Agilent Technologies 850-DS Sampling Station
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies 850-DS Sampling Station

Agilent Technologies 850-DS Sampling Station — Dissolution testing.

The Agilent 850-DS G7930A is a benchtop dissolution sampling station with an integrated 8-syringe rotary piston pump, 0.1–14 mL sample volumes and up to 36 timepoints per method. It differs from the optional G7931A filtration module by omitting automated filtration, and serves as the sampling front end for 708-DS, Apparatus 3/7 and NanoDis systems.

Full specifications

Core Performance & Functional Specifications
Wavelength range 190 nm
Baseline Flatness 0.001 Abs
Photometric accuracy 0.005 Abs
Printer Optional internal thermal printer
Filtration pore size (optional filter module) mfr 0.2 μm
Sample collection formats mfr 96 plates
Pump type mfr integrated syringe pump pump
Pump speed mfr variable variable
Automated cleaning mfr rinse port enables automated cleaning cycle cycl
Simultaneous system control mfr four systems systems
Software package mfr Cary WinUV WinUV
Number of apparatus mfr 1...2
Lamp type mfr Xenon flash flash
Instrument design mfr Scanning Scanning
Wavelength accuracy mfr 0.5 nm
Wavelength reproducibility mfr 0.1 nm
Stray light mfr 1.0
Safety standards mfr 2 only
Pumping system and speed mfr 8 mL/min
Sampling accuracy mfr 10
Sample volume per vial/tube mfr 0.1 mL
Number of samples per test mfr 36 change
Maximum run duration mfr 999 hours
Filtration options mfr 1 μm
Valves mfr PEEK body body
Needles mfr Anti-clogging stainless steel tips to prevent cl
Volumetric calibration mfr Simplified routine with stored calibration coeff
Vials mfr 12 mm
Test tubes mfr 16 mm
Cleaning volume mfr 0.1 mL
Microtiter plates mfr 96 tray
Method storage mfr 20
Method export mfr Printer, SD card or serial interface interface
PC instrument control mfr 1.7 Dissolut
Good laboratory practice features mfr User management via firmware or software, error
System and pumping speed mfr 8 mL/min
Sampling precision mfr 10
Path lengths mfr 0.2 mm
Système et vitesse de pompage mfr 8 ajustable
Précision d’échantillonnage mfr 10
Volume d’échantillon par flacon/tube mfr 0.1 mL
Fréquence d’échantillonnage mfr 2 minutes
Nombre d’échantillons par test mfr 36 plateau
Durée maximale d’exécution mfr 999 heures
Options de filtration mfr 1 μm
Vannes mfr Corps en PEEK PEEK
Aiguilles mfr Pointes en acier inoxydable anti-obturation pour
Étalonnage volumétrique mfr Routine simplifiée avec un coefficient d’étalonn
Flacons mfr 12 mm
Tubes à essai mfr 16 mm
Types de milieux mfr Compatible avec tous les milieux de dissolution
Remplacement de milieu mfr 100
Volume de nettoyage mfr 0.1 chantillonnage
Clear (effacer) mfr Efface toutes les valeurs entrées et remet les v
Save (enregistrer) mfr Enregistre tous les paramètres de la méthode et
Cleaning Cycle (cycle de nettoyage) Volume mfr Saisissez le volume qui sera pompé à travers les
# Cycles to Run (nombre de cycles à lancer) mfr Ce paramètre contrôle le nombre de répétitions d
Run (Lancement) mfr Lance un cycle de nettoyage avec les paramètres
Stop mfr Arrête un cycle de nettoyage. nettoyage.
Plaques microtitres mfr 96 plateau
Stockage de méthode mfr 20
Export de méthode mfr Imprimante, carte SD ou interface série rie
Imprimante mfr Imprimante thermique interne en option option
Fonctionnalités liées aux bonnes pratiques de laboratoire mfr Gestion des utilisateurs via micrologiciel ou lo
Caisson mfr Tous les matériaux sont recyclables recyclables
Spindle Speed (vitesse de l’arbre) mfr Vitesse de l’arbre souhaitée pour cette méthode.
Rotation Start Delay (retard de démarrage de rotation) mfr Temps après le démarrage de la méthode avant que
Apparatus 1/2/5/6 (appareil 1/2/5/6) mfr Sélection du type d’appareil approprié.
Initial Sample (échantillon initial) mfr Prise d’un échantillon avant de déposer une dose
RPM (tours/min) mfr Vitesse actuelle de l’arbre. arbre.
Duration (durée) mfr Durée pendant laquelle la rotation finale sera e
Import mfr Permet à l’utilisateur d’importer une méthode à
Cancel (annuler) mfr Permet de quitter les écrans de saisie de la mét
oC mfr Température actuelle du bain. bain.
Next Sample (échantillon suivant) mfr Temps avant la prochaine collecte d’échantillon.
Sample Time (visualisation des points d’acquisition) mfr Visualisation des points d’acquisition dans la m
Needle Position (Position d’aiguille) mfr Flèche vers le haut haut
Tray Position (Position de plateau) mfr Eject Eject
Pump Control (commande de la pompe) mfr Volume Volume
Maximum acquisition points per method mfr 36
Closed loop sampling mfr Yes Yes
Path lengths (mm) mfr 0.2
Automated dosage delivery mfr Yes Yes
Sampling mfr Simultaneous Simultaneous
Sample filtration (min. pore size μm) mfr 0.2 μm
Sample archival mfr Yes Yes
minimum reading interval mfr 45 seconds
lamp lifetime mfr 10 years
photometric linear range mfr 3.5 Abs
Dimensions, Weight & Mechanical Design
Slit Width 1.5 nm
Weight 27 kg
Dimensions (height × width × depth) mfr 600 mm
Enclosure mfr All materials are recyclable recyclable
path length (multicell) mfr 0.2...10 mm
path length (fiber optic tips) mfr 1...20 mm
Electrical, Power & Drive Specifications
Mains voltage mfr 115
Line frequency mfr 60 Hz
ISM classification mfr 1 A
Sampling frequency mfr 2 minutes
Power consumption mfr 8 A
Interfaces, Connections & Control
Display mfr color touchscreen touchscreen
Local user interface mfr 7 inch
Interface utilisateur locale mfr 7 pouces
Materials, Media & Process Compatibility
Media types mfr Compatible with all dissolution media containing
Media replacement mfr 100
Media Change (changement du milieu) mfr Lorsque cette option est activée, elle permet d’
Media Volume (volume de milieu) mfr Volume de milieu présent initialement dans les c
Operating Conditions, Environment & Protection
Operating Altitude 2000 m
Photometric noise mfr 0.0001 Abs
Humidity mfr 80 % RH

Description

Construction and series positioning

The 850-DS Sampling Station (model G7930A) is a standalone benchtop unit measuring 400 × 390 × 600 mm and weighing 27 kg, with a color touchscreen and an integrated 8-syringe rotary piston pump. Within the Dissolution Autosamplers series it is the base sampling station; the 850-DS Filtration Module (G7931A) is a separate option that adds automated filtration, increases height to 600 mm and width to 480 mm with the waste tray, and draws 8 A versus 2 A for the base unit. The base station handles sampling, media replacement and cleaning; filtration is enabled only when the G7931A module is installed.

Fluid handling and method behavior

Sample volume per vial or tube is 0.1–14 mL, doubling to 28 mL with dual sample, and sampling accuracy is 10 mL ± 2.5 %. The syringe pump runs at 6–12 mL/min, so timepoints cannot be closer than 2 minutes. Prime and purge volumes are 0–30 mL; cleaning volume is 0.1–15 mL for up to 9 cycles via the rinse port or sampling cannulas. At each timepoint the volume removed from each vessel equals sample volume + waste drop volume + a fixed 0.2 mL needle purge, and media replacement must be programmed to include that purge. Up to 36 timepoints are allowed per method, numbered consecutively, with a maximum run duration of 999 hours; beyond 36 samples the collection tray must be changed manually.

Installation, control and operation

Three 1/4-28 male to 1/8" barbed fittings on the left rear connect waste, media replacement and rinse containers via 1/4" silicone tubing; up to 8 sampling and 8 return lines run to the dissolution apparatus, and unused ports must be fitted with barbed fittings routed to waste. Media containers must sit at or below the 850-DS, and the waste inlet tube must never be below the waste liquid level. The unit accepts 115/230 V~ ±10 %, 60/50 Hz, 2 A through a country-specific Agilent power cord into a grounded outlet; there is no voltage selector or accessible fuse. Methods are stored in 40 memory locations, and the station can control one Agilent/Varian/VanKel dissolution apparatus or run synchronized with a second 850-DS using a 9-pin null modem cable (5075-0449) on the AUX port.

Integration & Operating Notes

Process Limits
Keep timepoints at least 2 minutes apart and within 36 per method

The syringe pump flow range of 6–12 mL/min and the minimum 2-minute sampling frequency set the shortest programmable interval. Methods allow up to 36 timepoints, numbered consecutively, and entering zero for the time deletes a timepoint. Runs can last up to 999 hours, but more than 36 samples requires a manual tray change.

850-DS Sampling Station User Manual, Timepoints, page 41; Performance data, page 16
Control & I/O
Enable Apparatus Control only for Agilent/Varian/VanKel apparatus

Dissolution Apparatus Control must be enabled only when the 850-DS is connected to another Agilent, Varian or VanKel dissolution apparatus. It must not be checked for a third-party apparatus with fixed probes, because the station would then issue control commands that conflict with the apparatus's own controller.

850-DS Sampling Station User Manual, Run Method, page 45
Control & I/O
Synchronized Start needs identical methods and a null modem cable

Two-collector synchronized operation requires the same method loaded in the same memory location on both 850-DS units and a 9-pin to 9-pin null modem cable (5075-0449) connected via the AUX port. Time Delayed start cannot be combined with Synchronized Start, so the start option must be chosen accordingly.

850-DS Sampling Station User Manual, Run Method, page 46
Maintenance
Run a validated cleaning method after every dissolution test

A validated cleaning method must be run after each dissolution test, using 0.1–15 mL cleaning volume for up to 9 cycles through the rinse port or the sampling cannulas. The main screen shows a maintenance-required notification and leak alarm; these must be addressed before the next run. Filter plates, where the G7931A module is fitted, are single-use with a maximum of 5 uses and a warning above 2 uses per filter.

850-DS Sampling Station User Manual, Sample Collection Parameters and Filtration Parameters, pages 38 and 40
Process Limits
Place media containers at or below instrument level

Media containers must be at the same level as or below the 850-DS, and the waste inlet tube must never be below the waste liquid level. Otherwise liquid can siphon back into the instrument. Up to 8 sampling and 8 return lines connect to the apparatus, and unused positions must be fitted with barbed fittings routed to waste.

850-DS Sampling Station User Manual, Tube Connections, page 25

Selection & Integration Q&A

When should I choose the 850-DS G7930A base station instead of adding the G7931A filtration module?

Choose the base G7930A when the method does not require automated filtration before collection; it draws 2 A and stands 400 mm high. Add the G7931A module when filtered samples are needed: it raises height to 600 mm, width to 480 mm with the waste tray, weighs 29 kg and draws 8 A, and must be installed by a certified Agilent representative.

What must be in place before connecting the 850-DS to a dissolution apparatus and a UV or HPLC collection path?

The station connects to the 708-DS, Apparatus 3, Apparatus 7, NanoDis or Multicell UV system, with up to 8 sampling and 8 return lines using 1/4-28 male to 1/8" barbed fittings and 1/4" silicone tubing. Apparatus Control may be enabled only for Agilent/Varian/VanKel apparatus. Collection trays for UV-Vis or HPLC are separate options, and the station can be driven by Dissolution Workstation Software or OpenLab Dissolution Workflow Manager.

What maintenance and consumables are required beyond the base delivery?

A validated cleaning method must be run after each test using 0.1–15 mL for up to 9 cycles. Consumables not included are cannula filters (1, 10, 35, 70 μm) and, with the G7931A module, single-use filter plates (0.22 or 0.45 μm, max 5 uses, warning above 2). A 9-pin null modem cable (5075-0449) is needed for synchronized two-unit operation, and country-specific power cables are ordered separately.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
850-DS Dissolution Sampling Station User Manual
Manual ·101 pp.
View PDF
PDF
Agilent Online UV-Vis Dissolution
Brochure ·5,3 МБ ·12 pp.
View PDF