Agilent Technologies Maintenance free, JetClean ion source for GC/MS instrument
Agilent Technologies Maintenance free, JetClean ion source for GC/MS instrument — Gas chromatography mass spectrometry gc ms.
The Agilent JetClean is a self-cleaning ion source accessory for Agilent single- and triple-quadrupole GC/MS systems, operating in Acquire & Clean or Clean Only mode. It extends manual source cleaning from every two weeks to once in nine months by bleeding hydrogen through the source at 0.13–3.52 mL/min (Clean Only) or 0.13–0.49 mL/min (Acquire & Clean).
Agilent Technologies GC MS Ion Sources Inlets Series — Model Comparison & Selection
| Model | Detection limit | Clean duration lower setpoint, min | Clean duration upper setpoint, min | Emission current lower setpoint, microampere | Emission current upper setpoint, microampere | Hydrogen flow acquire clean mode lower setpoint, mL/min |
|---|---|---|---|---|---|---|
| Maintenance free, JetClean ion source for GC/MS instrument Current | — | 1 | 120 | 10 | 35 | 0.13 |
| Ion Sources for Gas Chromatography/Mass Spectrometry | attogram | 1 | 120 | 10 | 35 | 0.13 |
Full specifications
| Clean Duration, Lower Setpoint | 1 min |
| Clean Duration, Upper Setpoint | 120 min |
| Hydrogen Flow (acquire & Clean Mode), Lower Setpoint | 0.13 mL/min |
| Hydrogen Flow (acquire & Clean Mode), Upper Setpoint | 0.49 mL/min |
| Hydrogen Flow (clean Only Mode), Lower Setpoint | 0.13 mL/min |
| Hydrogen Flow (clean Only Mode), Upper Setpoint | 3.52 mL/min |
| Cleaning gas mfr | hydrogen hydrogen |
| Manual source cleaning interval without JetClean mfr | once every two weeks weeks |
| Manual source cleaning interval with JetClean mfr | once in nine months months |
| Area %RSD, Benzo[a]anthracene mfr | 1.8 %RSD |
| Area %RSD, Chrysene mfr | 1.2 %RSD |
| Area %RSD, Benzo[b]fluoranthene mfr | 1.9 %RSD |
| Area %RSD, Benzo[a]pyrene mfr | 1.7 %RSD |
| Calibration linearity R2, Benzo[a]pyrene mfr | 0.9976 |
| MRM ratio, 252.1 -> 250.1 / 252.1 -> 252.1 mfr | 56.2 |
| MRM ratio, 250.1 -> 248.1 / 252.1 -> 252.1 mfr | 8.7 |
| Hydrogen flow step size mfr | 0.069 sccm |
| Emission Current, Lower Setpoint | 10 µA |
| Emission Current, Upper Setpoint | 35 µA |
| Source Temperature, Lower Setpoint | 150 °C |
| Source Temperature, Upper Setpoint | 350 °C |
Description
Construction and series positioning
- The JetClean is an ion-source modification, not a standalone instrument: it is factory-fitted on new Agilent 5977A/B, 7000B/C/D, 7010A/B and Intuvo-equipped 5977B/7000D systems, and is available as a retrofit to existing systems.
- Within the GC/MS line it sits alongside the 5977C single quad, 7000E/7010D triple quads and 7250 Q-TOF; it is the source-cleaning option rather than a different mass analyzer.
- Cleaning gas is hydrogen, supplied through dedicated plumbing described in the operating manual (G7077-90052).
Operating modes and behavior
- Two modes: Acquire & Clean (cleaning while acquiring data) and Clean Only (dedicated cleaning cycle).
- Hydrogen flow setpoints differ by mode: 0.13–0.49 mL/min in Acquire & Clean, 0.13–3.52 mL/min in Clean Only; flow step size is 0.069 sccm.
- Clean duration is settable from 1 to 120 min; emission current from 10 to 35 µA; source temperature from 150 to 350 °C.
- Reported method performance: area %RSD 2.8–4.9% for benzo[a]anthracene, chrysene, benzo[b]fluoranthene and benzo[a]pyrene; calibration linearity R² 0.9972 for benzo[a]pyrene.
Operation and maintenance effect
- Without JetClean, manual source cleaning is typically required every two weeks; with JetClean the interval extends to once in nine months.
- Configuration, parameter setpoints, hydrogen supply plumbing and troubleshooting are covered in the operating manual.
Integration & Operating Notes
Acquire & Clean mode is limited to 0.13–0.49 mL/min, while Clean Only permits up to 3.52 mL/min. Setting a Clean Only-level flow while acquiring data exceeds the Acquire & Clean envelope; verify the mode before entering a flow setpoint.
JetClean extends the manual cleaning interval from roughly every two weeks to once in nine months; it does not eliminate manual cleaning. Plan the nine-month service event into the instrument schedule.
The JetClean is offered as an option on new 5977A/B, 7000B/C/D, 7010A/B and Intuvo-equipped systems, and as a retrofit to existing systems. Confirm the host MS model and Intuvo configuration before specifying a retrofit.
Selection & Integration Q&A
Choose JetClean when sample load would otherwise force manual source cleaning about every two weeks; it extends the interval to once in nine months. If source cleanliness is not the bottleneck, the standard source on the 5977C, 7000E or 7010D needs no hydrogen plumbing or cleaning-gas supply.
It needs a hydrogen supply connected through the dedicated plumbing described in the operating manual, and a compatible host: 5977A/B, 7000B/C/D, 7010A/B, or Intuvo-equipped 5977B/7000D. Hydrogen flow is set per mode, 0.13–0.49 mL/min in Acquire & Clean and up to 3.52 mL/min in Clean Only.
Plan a manual source cleaning about once every nine months even with JetClean, versus roughly every two weeks without it. The hydrogen supply and its plumbing are system-side items to be provided; the operating manual covers configuration, setpoints and troubleshooting.