ACQUITY Premier System with 2D Technology Capabilities Guide

Technical Reference Summary

Covered Models & Document Scope

This document covers the Waters ACQUITY Premier System with 2D Technology, including configurations with up to three ACQUITY Premier solvent managers (binary or quaternary), up to three 515 pumps or one 1525 pump, a sample manager, a column manager (CM-A) with two 6-port 2-position valves and two independent heating zones, optional detectors, and up to three external valves. It describes system configuration in Empower 3 and MassLynx software, CM-A setup, external valve connection, flow path configurations, spare parts, and worksheets.

Section-by-Section Technical Breakdown

pp. 12
System Overview

The ACQUITY Premier System with 2D Technology employs solvent managers, a sample manager, a column manager, optical detectors, and mass spectrometers for multidimensional analyses. A 2D system can include up to three ACQUITY Premier solvent managers and up to three 515 pumps (or one 1525 pump). The CM-A is a required part of all 2D systems and incorporates programmable switching valves and two independent heating zones with active preheating.

  • Up to three ACQUITY Premier solvent managers and up to three 515 pumps (or one 1525 pump) can be configured.
  • The column manager (CM-A) is required for all 2D systems.
  • CM-A includes programmable switching valves and two independent heating zones with active preheating.
  • A conventional ACQUITY Premier system can be upgraded to include 2D technology.
pp. 13-19
System Configuration

System operation must be configured under Empower or MassLynx software. The CM-A must be operated in Advanced mode to enable 2D features. When adding or replacing an instrument component, the system configuration procedure must be repeated. The section details configuration steps for Empower 3, MassLynx, Waters Pump Control software, CM-A, and external valves.

  • Empower 3 with Driver Pack 4 is required for 2D technology under Empower control.
  • MassLynx with Driver Pack 3 or 4 is required for MassLynx-controlled 2D technology.
  • The CM-A must be configured in the console software regardless of whether MassLynx or Empower is used.
  • External valves connect to the CM-A via adapter cable 441000459 on the COM B connector.
  • Each external valve must have a unique address setting (1, 2, or 3).
pp. 20-25
System Components and Features

This section describes the components of the ACQUITY Premier system with 2D technology, including the column manager (CM-A), solvent managers, additional pumps, external valves, and tubing. The CM-A includes two 6-port, 2-position valves and two column troughs with independent temperature control. Typical configurations include two binary solvent managers or one quaternary and one binary solvent manager.

  • CM-A includes two 6-port, 2-position valves and two column troughs with independent temperature control.
  • A maximum of two active preheaters can be used when running 2D applications in the CM-A.
  • Column heaters (CM-Aux) cannot be used with 2D systems.
  • Up to three ACQUITY Premier solvent managers can be used; a quaternary solvent manager must be configured as the first-dimension pump (Pump 1).
  • Do not use a quaternary solvent manager for the second dimension.
  • Additional pumps (515 or 1525) can achieve flow rates up to 10 mL/min at pressures up to 34,473 kPa (344 bar, 5000 psi).
  • External valves are available in titanium (10 ports, 2 positions), PEEK (7 ports, 6 positions), and stainless steel (11 ports, 10 positions), all rated to 34,473 kPa (344 bar, 5000 psi).
  • Tubing options include High Performance Surfaces (HPS) or MP35N; use MP35N for corrosion resistance in high-ionic strength aqueous conditions.
pp. 26-40
Flow Path Configurations

This section provides examples of system configurations and their applications, including multiple-dimension 1-1, single-dimension 0-1 and 0-2, entry-level and advanced trap-and-elute, 1D/2D rapid switch, single heart-cut, and comprehensive 2D. The configurations support various workflows such as increasing injection volume up to 1000 uL, dual-stream loading, and heart-cutting.

  • Multiple-dimension 1-1 (single-stream loading) allows injection volumes up to 1000 uL without volume or mass overload.
  • Single-dimension 0-1 uses a single elution pump and a regeneration pump operating in isocratic mode only.
  • Single-dimension 0-2 uses dual elution pumps for two different analytical columns.
  • Entry-level trap-and-elute 1-1 (dual-stream loading) dilutes the sample to less than 5% before reaching the trap column.
  • Advanced trap-and-elute 2-1 and 1-2 configurations allow up to eight methods to be selected.
  • 1D/2D rapid switch 1-1 at-column dilution supports switching between 1D and 2D modes using a valve switch.
  • Single heart-cut 1-1-1 at-column dilution uses three pumps and a loop/trap combination.
  • Comprehensive 2D Premier uses high-pressure switching valves and sampling loops for on-line fraction collection.
pp. 41
Spare Parts

This section lists spare parts for the ACQUITY Premier System with 2D Technology, including tubing assemblies and a tee fitting with their engineering codes, sizes, materials, and order numbers.

  • Tubing assemblies are available in MP35N, HPS, ETFE, and PEEK materials.
  • Order numbers range from 700013666 to 700013679.
  • A tee fitting with .010 ID in HPS material is listed with order number 700013679.
pp. 42-51
Worksheets

This section contains printable blank worksheets and examples of completed worksheets to record CM-A setup, valve identity, and valve position. The worksheets include 2D programming tables and valve times, actions, connections, and positions.

  • Blank worksheets include a 2D programming worksheet and tables for two-valve and one-valve times, actions, connections, and positions.
  • Examples of completed worksheets show a 2D programming worksheet and valve times, actions, connections, and positions.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
A 2D system can include as many as three ACQUITY Premier solvent managers and as many as three 515 pumps (or one 1525 pump). p.12 ↗ System design must account for the number and type of pumps to support the desired flow path configuration. Whole series
The CM-A includes two 6-port, 2-position valves and two column troughs with independent temperature control. p.21 ↗ The CM-A provides the switching and temperature control necessary for multidimensional separations. Whole series
A maximum of two active preheaters can be used when running 2D applications in the CM-A. p.21 ↗ Preheater configuration is limited to two active preheaters for 2D operation. Whole series
Column heaters (CM-Aux) cannot be used with 2D systems. p.21 ↗ Do not attempt to use CM-Aux column heaters in a 2D configuration. Whole series
A quaternary solvent manager must be configured as the first-dimension pump (Pump 1) in Waters Pump Control software. p.24 ↗ When using a quaternary solvent manager, it must be assigned to Pump 1 to avoid retention time reproducibility issues. Systems with quaternary solvent manager
Do not use a quaternary solvent manager for the second dimension of an analysis. p.24 ↗ Use a binary solvent manager for the second dimension to avoid problems affecting retention time reproducibility. Whole series
Additional pumps (515 or 1525) can achieve flow rates as high as 10 mL/min at pressures up to 34,473 kPa (344 bar, 5000 psi). p.25 ↗ Ensure that system components downstream of additional pumps are rated for these pressures and flow rates. Systems with additional pumps
External valves are available with pressure ratings of 34,473 kPa (344 bar, 5000 psi) for titanium (10 ports, 2 positions), PEEK (7 ports, 6 positions), and stainless steel (11 ports, 10 positions). p.25 ↗ Select external valve material and port configuration based on application pressure and chemical compatibility. Systems with external valves
Each external valve must have a unique address setting (1, 2, or 3) that corresponds to the external valve identification (1, 2, or 3). p.19 ↗ Set the address dial on each valve correctly to ensure proper communication with the CM-A. Systems with external valves
The adapter cable (441000459) connects to the connector labeled "COM B" on the rear panel of the CM-A. p.18 ↗ Use the specified adapter cable for connecting external valves to the CM-A. Systems with external valves
To use 2D technology under Empower control, you must use Empower 3 software together with Driver Pack 4. p.13 ↗ Ensure Empower 3 and Driver Pack 4 are installed for 2D operation. Empower-controlled systems
MassLynx-controlled 2D technology requires MassLynx software with Driver Pack 3 or Driver Pack 4. p.14 ↗ Verify MassLynx and Driver Pack versions before configuring 2D technology. MassLynx-controlled systems
The CM-A must be configured with "2 columns" selected and Advanced mode with 2 for left and right positions and 6 for left and right ports. p.16 ↗ Follow the specified configuration to enable 2D features; do not select more than two columns. Whole series
You can install only one preheater in each trough, on either the left-hand or right-hand side. p.18 ↗ Plan preheater installation accordingly; select Active or Passive for column positions 1 and 2. Whole series
The trap-and-elute single-stream configuration allows injection volumes up to 1000 uL without volume or mass overload. p.26 ↗ Use this configuration to increase detection limits for applications requiring large injection volumes. Multiple-dimension 1-1 configuration
In dual-stream loading, the entire sample is diluted to less than 5% before reaching the trap column. p.31 ↗ Adjust the relative flow rates of the loader and diluter streams to vary the dilution ratio. Entry-level trap-and-elute 1-1 configuration
The regeneration pump in the single-dimension 0-1 configuration operates in isocratic mode only. p.28 ↗ Both pumps are restricted to their assigned circuits; no crossover option is available. Single-dimension 0-1 configuration
For HPLC (analytical column with 3.9 or 4.6 mm ID), a QSM pump is used; for UPLC (analytical column with 1.0 or 2.1 mm ID), a BSM pump is more suitable. p.26 ↗ Select pump type based on column dimensions and application requirements. Whole series
Use MP35N tubing for corrosion resistance in high-ionic strength aqueous conditions, such as those needed for separating biological samples. p.25 ↗ Choose MP35N tubing assemblies for biological or high-ionic strength applications. Whole series
When configuring the system for Comprehensive 2D Premier using the repeat feature, select the 2D repeat function above the Binary Solvent Manager gradient table and above the Column Manager event table. p.40 ↗ The time displayed on the last line of both tables defines the duration of one cycle, which repeats throughout a sample's run time. Comprehensive 2D Premier configuration