | USP/Ph. Eur. guideline |
| Certified absorbance value of filter | 0.3688 Abs p.3 ↗ | 1,100 nm |
| Certified absorbance tolerance of filter | ±0.0025 Abs p.3 ↗ | 1,100 nm |
| Measured absorbance (Cary 7000) | 0.3694 Abs p.3 ↗ | 1,100 nm, n = 3 |
| Absorbance error | 0.1894 % p.3 ↗ | 1,100 nm |
Operating Limits
- Absorbance accuracy requirement below 1 Abs: ±0.01% p.2 ↗
- Absorbance accuracy requirement above 1 Abs: ±1% p.2 ↗