Agilent Technologies Reliable routine analysis gas chromatography system
Agilent Technologies Reliable routine analysis gas chromatography system — Gas chromatography.
The Agilent 7820A GC System is a gas chromatograph available with EPC for all inlets and detectors or EPR for select inlets and detectors, operating from 8 to 425 °C with a maximum inlet temperature of 400 °C. It supports split ratios up to 250 and flow adjustment to 200 mL/min, with a 100 Hz data acquisition rate. It serves as the analytical core for routine GC separations in laboratory workflows.
Agilent Technologies GC Systems Series — Model Comparison & Selection
| Model | Operating temperature, °C | Sample volume, milliliter | Carrier gas | Collision gas nitrogen, mL/min | Column temperature, °C | Data rate, Hz |
|---|---|---|---|---|---|---|
| Reliable routine analysis gas chromatography system Current | 8...425 | — | — | — | — | — |
| Accurate and Reliable GC System with Faster Cycle Times | — | 0.002 | 2 | 1.5 | 85 | 10 |
Full specifications
| Maximum Data Acquisition Rate | 100 Hz |
| Peak Area, C13 | 45.2143 pA |
| Peak Area, C14 | 45.5790 pA |
| Peak Area, C15 | 457.8670 pA |
| Peak Area, C16 | 456.6820 pA |
| Retention Time, C13 | 3.7843 min |
| Retention Time, C14 | 4.4213 min |
| Retention Time, C15 | 5.0369 min |
| Retention Time, C16 | 5.6238 min |
| Maximum split ratio mfr | 250 |
| Minimum detection level (MDL) mfr | 3 tridecane |
| Linear dynamic range mfr | 107 |
| NPD selectivity mfr | 25000...1 gP/gC |
| NPD dynamic range mfr | 104 P |
| heated valve control mfr | two heated valves valves |
| PCOC minimum column id mfr | 0.250 id |
| instruments per PC (EZChrom Compact) mfr | two instruments instruments |
| Mains frequency mfr | 47.5...63 Hz |
| Operating temperature | 8...425 °C |
Description
Construction and series positioning
The 7820A GC System is the base gas chromatograph in a series that also includes the 7650A and 7693A automatic liquid samplers and the 7697A and 8697 headspace samplers. Unlike those dedicated front-end sampling devices, the 7820A is the separation and detection instrument itself. Two pneumatic configurations exist: EPC (electronic pneumatics control) for all inlets and detectors, and EPR (electronic pneumatics regulation) for select inlets and detectors. The system uses a 5-button keypad for local control and supports optional automated injectors and samplers.
Operating behavior
- The column oven operates from 8 to 425 °C; the inlet maximum is 400 °C.
- Split ratios up to 250 and flow adjustable to 200 mL/min define the inlet operating envelope.
- The maximum data acquisition rate is 100 Hz, sufficient for fast capillary peaks.
- The NPD selectivity is 25000:1 (gP/gC) with a dynamic range of 10^4; the linear dynamic range is 10^7 and the minimum detection level is 3 pg tridecane.
- Retention times and peak areas for C13–C16 are documented (e.g., C13 at 3.7843 min, 45.2143 pA; C16 at 5.6238 min, 456.682 pA), providing a reference for column and method performance.
Control, software, and installation
Local control is via a 5-button keypad. The system requires a mains frequency of 47.5–63 Hz. Software options include OpenLAB CDS Workstation/Workstation Plus, OpenLAB CDS VL Workstation/Workstation Plus, OpenLAB CDS ChemStation Edition (C.01.05 or higher), EZChrom Edition (A.04.05 or higher), OpenLAB CDS ChemStation VL, EZChrom VL, OpenLAB CDS EZChrom Compact, and DA Express (China and Russia only). Automated injectors and samplers are optional and not part of the base instrument.
Integration & Operating Notes
The instrument is specified for 47.5–63 Hz mains frequency. Operating outside this window risks unreliable electronics and invalid run data; verify site power quality before installation, especially on generator or unstable supply.
While the oven reaches 425 °C, the inlet maximum is 400 °C. Setting an inlet method above this limit can degrade septa, columns, and inlet hardware; method development must cap inlet temperature accordingly.
OpenLAB CDS ChemStation Edition requires C.01.05 or higher and EZChrom Edition requires A.04.05 or higher. Older software builds will not support the instrument; confirm the workstation version before commissioning.
The 7650A/7693A automatic liquid samplers and 7697A/8697 headspace samplers are separate optional accessories. Budget and plan for them explicitly if unattended operation is required; the base 7820A includes only the GC with keypad and pneumatics.
Selection & Integration Q&A
The 7820A is the gas chromatograph itself (oven, inlets, detectors, pneumatics). The 7697A and 8697 are headspace samplers that feed a GC; the 7650A and 7693A are liquid autosamplers. Choose the 7820A when you need the separation/detection instrument; add a sampler only for automated sample introduction.
The system requires a compatible Agilent data system: OpenLAB CDS ChemStation Edition C.01.05 or higher, EZChrom Edition A.04.05 or higher, or OpenLAB CDS Workstation/Workstation Plus. Local control is via the 5-button keypad. Confirm the workstation version and PC meet the software prerequisites before installation.
Automated liquid samplers (7650A, 7693A) and headspace samplers (7697A, 8697) are optional and not part of the base instrument. Columns, consumables, and the data-system software licenses are also separate. Plan for these items and for the software version requirements when budgeting.