Agilent Technologies 1260 Infinity Amino Acid Analysis System
Agilent Technologies 1260 Infinity Amino Acid Analysis System — Liquid chromatography.
The 1260 Infinity Amino Acid Analysis System is a dedicated LC configuration built around the 1260 Infinity II Vialsampler or Multisampler with a 100 µL analytical head and 100 µL loop, paired with an AdvanceBio AAA column (3.0 mm × 2.7 µm) and guard (4.6 mm). It performs automated OPA/FMOC derivatization at room temperature with fluorescence detection at 340/260 nm excitation and a 10-stage PMT gain, using a 0.4 M borate buffer at pH 10.2 and 2.5 mg/mL FMOC reagent. This modification is the application-specific amino acid workflow, distinct from the general-purpose Bio-Inert and Prime Bio siblings.
Agilent Technologies Bio LC Solutions Series — Model Comparison & Selection
| Model | Flow rate, liter/minute | Mass range | Max pressure, megapascal | Pumping speed, m³/h | Wavelength range | Sample volume, milliliter |
|---|---|---|---|---|---|---|
| 1260 Infinity Amino Acid Analysis System Current | — | — | — | — | — | — |
| 1260 Infinity III Bio-Inert LC System | 0.01 | — | 60 | 6×10⁻⁴ | 190...950 | — |
| 1290 Infinity III Bio LC | Biocompatible UHPLC system | — | 300...3,200 | — | — | — | 0.005 |
Full specifications
| Analytical Head Volume | 100 µL |
| Loop Volume | 100 µL |
| Vial Volume | 2 mL |
| AdvanceBio AAA LC column dimensions mfr | 3.0 mm |
| AdvanceBio AAA LC column particle size mfr | 2.7 μm |
| AdvanceBio AAA guard column dimensions mfr | 4.6 mm |
| Borate buffer concentration mfr | 0.4 M |
| Borate buffer pH mfr | 10.2 |
| FMOC reagent concentration mfr | 2.5 mg/mL |
| DTDPA reagent quantity mfr | 5 g |
| Fluorescence detection excitation (primary) mfr | 340 nm |
| Fluorescence detection excitation (secondary) mfr | 260 nm |
| PMT gain mfr | 10 |
| Gradient %B at 0.35 min mfr | 2 |
| Gradient %B at 13.4 min mfr | 57 |
| Sample thermostat option mfr | 101 |
| Diluent draw volume mfr | 80 μL |
| Wait time mfr | 0.6 minutes |
| Column mfr | 120 |
| Calibration concentration levels mfr | 20 |
| Caffeine stock solution concentration mfr | 100 μg/mL |
| Calibration function slope (Multisampler 100 μL, manual) mfr | 27.85 |
| Calibration function slope (Multisampler 100 μL, autosampler) mfr | 27.64 |
| Calibration function slope (Vialsampler 100 μL, manual) mfr | 27.76 |
| Calibration function slope (Vialsampler 100 μL, autosampler) mfr | 27.65 |
| Calibration function slope (Multisampler 900 μL, manual) mfr | 27.69 |
| Calibration function slope (Multisampler 900 μL, autosampler) mfr | 27.53 |
| Calibration function R2 (Multisampler 100 μL, manual) mfr | 0.99999 |
| Calibration function R2 (Multisampler 100 μL, autosampler) mfr | 1.00000 |
| Calibration function R2 (Vialsampler 100 μL, manual) mfr | 0.99999 |
| Calibration function R2 (Vialsampler 100 μL, autosampler) mfr | 0.99999 |
| Calibration function R2 (Multisampler 900 μL, manual) mfr | 0.99999 |
| Calibration function R2 (Multisampler 900 μL, autosampler) mfr | 1.00000 |
| Mobile phase A composition mfr | 10 |
| Signal-to-noise acceptance criterion mfr | 3 |
| AA standard concentration mfr | 1 nmol/μL |
| AA standard quantity mfr | 250 pmol |
Description
Construction and series positioning
- This is a complete amino acid analysis (AAA) application kit built on 1260 Infinity II LC modules: Flexible Pump (G7104C), Vialsampler (G7129C) or Multisampler (G7167A), Multicolumn Thermostat (G7116A), and Diode Array Detector HS (G7117C) configured for fluorescence.
- The default injector configuration is a 100 µL analytical head with a 100 µL loop; the Multisampler can alternatively be fitted with a 900 µL analytical head (option #163) and 900 µL loop (option #156).
- Unlike the 1260 Infinity III Bio-Inert LC System (600 bar, 10 mL/min, 190–950 nm) or the 1290 Infinity III Bio LC (300–3,200 mass range), this model is not a high-pressure or mass-spec front end; it is a derivatization-and-fluorescence workflow for amino acid quantification.
Derivatization and separation behavior
- Samples are derivatized at room temperature using OPA and FMOC chemistry; the borate buffer is 0.4 M at pH 10.2, FMOC reagent is 2.5 mg/mL, and DTDPA is supplied as 5 g.
- Separation uses an AdvanceBio AAA LC column (3.0 mm, 2.7 µm particles) with a 4.6 mm guard column. The gradient reaches 2 %B at 0.35 min and 57 %B at 13.4 min.
- Fluorescence detection uses 340 nm primary and 260 nm secondary excitation with PMT gain 10.
- Calibration standards are prepared at 1 nmol/µL (250 pmol injected); the injector program can perform serial dilutions, with calibration slopes of 27.85 (Multisampler 100 µL, manual) and 27.64 (Multisampler 100 µL, autosampler), and R² values of 0.99999–1.00000.
Installation and operation
- The injector program handles dilution: an 80 µL diluent draw volume and a 0.6 min wait time are used in the serial-dilution procedure.
- Sample thermostat option #101 is available for temperature control of the sample compartment.
- The system is assembled from discrete InfinityLab LC Series modules; the derivatization reagents, columns, and consumables are separate items from the base instrument delivery.
Integration & Operating Notes
The AdvanceBio AAA LC column is 3.0 mm with 2.7 µm particles, while the guard column is 4.6 mm; the guard must be installed on the matching hardware and is a consumable replaced independently of the analytical column. Confirm the column compartment (Multicolumn Thermostat G7116A) holds the 3.0 mm column format.
Selection & Integration Q&A
Choose this AAA system when the workflow is OPA/FMOC amino acid quantification with fluorescence detection at 340/260 nm and the AdvanceBio AAA column. The Bio-Inert LC System is a general-purpose 600 bar, 10 mL/min, 190–950 nm platform for bio-inert separations, not a derivatization kit.
The method is validated on a 1260 Infinity II Vialsampler (G7129C) or Multisampler (G7167A) with 100 µL analytical head and 100 µL loop, plus Flexible Pump G7104C, Multicolumn Thermostat G7116A, and Diode Array Detector HS G7117C. The Multisampler can alternatively use a 900 µL head (option #163) and 900 µL loop (option #156).
Sample thermostat option #101 is available for the sampler. The injector program performs serial dilution with an 80 µL diluent draw and 0.6 min wait; calibration slopes are 27.85 (Multisampler 100 µL, manual) and 27.64 (autosampler), with R² of 0.99999–1.00000.
Borate buffer (0.4 M, pH 10.2), FMOC reagent (2.5 mg/mL), DTDPA (5 g), amino acid standard (1 nmol/µL, 250 pmol), AdvanceBio AAA LC column (3.0 mm, 2.7 µm) and guard (4.6 mm), plus vials, caps, and septa are separate consumables ordered independently of the instrument.