Agilent Technologies 5110 ICP OES
Agilent Technologies 5110 ICP OES — Atomic spectroscopy.
The Agilent 5110 ICP-OES is a dual-view atomic emission spectrometer configured with a fully demountable Easy-fit torch. Its 1.8 mm injector and standard quartz outer tube-set suit aqueous and acidic samples, while the ceramic/alumina option extends TDS tolerance to 35%. It serves as the excitation and sample-introduction core for trace-element analysis in process and environmental laboratories.
Full specifications
| Injector Internal Diameter (id) | 1.8 mm |
| Maximum Total Dissolved Solids (tds) | 35 % |
| Most aqueous or acidic samples mfr | Standard quartz quartz |
| Volatile and semi-volatile organic samples mfr | 'Organic' high purity quartz quartz |
| injector inner diameter mfr | 2.4 mm |
| High-TDS matrices rich in alkali metals (including Li battery materials), hydrofluoric (HF) acid digests or fusions mfr | Ceramic outer outer |
| Cooling time before torch removal mfr | at least five minutes minutes |
| cooling time before torch removal mfr | cinque minuti minuti |
Description
Construction and series positioning
The 5110 is the base model of the 5100/5110 ICP-OES platform, distinguished from the 5800/5900 by its torch and sample-introduction configuration rather than by a different excitation source. The fully demountable Easy-fit torch is a five-part assembly: locking nut, injector, torch body, top seal, and outer tube-set. This modification is supplied with the standard quartz outer tube-set and a 1.8 mm injector, the general-purpose configuration for aqueous and acidic digests. The torch can be operated in either Radial View (RV) or Dual View (DV) mode, but only with the matching outer tube-set: the RV tube-set is shorter and lacks the top slot found on the DV version.
Behavior across sample matrices
The injector/outer-tube combination defines the working envelope. The 1.8 mm quartz injector handles most aqueous and acidic samples economically. Organic matrices require the organic high-purity quartz outer tube-set with a 0.8 mm (volatile) or 1.4 mm (semi-volatile) injector to resist solarization. High-TDS matrices, HF digests, and fusions call for the ceramic outer tube-set with a 2.4 mm inert alumina injector, which is rated to 35% total dissolved solids and offers the longest service life and resistance to devitrification. The ceramic kit does not include an injector; it must be ordered separately.
Installation, control, and operation
The torch ships without the injector installed. Before first use, the shipping plug is removed, the injector is inserted until it clicks into the retaining clamp, and the locking nut is hand-tightened only. During reassembly, the groove on the torch body must align with the 'T' mark on the outer tube-set, and the tube-set is pressed fully into the body. The quartz intermediate tube of the ceramic outer tube-set carries a fixed polymer adapter that must not be removed; the supplied tube-set tool aids assembly and disassembly. After plasma switch-off, the torch and compartment must cool for at least five minutes before removal, and all insertions and withdrawals must follow a straight, parallel path to avoid breaking quartz or ceramic components. Maintenance and cleaning procedures are referenced to publication 5994-8864EN.
Integration & Operating Notes
The G8020-47004 inert alumina injector is rated for high-TDS samples up to 35% total dissolved solids. Exceeding this limit risks salt deposition, injector blockage, and signal drift; dilute or pre-treat samples above this threshold before analysis.
The torch and torch compartment remain hot after the plasma is extinguished. Allow at least five minutes of cooling before attempting removal to prevent burns and thermal stress on the quartz or ceramic components.
Any angular force during insertion or removal can crack the quartz outer tube or the ceramic tube-set. Keep the torch axis aligned with the compartment guide throughout the stroke, and never lever the torch body against the compartment walls.
The locking nut (G8020-60810) secures the injector against the torch body. Using tools or excessive force can crush the injector ferrule and crack the torch body; tighten by hand until the injector is seated and the nut is snug.
The RV outer tube-set is shorter and lacks the top slot of the DV version. Fitting the wrong tube-set prevents correct seating and plasma ignition; confirm the observation mode (RV or DV) and select the corresponding outer tube-set part number before assembly.
Selection & Integration Q&A
Choose the standard quartz outer tube-set with 1.8 mm injector for routine aqueous and acidic digests; it is the most economical option. Switch to organic high-purity quartz with 0.8 or 1.4 mm injectors for organic matrices, or to the ceramic/2.4 mm alumina set for HF digests, fusions, and high-TDS samples up to 35%.
The fully demountable torch kit with ceramic outer tube-set does not include an injector. A compatible quartz or inert alumina injector must be ordered separately; for high-TDS and HF work, the 2.4 mm inert alumina injector G8020-47004 is the matching choice.
The torch must be inserted with a straight, parallel movement to avoid breaking quartz or ceramic components. The injector is inserted until it clicks into the retaining clamp, the locking nut is hand-tightened only, and the outer tube-set groove must align with the 'T' mark on the torch body.
Cleaning follows publication 5994-8864EN. Spares include top seal G8014-60023 (pack of 3), injector locking nut G8020-60810, and replacement injectors in quartz or inert alumina from 0.8 to 2.4 mm id. The base torch ships without the injector installed; the ceramic kit omits the injector entirely, and the tube-set tool is supplied only with the ceramic set.