Agilent Technologies 1290 Infinity II LC System

Agilent Technologies 1290 Infinity II LC System — Liquid chromatography.

The 1290 Infinity II LC System is a stackable UHPLC platform built around the G7120A High-Speed Pump (1300 bar, 5 mL/min, 10 µL delay volume) and a choice of Multisampler, Hybrid Multisampler, or Vialsampler. It is distinguished from the Flexible Pump variant by its binary high-pressure gradient and lack of a 4CGV, suiting narrow-bore sub-2 µm separations. Primary purpose: high-resolution, high-throughput liquid chromatography in analytical laboratories.

Full specifications

Core Performance & Functional Specifications
Analytical Flow Rate 5 mL/min
Baseline Drift 1 AU/h
Carryover 9 ppm
Data Rate 240 Hz
Maximum Individual Bottle Volume 2 L
Maximum pressure 1300 bar
Maximum System Pressure 1300 bar
Sample Capacity 6144 samples
Solvent Cabinet Maximum Volume 8 L
Delay volume mfr 10 μl
Gradient composition range mfr 1...99
Number of columns mfr eight columns columns
Sensitivity gain vs conventional flow cells mfr 10 times
Inline filter nominal pore size mfr 0.3 μL
Purge flow rate mfr 2.5...3 mL/min
Seal wash solvent mfr 10 water
Default IP address mfr 192.168.254.11
Delay volume (High-Speed Pump) mfr 10 μl
Gradient composition range (High-Speed Pump) mfr 1...99
Number of columns in MCT mfr eight columns columns
Sensitivity gain vs conventional flow cells (DAD) mfr 10 times
Red mfr 0.12 mm
Green mfr 0.17 mm
Blue mfr 0.25 mm
Orange mfr 0.50 mm
Double mfr 7116 A/B
Single mfr 7114 A/B
Dimensions, Weight & Mechanical Design
Flow Cell Path Length 60 mm
Capillary inner diameter mfr 0.12 mm
Capillary inner diameter (red) mfr 0.12 mm
Capillary volume per length (red) mfr 0.113 cm
Capillary inner diameter (green) mfr 0.17 mm
Capillary volume per length (green) mfr 0.227 cm
Capillary inner diameter (blue) mfr 0.25 mm
Capillary volume per length (blue) mfr 0.491 cm
Capillary inner diameter (orange) mfr 0.50 mm
Operating Conditions, Environment & Protection
Column Compartment Temperature 80 °C
Column thermostatting range mfr 20...110 °C
Detector noise mfr 0.6 μAU/cm
Column thermostatting cooling mfr 20 °C
Column thermostatting heating mfr 110 °C
Detector noise (DAD) mfr 0.6 μAU/cm
Detector noise (VWD) mfr 1.5 μAU

Description

Architecture and series positioning

Behavior under load and separation performance

Installation, connection, and control

Integration & Operating Notes

Process Limits
DMF mobile phase corrodes the PVDF leak sensor

Dimethylformamide is incompatible with the PVDF leak sensor material and leads to corrosion. Do not use DMF as mobile phase; check the leak sensor regularly for corrosion and do not place metal parts on the leak panel or sensor.

System Manual, page 41 Leak Sensor
Electrical
LAN cable must land on the highest-data-rate module

LAN configuration is executed from the module with direct LAN connection to the controller software, which must be the module with the highest data rate, usually the detector. Rear DIP switches set addressing: all DOWN = default IP 192.168.254.11; switch 4 UP = DHCP; switch 5 UP = STORED address. MAC address is usable only in DHCP mode.

System Manual, pages 43-44 LAN Configuration
Process Limits
Solvent cabinet limited to 8 L total, 2 L per bottle

Plan mobile phase volumes and bottle sizes to stay within the 8 L total cabinet capacity and 2 L per-bottle limit. Use only HPLC-grade solvents and water filtered through 0.2 µm filters; exchange water-based solvents daily and use solvent volume within 1–2 days.

System Manual, pages 32 and 51
Maintenance
Purge each channel at 2.5–3 mL/min for 5 min before runs

The Purge function fills the pump, exchanges solvent, and removes air bubbles. Purge each channel with fresh solvent at 2.5–3 mL/min for 5 min before operation, then equilibrate the system. Conditioning is used daily when starting the pump and to minimize pressure ripple.

System Manual, page 49 Best Practices
Maintenance
Replace seal wash solvent weekly; do not recycle it

Change seal wash solvent (10% isopropanol in water) and bottle weekly; do not recycle seal wash solvent. Seal Wash extends seal lifetime when using buffers with elevated salt concentrations or volatile solvents with non-volatile additives.

System Manual, pages 49-51
Maintenance
Back-flush the optional inline filter weekly

The optional inline filter (5067-5407, stainless steel) has a nominal filter pore size of 0.3 µL, very small internal volume, is specified for high pressures, and can be back-flushed. It is recommended to protect the downstream system from blockages, for solvent combinations that can form precipitation, and for buffer applications; back-flush weekly.

System Manual, pages 51-52 Optional Inline Filter
Mounting
Drain connectors excluded from the bottom module

Install the Double Drain Connector (5043-1836) on the Multicolumn Thermostat and Single Drain Connectors (5043-1834) on other modules, excluding the bottom module, which must be connected via waste tubing. Drain connectors improve leak drainage for low-flow leaks of high-viscosity solvents.

System Manual, pages 36-39 Drain Connectors Installation
Process Limits
Waste tubes must not immerse in liquid

Agilent recommends the 6 L waste can with 1 Stay Safe cap GL45 with 4 ports (5043-1221) for safe waste disposal. If using a custom waste solution, tubes must not immerse in the liquid, and waste flow must not be restricted at bends or joints.

System Manual, page 40 Waste Concept

Selection & Integration Q&A

When should I choose the High-Speed Pump configuration of the 1290 Infinity II over the Flexible Pump variant?

Pick the High-Speed Pump (G7120A) for binary high-pressure gradients with narrow-bore columns and sub-2 µm particles: 1300 bar, 5 mL/min, 10 µL delay volume, 1–99% gradient. The Flexible Pump (G7104A) adds the 4CGV gradient valve for low-pressure mixing and buffer channels; the 4CGV is not available on the High-Speed Pump.

What are the requirements for connecting the system to the control software and network?

Connect the LAN cable to the module with the highest data rate, usually the detector, and run LAN configuration from that module. Set rear DIP switches: all DOWN = 192.168.254.11; switch 4 UP = DHCP; switch 5 UP = STORED. MAC address is usable only in DHCP mode. Add the system in Lab Advisor by name, address, and type.

What mounting and environment limits apply to the stack configuration?

The system can be installed as a single stack of maximum four modules, two stacks, an InfinityLab Flex Bench, or an InfinityLab Benchtop, chosen to minimize delay volume. Install drain connectors on all modules except the bottom one, which uses waste tubing. The leak sensor stops the system when the detection level is reached, so ensure leak plane outlets and waste tubes are unrestricted.

What maintenance schedule and accessories are required beyond the base delivery?

Daily: replace water/buffer solvents, check seal wash, purge each channel at 2.5–3 mL/min for 5 min, equilibrate. Weekly: change seal wash solvent (10% isopropanol in water) and inspect solvent filters; back-flush the optional inline filter (5067-5407). Recommended accessories not in base delivery: 6 L waste can 5043-1221, drain connectors 5043-1836/5043-1834, and the HPLC System Tool Kit.