Agilent Technologies 280Z AA
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies 280Z AA

Agilent Technologies 280Z AA — Atomic spectroscopy.

The Agilent 280Z AA is a flame atomic absorption spectrometer with a Zeeman furnace configuration, part of the 240/280 series. It is distinguished by its flame safety envelope: acetylene supply 65–100 kPa, exhaust ≥6 m³/min, and UV Risk Group 3 exposure limits. It is used for trace metal determination in flame and furnace modes.

Full specifications

Core Performance & Functional Specifications
Acetylene Purity 99.5 %
Direct Viewing Distance From Flame 0.2 m
Exempt Distance From Flame 7.1 m
Maximum Flame Door Open Exposure 13 minutes
Viewing Distance Rg1 2.3 m
Viewing Distance Rg2 0.8 m
Exhaust ventilation rate mfr 6 minute
Exhaust fan distance from flame mfr 3 meters
UV intensity emitted mfr 37.7
UV hazard classification mfr 3
Burner slot replacement limit, nitrous-oxide/acetylene mfr 0.47 mm
Burner slot replacement limit, air/acetylene mfr 0.54 mm
Acetylene maximum pressure mfr 105 kPa
Acetylene fuel supply pressure range mfr 65 kPa
Acetylene cylinder minimum pressure before replacement mfr 700 kPa
Acetylene maximum consumption rate mfr 1 hour
Solvent minimum specific gravity mfr 0.75
Acetylene fuel supply pressure mfr 65...100 kPa
Acetylene cylinder replacement pressure mfr 700 kPa
Burner slot replacement threshold mfr 0.54 mm
Minimum solvent specific gravity mfr 0.75
Schedule mfr 62
Cleaning mfr 63
Gas cylinders mfr 64
Preventing flashbacks mfr 65
Spare parts mfr 65
Ultraviolet Hazard mfr 3
Acetylene cylinder minimum pressure mfr 700 kPa
Acetylene maximum consumption mfr 1 hour
Zeeman magnetic field peak mfr 0.8 Tesla
Zeeman magnet exclusion distance mfr 300 mm
UV intensity mfr 37.7
Hollow cathode lamp capacity 240 series mfr Four lamp capacity capacity
Hollow cathode lamp capacity 280 series mfr Eight lamp capacity capacity
Dimensions, Weight & Mechanical Design
Burner slot width maximum, N2O/Acetylene Mark 7 mfr 0.46 mm
Burner slot width maximum, Air/Acetylene Mark 7 mfr 0.54 mm
Burner slot maximum width mfr 0.54 mm
Nitrous-oxide/acetylene burner slot maximum width mfr 0.47 mm
Air/acetylene burner slot maximum width mfr 0.54 mm
Mark 7 N2O/acetylene burner slot maximum width mfr 0.46 mm
Mark 7 air/acetylene burner slot maximum width mfr 0.54 mm
Electrical, Power & Drive Specifications
Non-Zeeman 240/280FS fuse rating mfr 4 mm
Zeeman circuit breaker rating mfr 15 breaker
Zeeman fuse T3.15 A rating mfr 3.15 mm
Zeeman fuse F315 mA rating mfr 315 mm
Zeeman fuse F1A rating mfr 1 mm
Zeeman fuse T6.3 A rating mfr 6.3
Zeeman fuse F800 mA rating mfr 800 mm

Description

Construction and series positioning

The 280Z AA belongs to the Agilent 240/280 AA series, which includes flame-only, furnace (Zeeman or D2), and Duo configurations. This model is a Zeeman furnace variant, meaning it combines flame operation with a Zeeman-effect graphite furnace for trace analysis. The manual covers the whole series, so safety and maintenance facts apply broadly; the 280Z AA specifically includes the Zeeman furnace module. No sibling models are listed for direct comparison, but the series spans flame and furnace capabilities.

Behavior under load and media

Flame operation requires acetylene fuel at 65–100 kPa (never exceeding 105 kPa) and an oxidant of air or nitrous oxide (never oxygen). The burner slot width must be monitored: maximum 0.46 mm for N2O/acetylene and 0.54 mm for air/acetylene; exceeding these limits requires burner replacement. The liquid trap must be filled with the same solvent as samples, and solvents with specific gravity below 0.75 must not be used, as they can breach the trap seal and cause flashback. UV emission is 37.7 mW/m², classified Risk Group 3 under IEC 62471, so direct viewing at ≤0.2 m is hazardous; with the flame door open, exposure must not exceed 13 minutes per 8-hour period.

Installation, connection, and control

The instrument requires an exhaust system providing at least 6 m³/min, with the exhaust fan located at least 3 meters from the flame. Gas connections use left-hand threads for fuel gases and right-hand threads for support gases to prevent cross-connection. Acetylene cylinders must be replaced when pressure falls below 700 kPa, and consumption must stay below 1/7 of cylinder contents per hour to avoid acetone carryover. The flame door interlock prevents ignition if the door is opened during ignition; bypassing is prohibited. Daily leak tests at all joints with soapy water are mandatory. The manual also covers SpectrAA software (version 5.6.8 or higher) and USB-GPIB communications for control and data acquisition.

Integration & Operating Notes

Process Limits
Acetylene pressure must never exceed 105 kPa

Supply pressure must be regulated to 65–100 kPa; pressures above 105 kPa risk spontaneous acetylene decomposition. The regulator must be set within this range and verified before each use, as higher pressures can cause explosion.

Safety Practices and Hazards, p.19
Process Limits
Burner slot width must be checked against replacement limits

For N2O/acetylene Mark 7 burners, slot width must not exceed 0.46 mm; for air/acetylene, 0.54 mm. If exceeded, the burner must be replaced. Regular inspection and cleaning are critical to prevent flashback.

Safety Practices and Hazards, p.25 and p.21
Process Limits
Solvent specific gravity must be at least 0.75

Using a solvent with specific gravity below 0.75 can breach the liquid trap seal, leading to flashback. The liquid trap must be filled with the same solvent as the samples, and the drain tube free end must remain above the waste vessel liquid level.

Safety Practices and Hazards, p.17 and p.22
Electrical
Verify operating voltage before connecting to mains

The instrument must not be connected to mains until the operating voltage is verified for the specific outlet. Use only the Agilent-supplied power cord or an equivalent specified in the AA Site Preparation Guide to ensure proper grounding and electrical safety.

Safety Practices and Hazards, p.9
Control & I/O
Flame door interlock prevents ignition when door is open

If the flame door is opened during ignition, the interlock prevents ignition. If the interlock is lost or malfunctioning, contact Agilent immediately; bypassing the interlock is prohibited as it is a critical safety feature.

Safety Practices and Hazards, p.14
Maintenance
Perform daily leak tests at all gas joints

Before each use, test all joints and seals for leaks using soapy water or a proprietary solution; never use a naked flame. This daily check is mandatory to prevent gas accumulation and fire hazards.

Safety Practices and Hazards, p.12
Maintenance
Weekly inspect O-rings in burner and spray chamber

Check three O-rings inside the burner, spray chamber, and bung, and two sealing gas inlets to the spray chamber. Damaged O-rings can cause gas leakage, fire, or flashback; replace immediately if worn or damaged.

Maintenance and Troubleshooting, p.62-63
Cooling & Media
Nitrous oxide regulator may require heating to prevent freezing

When using nitrous oxide, decompression cooling can freeze the regulator. Use an in-line or wrap-around heater to prevent regulator malfunction and possible flashback.

Safety Practices and Hazards, p.20
Mounting
Exhaust fan must be at least 3 meters from the flame

The exhaust system must provide at least 6 m³/min and the fan must be located at least 3 meters from the flame to avoid drawing flame or fumes improperly. The instrument must be vented to outside air via a self-contained collector hood and ducting; never vent within the building.

Safety Practices and Hazards, p.10

Selection & Integration Q&A

When should I choose the 280Z AA over a flame-only or D2 furnace variant in the 240/280 series?

Choose the 280Z AA when you need both flame AA for higher concentrations and Zeeman furnace AA for trace levels with background correction. A flame-only variant suffices for routine flame work; a D2 furnace variant may be chosen if Zeeman correction is not required.

What are the requirements for the acetylene supply and gas connections to the 280Z AA?

Use instrument-grade acetylene (≥99.5% purity) in acetone, with supply pressure 65–100 kPa (never >105 kPa). Fuel gas fittings use left-hand threads; support gas fittings use right-hand threads. Replace cylinders below 700 kPa and keep consumption <1/7 of cylinder per hour.

What are the laboratory ventilation and UV exposure limits for operating the 280Z AA?

Exhaust must provide ≥6 m³/min with fan ≥3 m from flame. UV emission is 37.7 mW/m² (Risk Group 3); direct viewing at ≤0.2 m is hazardous. With flame door open, limit exposure to 13 minutes per 8-hour period and wear UV-protective eyewear.

What maintenance items and accessories are required but not included with the base instrument?

The base delivery does not include gas regulators, exhaust system, or UV-protective eyewear. Daily leak tests and weekly O-ring inspections are mandatory. Spare parts must be Agilent-supplied; order via Agilent website. Burners and O-rings are consumables requiring periodic replacement.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Agilent AA Spectrometer User's Guide
Manual ·68 pp.
View PDF