Agilent 1260 Infinity III Vialsampler with 100 µL analytical head and 100 µL loop – Injection volume: 1 µL – Needle wash: 3 s with acetonitrile: 0.1 M HCl (50:50; v:v) – Sampling speeds: draw speed: 200 µL/min; eject speed: 400 µL/min – Needle height position: offset 0.0 mm – Sample temperature: 8 °C Agilent 1260 Infinity III Multisampler with 100 µL analytical head and 100 µL loop – Injection volume: 1 µL – Needle wash: 3 s with acetonitrile: 0.1 M HCl (50:50; v:v) – Sampling speeds: draw speed: 200 µL/min; eject speed: 400 µL/min – Needle height position: use vial bottom sensing, offset 0.0 mm – Sample temperature: 8 °C
Table 3. Calibration linearity, sensitivity, and repeatability of amino acid analysis using injector workflows on an Agilent 1260 Infinity III Multisampler. Ten consecutive injections of the amino acid mixture with a final concentration of 22.5 pmol/µL for each amino acid and 50 pmol/µL ISTDs were used for relative standard deviation (RSD) calculations. The limit of detection (LOD) was determined by using a threshold of 3x the signal‑to‑noise (S/N) value.
Peak No.
Compound
RT (min)
RT RSD (%)
Area RSD (%)
LOD (pmol/µL)
Calibration Range (pmol/µL)
Calibration Type
R²
1
L-Aspartic acid
1.07
0.61
0.34
0.227
0.45 to 90
Linear
0.99999
2
L-Glutamic acid
1.56
0.60
0.36
0.117
0.45 to 90
Linear
0.99997
3
L-Asparagine
3.47
0.07
0.25
0.112
0.45 to 90
Linear
0.99990
4
L-Serine
3.68
0.06
0.29
0.068
0.45 to 90
Linear
0.99992
5
L-Glutamine
4.20
0.04
0.23
0.121
0.45 to 90
Linear
0.99986
6
L-Histidine
4.38
0.03
0.98
0.143
0.45 to 90
Quadratic
0.99999
7
Glycine
4.62
0.03
0.32
0.090
0.45 to 90
Linear
0.99997
8
L-Threonine
4.77
0.03
0.20
0.122
0.45 to 90
Linear
0.99997
9
L-Arginine
5.38
0.02
0.21
0.107
0.45 to 90
Linear
0.99998
10
L-Alanine
5.72
0.02
0.21
0.110
0.45 to 90
Linear
0.99998
11
L-Tyrosine
6.69
0.02
0.30
0.110
0.45 to 90
Linear
0.99998
12
L-Cysteine
7.76
0.04
1.71
1.399
2.25 to 90
Quadratic
0.99994
13
L-Valine
8.11
0.03
0.23
0.109
0.45 to 90
Linear
0.99997
14
L-Methionine
8.28
0.03
0.91
0.142
0.45 to 90
Linear
0.99998
15
L-Norvaline*
8.54
0.03
0.36
NA
NA
NA
NA
16
L-Tryptophan
8.96
0.03
0.49
0.133
0.45 to 90
Linear
0.99996
17
L-Phenylalanine
9.24
0.02
0.21
0.126
0.45 to 90
Linear
0.99998
18
L-Isoleucine
9.39
0.02
0.33
0.116
0.45 to 90
Linear
0.99999
19
L-Leucine
9.90
0.02
0.34
0.121
0.45 to 90
Linear
0.99999
20
L-Lysine
10.30
0.02
2.32
0.39
0.45 to 90
Quadratic
0.99993
21
Sarcosine*
12.35
0.02
3.29
NA
NA
NA
NA
22
L-Proline
12.84
0.02
3.18
1.606
2.25 to 90
Quadratic
0.99943
* ISTD
Table 4. Calibration linearity, sensitivity, and repeatability of amino acid analysis using injector workflows on an Agilent 1260 Infinity III Vialsampler. Ten consecutive injections of the amino acid mixture with a final concentration of 22.5 pmol/µL for each amino acid and 50 pmol/µL ISTDs were used for RSD calculations. The LOD was determined by using a threshold of 3x the S/N value.
Peak No.
Compound
RT (min)
RT RSD (%)
Area RSD (%)
LOD (pmol/µL)
Calibration Range (pmol/µL)
Calibration Type
R²
1
L-Aspartic acid
1.194
0.338
0.469
0.167
0.45 to 90
Linear
0.99994
2
L-Glutamic acid
1.808
0.456
0.632
0.267
0.45 to 90
Linear
0.99995
3
L-Asparagine
3.657
0.139
0.512
0.137
0.45 to 90
Linear
0.99980
4
L-Serine
3.875
0.110
0.563
0.137
0.45 to 90
Linear
0.99997
5
L-Glutamine
4.389
0,081
0.623
0.141
0.45 to 90
Linear
0.99967
6
L-Histidine
4.573
0.079
1.313
0.254
0.45 to 90
Quadratic
0.99972
7
Glycine
4.829
0.079
0.559
0.166
0.45 to 90
Linear
0.99996
8
L-Threonine
4.966
0.075
0.541
0.213
0.45 to 90
Linear
0.99984
9
L-Arginine
5.569
0.070
0.588
0.184
0.45 to 90
Linear
0.99998
10
L-Alanine
5.933
0.050
0.639
0.186
0.45 to 90
Linear
0.99997
11
L-Tyrosine
6.912
0.078
0.643
0.163
0.45 to 90
Linear
0.99998
12
L-Cysteine
7.991
0.048
1.181
0.886
0.90 to 90
Quadratic
0.99973
13
L-Valine
8.326
0.044
0.676
0.149
0.45 to 90
Linear
0.99998
14
L-Methionine
8.517
0.037
2.892
0.563
0.90 to 90
Linear
0.99962
15
L-Norvaline*
8.764
0.034
0.573
NA
NA
NA
NA
16
L-Tryptophan
9.209
0.026
0.653
0.156
0.45 to 90
Linear
0.99981
17
L-Phenylalanine
9.479
0.025
0.675
0.187
0.45 to 90
Linear
0.99999
18
L-Isoleucine
9.615
0.025
0.525
0.169
0.45 to 90
Linear
0.99997
19
L-Leucine
10.137
0.017
0.581
0.189
0.45 to 90
Linear
0.99997
20
L-Lysine
10.535
0.012
1.287
0.564
0.45 to 90
Quadratic
0.99999
21
Sarcosine*
12.578
0.016
3.521
NA
NA
NA
NA
22
L-Proline
13.067
0.013
2.961
0.477
0.90 to 90
Quadratic
0.99965
* ISTD
Table 5. Caffeine calibrations using calibration solutions prepared with the Agilent 1260 Infinity III Multisampler and Agilent 1260 Infinity III Vialsampler.
Autosampler
Calibration Function
Correlation R2
Deviation Control Sample (%)
Agilent 1260 Infinity II Multisampler
y = 28.3784x – 0.4930
R² = 0.99994
3.58%
Agilent 1260 Infinity II Vialsampler
y = 28.5657x – 0.7107
R² = 0.99999
0.79%
Engineering Facts
13 facts · every claim linked to a page
Selection
The autosampler default configuration uses a 100 µL analytical head and 100 µL sample loop, so method volumes must fit this setup.
Determines whether the standard injector hardware can deliver the required sample and reagent volumes without modification.
Source
“100 µL analytical head and 100 µL sample loop (default setup)”
Selection
The Vialsampler requires the integrated column compartment while the Multisampler requires an additional Multicolumn Thermostat, affecting system configuration.
Impacts the module list and bench space needed when choosing between the two autosamplers.
Source
“Multisampler (G7167A) with 100 µL analytical head and 100 µL sample loop (default setup) and additional Multicolumn Thermostat (G7116A)”
Compliance
The injector workflow software requires OpenLab CDS version 2.8 or later.
Must be verified before purchase/installation to ensure the automation workflows are supported.
Source
“Agilent OpenLab CDS version 2.8 or later”
Operation
Reagents in the autosampler must be replaced daily and stored at 4 to 8 °C.
Defines the daily operating routine and sample compartment temperature control needed for reliable derivatization.
Source
“All reagents should be stored at 4 to 8 °C, and reagents in the autosampler should be exchanged daily.”
Maintenance
OPA/FMOC reagents must be aliquoted into amber vials and used within one week after opening.
Sets the reagent shelf-life limit that governs consumable planning and workflow scheduling.
Source
“stored for no longer than a week”
Operation
Stock solutions should be aliquoted before freezing to avoid repeated freeze-thaw cycles.
Prevents degradation of standards that would compromise calibration accuracy.
Source
“It is recommended to distribute the stock solutions before freezing them at –20 °C to avoid freeze-thaw cycles.”
Operation
Flat glass vial inserts are required to ensure proper mixing during the dilution series workflow.
Specifies the correct vial insert type needed for reliable automated serial dilutions.
Source
“the use of flat glass vial inserts (part number 5181-3377) is necessary to ensure proper mixing of the dilution.”
Operation
The Multisampler uses vial bottom sensing with a 0.0 mm offset for amino acid analysis, while the caffeine method uses a –3.0 mm offset.
Needle height settings must be configured per method to ensure correct sample aspiration.
Source
“Needle height position: use vial bottom sensing, offset 0.0 mm”
Performance
The amino acid method separates 20 amino acids and two ISTDs within 17 minutes.
Defines the achievable throughput and run time for the derivatization workflow.
Source
“The chromatogram shows a successful separation of 20 amino acids and two ISTDs within 17 minutes.”
Performance
Most compounds achieved RT RSD below 0.1% and peak area RSD below 1% over ten injections.
Demonstrates the precision achievable with injector workflows for method validation planning.
Source
“showing values below 0.1% for RT RSD and below 1% for peak area RSD”
Performance
LOD was mostly below 0.4 pmol/µL for all amino acids except L-cysteine and L-proline.
Sets realistic sensitivity expectations and flags analytes needing special attention.
Source
“excellent sensitivity with LOD mostly below 0.4 pmol/µL was obtained for all amino acids except L-cysteine and L-proline”
Performance
The caffeine dilution workflow covers a serial dilution range of 25 to 0.04 µg/mL.
Defines the usable calibration range for the automated dilution series.
Source
“a serial dilution of the caffeine stock solution within a range of 25 to 0.04 µg/mL”
Performance
The Vialsampler gave a 0.79% deviation on the control sample versus 3.58% for the Multisampler in caffeine calibration.
Helps select the autosampler based on dilution accuracy requirements.