Fraction Collector III Operator's Guide
Technical Reference Summary
Covered Models & Document Scope
Waters Fraction Collector III (WFC III), a stand-alone programmable fraction collection module. The document covers the base unit, standard components (standard test tube rack, 3-way valve, I/O terminal connector, cable cores GRFC-10/GRFC-8/GRFC-9, Teflon sample tubing 2.0 mm OD x 1.0 mm ID x 1 m, silicone drain tubing 8.0 mm OD x 5.0 mm ID x 0.5 m, tube suspenders, LC connection kit, power cord), and optional racks: prep funnel/funnel rack, 17-mm vial rack, 28-mm vial rack, micro plate rack, Eppendorf tube rack, carousel rack, multi-purpose rack. A low residual 3-way valve is available as an option. Collection configurations: single unit, dual units (MassLynx-controlled, serial connection), and multiple units (up to 16 WFC III units via automatic selector valve).
Section-by-Section Technical Breakdown
Introduces the WFC III as a programmable stand-alone fraction collection module compatible with Waters Autopurification MS/UV systems using FractionLynx, Waters PrepLC systems (Delta 600, 4000 Series, PrepLC 2000), and other Waters and non-Waters HPLC modules. Describes three collection configurations: single unit, dual units for peak and between-peak collection (MassLynx only, serial connection), and multiple units for high-throughput autopurification using an automatic selector valve.
- Single WFC III collects fractions from an HPLC or LC/MS system.
- Dual WFC III units collect peak fractions and volume between peaks; supported only in MassLynx-controlled systems and must be connected serially.
- Multiple WFC III units require an automatic selector valve connected to up to eight units or eight pairs; maximum 16 units per system.
- Compatible with Waters Autopurification MS and UV systems using FractionLynx and any Waters PrepLC system, specifically Delta 600, 4000 Series, and PrepLC 2000.
Lists the standard components shipped with the WFC III and describes the optional racks that replace the standard tube rack. The 3-way valve diverts solvent flow to waste as the dropper assembly moves between vessels; a low residual valve option provides approximately 40% of the standard valve flow rate.
- Standard components include standard test tube rack, 3-way valve (not installed upon shipment), I/O terminal connector, cable cores GRFC-10, GRFC-8, GRFC-9, Teflon sample tubing 2.0 mm OD x 1.0 mm ID x 1 m, silicone drain tubing 8.0 mm OD x 5.0 mm ID x 0.5 m, tube suspenders (2), LC connection kit, power cord.
- Prep funnel and funnel rack: up to 128 large bottles or flasks; up to four prep funnels (32 fractions each); funnel rack is not approved by CSA certification.
- 17-mm vial rack holds 120 vials; 28-mm vial rack holds 56 vials.
- Micro plate rack holds four 96-well micro titer plates (standard or deep-well); use with extension nozzle.
- Eppendorf tube rack holds 120 1.5-mL tubes.
- Carousel rack accommodates one or two 24-vial Alliance carousels.
- Multi-purpose rack accepts one Waters 2700 Sample Manager rack, up to two Gilson FC 204 racks, up to eight GeneVac test tube holders, or a combination of one Gilson FC 204 rack and up to four GeneVac holders.
Describes the utility, simple, and advanced operating modes of the WFC III. Utility modes include DISPLAY RESULT, CONDITIONS, RACK SELECTION, BOTTLE POSITION, and TEST. Simple modes (TIME, DROP, SIGNAL, VOLUME) and advanced modes (PEAK, WINDOW, MANUAL, SAMPLING) support the multi-sample option except MANUAL.
- Utility modes: DISPLAY RESULT (table or graph; graph only for PEAK, WINDOW, MANUAL), CONDITIONS, RACK SELECTION, BOTTLE POSITION, TEST (accessed by selecting CONDITIONS and pressing END).
- Simple modes: TIME (time per vessel), DROP (drops per vessel), SIGNAL (count per vessel using external signal), VOLUME (volume per vessel); all can use multi-sample option.
- Advanced modes: PEAK (discriminates by peak number, absolute value, slope angle, or both), WINDOW (collects only samples complying with parameters within preset time windows; can set PEAK parameters within WINDOW), MANUAL (TEACHING and PLAYBACK functions), SAMPLING (takes samples at specified time intervals; can pool fractions).
Specifies the environmental and power requirements for installing the WFC III, required installation materials, and power connection procedure. The unit accepts 100 to 240 Vac, 50 to 60 Hz.
- Temperature: 4 to 40 °C (39 to 104 °F); relative humidity: 20 to 85%, noncondensing.
- Bench space: W 12 1/4 in. x D 16 1/4 in. x H 14 1/4 in. (W 311 mm x D 413 mm x H 362 mm).
- Power: 0.23 A at 100 V, 0.2 A at 115 V, 0.15 A at 240 V, 50 to 60 Hz.
- Required materials: Waters PEEK tubing cutter (part number WAT031795) or similar, wire insulation strippers, small screwdriver for I/O terminal.
- Do not lift the unit by the carriage rails; always lift with one hand under the base.
- Front and top housing are ABS plastic and can be damaged by solvents; wipe up spills immediately.
Describes how to connect sample tubing, drain tubing, and the optional 3-way valve. Sample tubing must be cut at 90° and protrude 3 to 5 mm from the bottom of the mounting plate; the tubing collar is hand-tightened only.
- Cut sample tubing at a 90° angle; Waters PEEK tubing cutter WAT031795 recommended.
- Insert tubing until it protrudes 3 to 5 mm from the bottom of the mounting plate; hand tighten tubing collar only.
- Tubing collar designed for 2.0-mm OD tubing; for larger OD, join a segment to standard 2.0-mm OD tubing; for smaller OD, wrap tape around tubing at the collar.
- Stainless steel tubing must be connected to a section of Teflon tubing before passing through the dust cover using LC Connection Kit WAT037048.
- Install tubing suspenders if sample tubing is too long or loose to prevent entanglement on dropper assembly or carriage.
Covers making I/O signal connections to the WFC III, including polarity requirements for signal input. The positive connection must be made to the even-numbered pin and the negative connection to the odd-numbered pin.
- I/O terminal connector is attached to the rear of the WFC III body.
- Positive signal connection must be made to the even-numbered pin; negative connection to the odd-numbered pin.
- Connector cable cores GRFC-10, GRFC-8, and GRFC-9 are supplied.
Describes installation of the standard test tube rack and optional racks, including test tube supports and procedures for prep funnel, 28-mm vial, micro plate, carousel, 17-mm vial, Eppendorf, and multi-purpose racks.
- Standard test tube rack installation includes test tube supports.
- Optional racks replace the standard tube rack and are installed per the procedures in section 2.8.
- The prep funnel rack is not approved by CSA certification.
Covers powering on the unit, initializing memory, LCD display and keypad overview, and preliminary setup including setting conditions, selecting the rack, programming bottle positions, and dropper assembly movement.
- Power on the unit and initialize memory as described in section 3.1.1.
- LCD display and keypad functions are described in section 3.2.
- Preliminary setup includes setting conditions, selecting the rack, programming bottle positions, and dropper assembly movement.
Describes collection using TIME, DROP, SIGNAL, and VOLUME modes, including parameter setting, running the program, and sample programs for each mode.
- TIME mode collects fractions at set time intervals per vessel.
- DROP mode collects a set number of drops per vessel.
- SIGNAL mode collects a set count per vessel using an external signal.
- VOLUME mode collects a set volume per vessel.
- All simple modes support the multi-sample option.
Describes collection using PEAK discrimination, WINDOW time windows, MANUAL programmed sequence (TEACHING/PLAYBACK), and SAMPLING intervals, plus displaying results as table or graph and using event markers to track peaks.
- PEAK mode discriminates peaks by peak number, absolute value, slope angle, or absolute value and slope angle.
- WINDOW mode collects only samples complying with specified parameters within preset time windows; PEAK parameters can be set within WINDOW.
- MANUAL mode uses TEACHING to record a sequence of commands and PLAYBACK to replay it automatically.
- SAMPLING mode takes samples at specified time intervals and can pool repeated samples into the same vessel.
- Graph display is available only for PEAK, WINDOW, and MANUAL collection runs.
- Event markers can be used to track peaks.
Provides maintenance guidelines, spare parts reference, and test mode diagnostics for the dropper assembly movement, panel keys, 3-way valve, I/O terminal, chromatography signal, and drop sensor. The drop sensor test requires a difference of more than 300 between uncovered and covered light source hole readings to confirm correct operation.
- Spare parts are listed in Appendix C; parts not listed are not recommended for customer replacement.
- Test mode is accessed by selecting CONDITIONS on the MAIN MENU and pressing END.
- Drop sensor test: record drop level with light source hole uncovered, cover hole, record value; difference greater than 300 indicates correct operation.
- Diagnostics include testing dropper assembly movement, panel keys, 3-way valve, I/O terminal, chromatography signal, and drop sensor.
Provides basic troubleshooting steps and tables for general system, signal, and collection troubleshooting. Basic steps include checking simple causes, comparing current operation with normal operation, and identifying symptoms in order.
- Basic troubleshooting: check simple things first (unplugged or improperly connected instruments), compare current operation with normal operation, keep a daily log, run test chromatograms regularly.
- Fill out a Normal Operation Checklist at installation and each time a new method is developed.
- Troubleshooting tables cover general system, signal, and collection problems.
- For signal connections, positive must be on even-numbered pin and negative on odd-numbered pin.
Appendix B contains WFC III specifications, including 3-way valve specifications referenced from the components section.
- 3-way valve specifications are provided in Appendix B (page 132).
- Low residual 3-way valve flow rate is approximately 40% of the standard valve flow rate.
Appendix C lists Fraction Collector III spare parts and accessories.
- Spare parts are listed in section C.1.
- Accessories are listed in section C.2.
Appendix D provides information on the compatibility of the WFC III components with a variety of chemicals, including clean solvents and solvent quality.
- D.1 Introduction covers clean solvents and solvent quality.
- D.2 provides component chemical compatibility information.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Operating temperature range: 4 to 40 °C (39 to 104 °F); relative humidity 20 to 85%, noncondensing. p.28 ↗ | Install the WFC III only in a controlled laboratory environment within these limits to ensure reliable operation. | Whole series |
| Bench space required: W 12 1/4 in. x D 16 1/4 in. x H 14 1/4 in. (W 311 mm x D 413 mm x H 362 mm). p.28 ↗ | Plan bench space accordingly, leaving room for fluidic and electrical connections. | Whole series |
| Power consumption: 0.23 A at 100 V, 0.2 A at 115 V, 0.15 A at 240 V, 50 to 60 Hz. p.28 ↗ | Ensure the power outlet and circuit can supply the required current at the local voltage. | Whole series |
| Power input: 100 to 240 Vac, 50 to 60 Hz. p.29 ↗ | The unit can be used worldwide with an appropriate regional power cord. | Whole series |
| Sample tubing must be cut at a 90° angle and inserted until it protrudes 3 to 5 mm from the bottom of the mounting plate. p.31 ↗ | Correct tubing protrusion ensures proper drop formation and reliable collection. | Whole series |
| Tubing collar is designed for 2.0-mm OD tubing; for larger OD, join a segment to standard 2.0-mm OD tubing; for smaller OD, wrap tape around tubing at the collar. p.31 ↗ | Allows use of alternative tubing sizes while maintaining a secure connection. | Whole series |
| Stainless steel tubing must be connected to a section of Teflon tubing before passing through the dust cover using LC Connection Kit WAT037048. p.32 ↗ | Prevents damage and ensures proper fluidic connection when using stainless steel tubing. | Whole series |
| Positive signal connection must be made to the even-numbered pin; negative connection to the odd-numbered pin. p.38 ↗ | Incorrect polarity will prevent signal input and collection in SIGNAL and PEAK modes. | Whole series |
| Maximum number of WFC III units that can be connected to a system is 16. p.17 ↗ | Plan high-throughput configurations within this limit using an automatic selector valve. | Multiple WFC III configuration |
| Dual WFC III units must be connected serially and are supported only in MassLynx-controlled systems. p.17 ↗ | Dual configuration requires MassLynx software and serial connection. | Dual WFC III configuration |
| Low residual 3-way valve flow rate is approximately 40% of the standard valve flow rate. p.20 ↗ | Select the low residual valve when reduced flow resistance is required; account for lower flow capacity. | Optional low residual 3-way valve |
| Drop sensor test: difference between drop level with light source hole uncovered and covered must be more than 300 for correct operation. p.113 ↗ | Use this test to verify drop sensor functionality during diagnostics. | Whole series |
| 17-mm vial rack holds 120 vials; 28-mm vial rack holds 56 vials; Eppendorf tube rack holds 120 1.5-mL tubes. p.21 ↗ | Choose rack based on sample volume and number of fractions required. | Optional racks |
| Micro plate rack holds four 96-well micro titer plates (standard or deep-well) and should be used with the extension nozzle. p.22 ↗ | Use extension nozzle for proper collection into micro plates. | Micro plate rack |
| Prep funnel rack can hold up to 128 large bottles or flasks using up to four prep funnels (32 fractions each); funnel rack is not approved by CSA certification. p.20 ↗ | Use for large-volume fraction collection; note CSA certification limitation. | Prep funnel and funnel rack |
| Carousel rack accommodates one or two 24-vial Alliance carousels. p.23 ↗ | Allows collection into 2-mL vials using Alliance carousels. | Carousel rack |
| Multi-purpose rack accepts one Waters 2700 Sample Manager rack, up to two Gilson FC 204 racks, up to eight GeneVac test tube holders, or a combination of one Gilson FC 204 rack and up to four GeneVac holders. p.23 ↗ | Provides flexibility for various sample container types. | Multi-purpose rack |
| Required installation materials include Waters PEEK tubing cutter (part number WAT031795) or similar, wire insulation strippers, and a small screwdriver. p.27 ↗ | Have these tools available before starting installation. | Whole series |
| Front and top housing are ABS plastic and can be damaged by solvents; wipe up solvent spills immediately. p.28 ↗ | Avoid solvent contact with housing to prevent damage. | Whole series |
| Do not lift the unit by the carriage rails; always lift with one hand under the base. p.28 ↗ | Prevents damage to the carriage mechanism during unpacking or moving. | Whole series |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Unit does not power on p.114 ↗ | Power cord not connected | Check power cord. | Table 6-1 |
| Unit does not power on p.114 ↗ | No power at outlet | Check line voltage. | Table 6-1 |
| Unit does not power on p.114 ↗ | Power supply fuse blown | Call Waters Technical Service. | Table 6-1 |
| No response from keypad p.114 ↗ | Keypad is broken | Call Waters Technical Service. | Table 6-1 |
| Screen is blank p.115 ↗ | Unit not powered on | Power on unit. | Table 6-1 |
| Startup diagnostic test failure p.115 ↗ | Not specified | Power off and on again. If failure not corrected, call Waters Technical Service. | Table 6-1 |
| Display dim p.115 ↗ | Not specified | Adjust LCD contrast using CONDITIONS mode. | Table 6-1 |
| Display is broken p.115 ↗ | Not specified | Call Waters Technical Service. | Table 6-1 |
| Front panel software "lock up" p.115 ↗ | Software problem or power line problem | Initialize unit's memory (see "Initializing the memory" on page 48). If failure not corrected, call Waters Technical Service. | Table 6-1 |
| LCD displays nonsense p.115 ↗ | Strong noise (static electric discharge) | Check for nearby noise source and remove. | Table 6-1 |
| LCD is completely dark or white p.115 ↗ | CPU was not reset | Power off unit, then power on unit. If failure not corrected, call Waters Technical Service. | Table 6-1 |
| LCD is completely dark or white p.115 ↗ | Strong noise (static electric discharge) | Check for nearby noise source and remove. | Table 6-1 |
| Displays power failure message without a power failure p.115 ↗ | Bad connection in AC plug | Check and replace. | Table 6-1 |
| Displays power failure message without a power failure p.115 ↗ | Power switch turned off during run | Turn on power switch. | Table 6-1 |
| Parameters not saved p.115 ↗ | Bad battery | Call Waters Technical Service. | Table 6-1 |
| Cannot input chromatogram [SIGNAL] p.116 ↗ | Bad connecting cord | Check and replace. | Table 6-2 |
| Cannot input chromatogram [SIGNAL] p.116 ↗ | Bad I/O connector | Check and replace. | Table 6-2 |
| Cannot input chromatogram [SIGNAL] p.116 ↗ | Wrong connection with +/− | Check +/− connection and connect properly. The positive connection must be made to the even-numbered pin; the negative connection to the odd-numbered pin. | Table 6-2 |
| [EXT START] signal not accepted p.116 ↗ | Bad connecting cord | Check and replace. | Table 6-2 |
| [EXT START] signal not accepted p.116 ↗ | Bad I/O connector | Check and replace. | Table 6-2 |
| [EXT END] signal not accepted p.116 ↗ | Bad connecting cord | Check and replace. | Table 6-2 |
| [EXT END] signal not accepted p.116 ↗ | Bad I/O connector | Check and replace. | Table 6-2 |
| [EXT COUNT] signal not accepted p.116 ↗ | Bad connecting cord | Check and replace. | Table 6-2 |
| [EXT COUNT] signal not accepted p.116 ↗ | Bad I/O connector | Check and replace. | Table 6-2 |
| Cannot input [EVENT MARKER] into a recorder p.116 ↗ | Bad connecting cord | Check and replace. | Table 6-2 |
| Cannot input [EVENT MARKER] into a recorder p.116 ↗ | Bad I/O connector | Check and replace. | Table 6-2 |
| [EVENT MARKER 2] signal not accepted p.116 ↗ | Bad connecting cord | Check and replace. | Table 6-2 |
| [EVENT MARKER 2] signal not accepted p.116 ↗ | Bad I/O connector | Check and replace. | Table 6-2 |
| Collection position out of alignment on X-axis (right/left) p.116 ↗ | Incorrect length of sample tubing | Check and connect properly. | Table 6-3 |
| Collection position out of alignment on X-axis (right/left) p.116 ↗ | X-slider mechanism may be corroded or dusty | Clean and lubricate with silicone oil. | Table 6-3 |
| Collection position out of alignment on Y-axis (front/back) p.116 ↗ | Incorrect length of Teflon tubing | Check and connect properly. | Table 6-3 |
| Collection position out of alignment on Y-axis (front/back) p.116 ↗ | Y-slider mechanism may be corroded or dusty | Clean and lubricate with silicone oil. | Table 6-3 |
| Collection volume varies in DROP mode p.117 ↗ | Other light sources interfere with drop sensor | Check to see if light from windows or other sources shines on sensor. Remove or block other light sources or relocate WFC III. | Table 6-3 |
| Collection volume varies in DROP mode p.117 ↗ | Vibration interferes with proper collection | Check for compressor or other nearby source of vibration. Remove source of vibration or relocate WFC III. | Table 6-3 |
| Collection volume varies in DROP mode p.117 ↗ | Bubbles in buffer | Check sealing at joint for completeness. Reseal if necessary. | Table 6-3 |
| Collection volume varies in DROP mode p.117 ↗ | Crystals stuck to end of tubing at dropper assembly | Remove crystals stuck to tubing and flush tubing with clean solvent. | Table 6-3 |
| Collection volume varies in DROP mode p.117 ↗ | Tubing end at dropper assembly not cut straight | Recut tubing at a 90° angle. | Table 6-3 |
| Collection volume varies in DROP mode p.117 ↗ | Test tubes and dropper assembly not centered | Check the positioning and readjust to center. | Table 6-3 |
| Collection volume varies in DROP mode p.117 ↗ | Tubing at dropper assembly too short | Adjust tubing so that 3 to 5 mm protrudes from dropper assembly. | Table 6-3 |
| No collection in DROP mode p.117 ↗ | Protruding length of tubing at dropper assembly is longer than 5 mm | Adjust tubing so that 3 to 5 mm protrudes from dropper assembly. | Table 6-3 |
| No collection in DROP mode p.117 ↗ | Other light sources interfere with drop sensor | Check to see if light from windows or other sources shines on sensor. Remove or block other light sources or relocate WFC III. | Table 6-3 |
| No collection in DROP mode p.117 ↗ | Teflon tubing interferes with light pass | Check and adjust tubing. | Table 6-3 |
| Collection times vary in TIME and VOLUME modes p.118 ↗ | Bubbles in buffer | Check sealing at joint for completeness. Reseal if necessary. | Table 6-3 |
| Collection times vary in TIME and VOLUME modes p.118 ↗ | Crystals stuck to end of tubing at dropper assembly | Remove crystals stuck to tubing and flush tubing with clean solvent. | Table 6-3 |
| Collection times vary in TIME and VOLUME modes p.118 ↗ | Tubing end at dropper assembly not cut straight | Recut tubing at a 90° angle. | Table 6-3 |
| Collection times vary in TIME and VOLUME modes p.118 ↗ | Improper positioning of both center of tubes and dropper assembly | Check tube rack position and reinstall properly. | Table 6-3 |
| Incorrect count time in TIME mode p.118 ↗ | Extreme variation in operating temperatures | Control ambient temperature. | Table 6-3 |
| Incorrect count time in TIME mode p.118 ↗ | Software problem | Check for reproducibility of error, replace ROM if bad. | Table 6-3 |
| Cannot collect in PEAK mode both or either absolute and slope values p.118 ↗ | Bad connecting cords between detector and WFC III | Check and replace. | Table 6-3 |
| Cannot collect in PEAK mode both or either absolute and slope values p.118 ↗ | Bad I/O connector | Check and replace. | Table 6-3 |
| Cannot collect in PEAK mode both or either absolute and slope values p.118 ↗ | Connection polarity +/− is incorrect | Check +/− connection and connect properly. The positive connection must be made to the even-numbered pin; the negative connection to the odd-numbered pin. | Table 6-3 |
| Cannot collect in PEAK mode both or either absolute and slope values p.118 ↗ | Collection parameters set incorrectly | Check and reset. | Table 6-3 |