Induztr
Engineering Digest Document Viewer Related Models

Protein Sizing and Quantification with the Agilent ProteoAnalyzer System

Technical Reference Summary

Covered Models & Document Scope

Agilent ProteoAnalyzer system with the Agilent Protein Broad Range P240 kit; technical overview covering protein sizing and quantification (BSA and CAII examples), reduced samples, 10–240 kDa sizing range, 2–2,000 ng/µL concentration range.

Technical Data

Parameter Value Condition
Samples analyzed simultaneously per run up to 12 samples p.1 ↗
Run time approximately 30 min p.1 ↗
Sizing range 10 to 240 kDa p.1 ↗
Detection concentration range 2 to 2,000 ng/µL p.1 ↗
Typical sizing accuracy (LM only) < 15 % p.3 ↗ BSA, CAII
Typical sizing accuracy (LM and UM) < 10 % p.3 ↗ BSA, CAII
Average size, BSA 67.99 to 69.84 kDa p.4 ↗ LM only, Instruments 1–3
Average size, CAII 27.17 to 27.87 kDa p.5 ↗ LM only, 2 to 2,000 ng/µL
Lower marker size 6 kDa p.2 ↗
Upper marker precision benefit higher precision for samples smaller than 200 kDa p.2 ↗ UM added
Sample buffer 1x PBS (30 mM Tris-HCl, 26 mM NaH2PO4, 41 mM Na2HPO4, 79 mM NaCl, pH 8.5) p.2 ↗
CAII extinction coefficient 0.575 p.2 ↗ Nanodrop

Operating Limits

  • Sizing range: 10 to 240 kDa p.1 ↗
  • Detection concentration range: 2 to 2,000 ng/µL p.1 ↗
  • Maximum samples per run: 12 samples p.1 ↗
  • Run time: approximately 30 min p.1 ↗