Technical Reference Summary

Covered Models & Document Scope

This document (715003723 Rev. B, December 2, 2013) covers the Waters Micro Solvent Stabilizer Kit, an accessory that modifies a Waters system containing a mass spectrometer to sparge or blanket solvents with helium. It applies to systems using one or two solvent bottles up to 1000 mL with GL45 DIN thread (ISO 4796). No model-specific variants or ordering codes beyond the kit part number are given.

Section-by-Section Technical Breakdown

pp. 2
Required Materials and Tools

Lists the helium source, solvent bottles, regulator assembly, and tools needed before installation. The helium supply must be highest purity with a regulated output of 621–689 kPa (6.2–6.9 bar, 90–100 psi). Bottles must be GL45 DIN thread per ISO 4796 and no larger than 1 L.

  • Helium source: highest purity, regulated output 621–689 kPa (6.2–6.9 bar, 90–100 psi)
  • One or two solvent bottles up to 1 L maximum, GL45 DIN thread per ISO 4796
  • Regulator assembly and mounting bracket supplied in the installation kit
  • Required tools: razor knife and T20 TORX driver
pp. 2-4
Installing the Regulator Assembly

Describes mounting the regulator assembly on the right-hand side of the solvent tray. The regulator attaches to its mounting bracket with two M4×10 screws, and the bracket is secured to the four bosses on the bottom of the solvent tray with four M4×10 screws.

  • Regulator assembly mounts on the right-hand side of the solvent tray
  • Attach regulator to mounting bracket using T20 TORX driver and two M4×10 screws
  • Locate four bosses on the bottom of the solvent tray
  • Secure bracket to tray bottom with four M4×10 screws
pp. 4-6
Making the Solvent Connections

Covers preparing bottle caps and solvent tubing for 500-mL or 1000-mL bottles. It specifies trimming Teflon lines for 500-mL bottles, installing Omnifit fittings and ferrules, fitting check valves and bubbler filters, relocating solvent filters, and cutting protective netting to bottle height.

  • For 500-mL bottles, trim approximately 10 cm from each of the 3 Teflon tubing lines under the bottle-caps
  • Replacement tubing: 1/8-inch OD × 1/16-inch ID Teflon
  • Install check valve finger-tight on one of the three tubing ports
  • Bubbler filter and check valve must attach to the same port
  • Cut bottle netting to approximately 17 cm for 500-mL and 22 cm for 1000-mL reservoirs
  • Netting is suitable for 500-mL and 1000-mL bottles less than 10 cm in diameter
pp. 6-8
Making the Gas Connections

Details the helium gas path from regulator to bottle caps, including tubing lengths, check-valve orientation, Y-fitting, and regulator fittings. The in-line check valve arrow must point away from the regulator, and regulator connections require PEEK fittings with non-tapered ferrules.

  • Tubing from regulator gas outlet to in-line check valve: 13–15 cm
  • Tubing from in-line check valve to Y-fitting: 5–7 cm
  • Tubing from Y-fitting to each bottle-cap check valve: 46–51 cm
  • In-line check valve uses blue nut and clear tapered ferrule; arrow must point away from regulator
  • Y-fitting uses tapered ferrules and large nuts
  • Regulator connections use PEEK fittings and non-tapered ferrules
  • Kit includes 1/8-inch NPT fitting and 1/8-inch to 1/4-inch bushing for various helium sources
pp. 9-11
Operation

Explains sparging and blanketing procedures, solvent addition, and helium source replacement. Helium source pressure must be 621–689 kPa (6.2–6.9 bar, 90–100 psi), and bottle-cap shutoff valves must be closed when initially supplying helium and when filling bottles or changing the source.

  • All solvent tubing must be free of liquid before operation
  • Sparge: open all bottle-cap shutoff valves
  • Blanket: close shutoff valve at the unused port and keep all other port valves open
  • When adding solvents: stop solvent flow, close all bottle-cap shutoff valves, remove cap, fill, reinstall cap, open valves, sparge, blanket, restart flow
  • Valve handle horizontal = closed; vertical = open
  • When replacing helium source: stop solvent flow, close all bottle-cap shutoff valves, disconnect depleted source, connect new source
pp. 11-12
Troubleshooting

Provides a symptom/cause/solution table for gas consumption, pressure-relief hissing, no gas flow, and inability to sparge. It also states the regulator default setting and pressure-relief behavior.

  • Regulator is set to approximately 31 kPa (0.31 bar, 4.5 psi)
  • Pressure relief valve partially opens above approximately 31 kPa and allows 100% venting at approximately 70 kPa (0.7 bar, 10 psi)
  • Do not tamper with the regulator; overpressure can burst the solvent bottle or separate gas tubing
  • Use a leak detector such as Snoop to find gas leaks
pp. 13
Solvent Stabilization Kit Installed on Solvent Tray

Shows the suggested installation layout with component locations and suggested tubing lengths. It identifies the required cap check valve and in-line check valve and gives lengths for the gas connections.

  • Suggested tubing length from regulator to in-line check valve: 12.7–15.2 cm
  • Suggested tubing length from Y-connector to each bottle cap: 2× 45.7–50.8 cm
  • Suggested tubing length from in-line check valve to Y-connector: 5.1–7.6 cm
  • Cap check valve and in-line check valve are required

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Helium source regulated output must be between 621 and 689 kPa (6.2 and 6.9 bar, 90 and 100 psi). p.2 ↗ Select a helium regulator and supply that can maintain this pressure range during sparging and blanketing. Whole kit
Solvent bottles must be no larger than 1000 mL and have GL45 DIN thread per ISO 4796. p.2 ↗ Verify bottle compatibility before installing the cap assemblies. Whole kit
For 500-mL bottles, trim approximately 10 cm from the end of each of the 3 Teflon tubing lines under the bottle-caps. p.4 ↗ Tubing length must be adjusted to bottle size to avoid excess tubing in smaller bottles. 500-mL bottle configuration
Replacement solvent tubing is 1/8-inch OD × 1/16-inch ID Teflon. p.4 ↗ Use this tubing specification if the trimmed lines need replacement. Whole kit
Bottle netting length should be approximately 17 cm for 500-mL reservoirs and approximately 22 cm for 1000-mL reservoirs. p.5 ↗ Cut netting to bottle height to protect bottles from surface damage. 500-mL and 1000-mL bottles
Netting is suitable for 500-mL and 1000-mL bottles less than 10 cm in diameter. p.6 ↗ Check bottle diameter before using the supplied netting. Whole kit
Tubing from regulator gas outlet to in-line check valve must be cut 13 to 15 cm long. p.6 ↗ Use this length to connect the regulator outlet to the in-line check valve. Whole kit
Tubing from in-line check valve to Y-fitting must be cut 5 to 7 cm long. p.7 ↗ Use this short section between the check valve and Y-fitting. Whole kit
Tubing from Y-fitting to each bottle-cap check valve must be cut 46 to 51 cm long. p.8 ↗ Cut two sections of this length when using two bottles. Whole kit
The in-line check valve arrow must point away from the regulator; otherwise helium cannot flow to the solvent bottles. p.6 ↗ Verify check-valve orientation during installation. Whole kit
Regulator connections must use PEEK fittings and non-tapered ferrules. p.8 ↗ Use the specified fittings to ensure leak-free gas connections at the regulator. Whole kit
The regulator is set to approximately 31 kPa (0.31 bar, 4.5 psi) and must not be tampered with. p.8 ↗ Do not adjust the regulator; overpressure can burst the solvent bottle or separate gas tubing. Whole kit
The pressure relief valve partially opens above approximately 31 kPa and allows 100% venting at approximately 70 kPa (0.7 bar, 10 psi). p.11 ↗ Hissing from the relief valve indicates the regulator setting has been increased above default. Whole kit
When initially supplying helium to the regulator assembly, the bottle-cap shutoff valves must be closed and reopened only after the helium source valve is opened. p.9 ↗ Follow this sequence to avoid pressure surges or improper gas flow. Whole kit
To sparge, open all bottle-cap shutoff valves; to blanket, close the shutoff valve at the unused port and keep all other port valves open. p.10 ↗ Use valve positions to switch between sparging and blanketing. Whole kit
When adding solvents, stop solvent flow, close all bottle-cap shutoff valves, remove the cap assembly, fill the bottle, reinstall the cap, open all valves, sparge, blanket, and restart solvent flow. p.9 ↗ Follow this sequence to avoid introducing air or losing gas stabilization. Whole kit
Suggested tubing lengths in the installed layout are 12.7–15.2 cm from regulator to in-line check valve, 5.1–7.6 cm from in-line check valve to Y-connector, and 2× 45.7–50.8 cm from Y-connector to bottle caps. p.13 ↗ Use these lengths as a guide for a tidy installation. Whole kit

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Gas is too quickly consumed p.11 ↗ Leaks from the gas connections or tubing Use a leak detector, like Snoop®, to detect any leaks in the gas connections or tubing.
Gas is too quickly consumed p.11 ↗ Excessive sparging; regulator setting accidentally increased Adjust the regulator to the required 31 kPa (0.31 bar, 4.5 psi).
Hissing from pressure relief valve (Pressure relief valve opens to 100% venting at about 70 kPa (0.7 bar, 10 psi), which is about twice the default setting.) p.11 ↗ Regulator setting accidentally increased Adjust the regulator to the required 31 kPa (0.31 bar, 4.5 psi).
No gas flow p.12 ↗ Bottle-cap shutoff valve at the check-valve port is closed Open bottle-cap shutoff valve at check valve port.
No gas flow p.12 ↗ Helium source turned off or depleted Ensure that the helium source pressure is between 621 and 689 kPa (6.2 and 6.9 bar, 90 and 100 psi). Replace helium supply.
No gas flow p.12 ↗ Regulator accidentally decreased Adjust the regulator to the required 31 kPa (0.31 bar, 4.5 psi). Ensure that the helium source pressure is between 621 and 689 kPa (6.2 and 6.9 bar, 90 and 100 psi).
Unable to sparge p.12 ↗ Unused shutoff valve on cap closed Open unused shutoff valve on bottle-cap.
Unable to sparge p.12 ↗ Helium source empty Replace helium source.