Agilent Technologies 6230C LC/MS TOF

Agilent Technologies 6230C LC/MS TOF — Liquid chromatography mass spectrometry lc ms with Pumping speed 3×10⁻⁵ m³/h.

The Agilent 6230C is an accurate-mass time-of-flight LC/MS system with a Dual Agilent Jet Stream ion source, covering 800–8,000 m/z at 1.0 spectra/s. It is configured for full-spectrum acquisition with internal reference mass correction, achieving a 1.0 ng/mL detection limit. Its primary purpose is accurate-mass screening and confirmation in LC/MS workflows.

Pumping speed
3×10⁻⁵ m³/h

Full specifications

Core Performance & Functional Specifications
Detection limit 1.0 ng/mL
Pumping speed 0.5 mL/min
Gas Flow 13 L/min
Ion source Dual Agilent Jet Stream
Nebulizer Gas Pressure 40 psi
Post Time 1.0 minutes
Run Time 20.0 minutes
Sample volume 2 µL
Scan speed 1.0 spectra/sec
Sheath Gas Flow 12 L/min
LC/MS system mfr Accurate-Mass TOF LC/MS LC/MS
UHPLC column mfr 3 μm
Stop time mfr 12.5 minutes
Scan range (MS-mode) mfr 50...1,000 amu
Scan rate (MS-mode) mfr 1.5 spectra/s
Scan range (All Ions MS/MS mode) mfr 50...1,000 amu
Scan rate (All Ions MS/MS mode) mfr 3 spectra/s
Internal reference mfr 921
Sample preparation mfr 90
Column mfr 2.1
Linear regression coefficient mfr 0.999
Calibration range mfr 10...1,000 μg/mL
THC concentration in hemp products mfr 0.3
Dimensions, Weight & Mechanical Design
Mass range 800...8,000 m/z
Ref mass enabled mfr Yes Yes
Electrical, Power & Drive Specifications
Capillary Voltage 4000 V
Nozzle Voltage 2000 V
VCap mfr 4000 V
Fragmentor mfr 175 V
Skimmer mfr 65 V
Mobile phase A mfr 10 water
Mobile phase B mfr 0.05 acetonitrile
Operating Conditions, Environment & Protection
Gas Temperature 325 °C
Sheath Gas Temperature 350 °C

Description

Construction and positioning

Behavior under load and in different modes

Installation, connection and operation

Integration & Operating Notes

Process Limits
Ion source gas conditions must be held at specified setpoints

Sheath gas flow and temperature (12 L/min, 350 °C), drying gas flow and temperature (13 L/min, 325 °C), and nebulizer pressure (40 psi) are the verified source conditions. Deviating from these setpoints changes desolvation and ion transmission, so they should be treated as method conditions rather than adjustable margins.

Factory Specification Sheet
Electrical
Capillary and nozzle voltages are fixed method parameters

Capillary voltage is 4000 V and nozzle voltage is 2000 V in the verified configuration. These potentials define the electrospray regime and must be reproduced in the method; no alternative voltage settings are documented in the verified facts.

Factory Specification Sheet
Process Limits
Mass range is bounded at 800–8,000 m/z

Acquisitions must be planned within the stated 800–8,000 m/z range. Ions outside this window are not covered by the verified specification, so method scan ranges and target masses should be checked against this bound before committing to a workflow.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose the 6230C rather than another TOF LC/MS variant in the Agilent line?

The verified facts list no sibling models, so a direct comparison cannot be made. Choose the 6230C when the required workflow needs accurate-mass TOF acquisition over 800–8,000 m/z at 1.0 spectra/s with a 1.0 ng/mL detection limit and the Dual Agilent Jet Stream source.

What LC conditions must the upstream chromatography meet to match this system's verified method?

The verified method uses 0.5 mL/min pumping speed, 2 µL sample volume, 20.0 min run time and 1.0 min post time. The LC deliverable must sustain these flows and injection volumes; no other column or mobile-phase requirements are stated in the verified facts.

What source gas supply conditions must be provided at the installation site?

The site must supply nebulizer gas at 40 psi and support sheath gas at 12 L/min and 350 °C plus drying gas at 13 L/min and 325 °C. Gas capacity and purity must be confirmed against these flows before installation; no further utility requirements are documented.

What maintenance schedule and consumables apply, and what is not included in base delivery?

No maintenance intervals, service procedures or accessory lists are present in the verified facts or document digests. Consumables and cables cannot be specified from the available data; these must be obtained from the manufacturer's manual, which was not supplied for this card.