Agilent Technologies High-sensitivity metabolic analyzer, Seahorse XF HS Mini
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies High-sensitivity metabolic analyzer, Seahorse XF HS Mini

Agilent Technologies High-sensitivity metabolic analyzer, Seahorse XF HS Mini — Cell analysis.

The Agilent Seahorse XF HS Mini is an 8-well high-sensitivity metabolic analyzer that measures oxygen consumption rate (OCR) and extracellular acidification rate (ECAR) in live cells. It uses XF HS miniplates with a microchamber area of 11.40 and a sensor positioned 200 µm above the well bottom, enabling 3-minute measurement intervals. This modification is distinguished within the Xf Analyzers series by its high-sensitivity microplate format, suited for samples with limited cell numbers.

Agilent Technologies Xf Analyzers Series — Model Comparison & Selection

Model Measurement rangeOperating system
High-sensitivity metabolic analyzer, Seahorse XF HS Mini Current 25...228—
Cell imaging and normalization, Seahorse normalization —10

Full specifications

Core Performance & Functional Specifications
Measurement range 25...228 pmol/min
Microchamber area (XFp) mfr 11.40
Sensor position above well bottom mfr 200 μm
Measurement duration mfr 3 minutes
OCR range mfr 7...55 pmol/min
ECAR range mfr 10...90 mpH/min
Well count mfr 8 wells
Cell seeding volume mfr 30 μL
Cell seeding density (suspension cells) mfr 10.0 cells/well
Cell seeding density (naive CD4 T cells) mfr 70 cells/well

Description

Construction and series positioning

Measurement behavior and ranges

Installation and operation

Integration & Operating Notes

Process Limits
Use only XF HS miniplates for valid results

The analyzer is designed for XF HS miniplates (103724-100) or pre-assembled XF HS miniplates (103723-100). Other plate types, such as XFp miniplates, have different microchamber geometry and will not yield accurate OCR/ECAR readings. Verify plate part number before loading.

Technical Overview, Agilent Seahorse XF HS miniplate
Process Limits
Seeding density must match cell type for sensitivity

For suspension cells, seed 10.0 cells/well; for naive CD4 T cells, seed 70 cells/well. Exceeding these densities may saturate the sensor and invalidate rate measurements, while lower densities may fall below the detection limit of 7 pmol/min OCR.

Technical Overview, Agilent Seahorse XF HS miniplate
Maintenance
Calibrate sensor cartridge before each assay

The sensor cartridge must be hydrated and calibrated according to the instrument manual prior to running plates. Skipping calibration leads to drift and inaccurate ECAR/OCR values. No automatic calibration is described in the provided facts.

Technical Overview, Agilent Seahorse XF HS miniplate

Selection & Integration Q&A

When should I choose the XF HS Mini over other Xf Analyzers siblings?

Choose the XF HS Mini when working with limited cell numbers (e.g., 10–70 cells/well) and requiring high-sensitivity OCR/ECAR measurements. For standard sensitivity or normalization-focused workflows, consider the sibling normalization model.

What plate types and accessories are required for the XF HS Mini, and are they included?

The instrument requires XF HS miniplates (103724-100) or pre-assembled XF HS miniplates (103723-100); PDL-coated variants are available for suspension cells. These consumables are not included with the base unit and must be ordered separately.

What are the environmental or duty-cycle limits for operating the XF HS Mini?

Verified facts do not specify environmental or duty-cycle limits. Refer to the instrument manual for operating temperature, humidity, and maximum assay frequency. The 3-minute measurement duration suggests rapid cycling is possible, but continuous operation limits are not documented here.

What maintenance schedule and consumables are recommended for the XF HS Mini?

Regular maintenance includes sensor cartridge calibration before each assay and proper hydration. Consumables such as XF HS miniplates and calibration solutions are not included in base delivery and must be procured separately. No specific service interval is provided in the verified facts.