Agilent Technologies Cary 660 FTIR
Agilent Technologies Cary 660 FTIR — Molecular spectroscopy with Flow rate 10 liter/minute.
The Cary 660 is a mid-series FTIR spectrometer in the Cary 600 Series, sharing the penthouse architecture with the 670 and 680 while the 610 microscope is a separate accessory. It uses a 632.8 nm continuous laser (<1 mW embedded, <600 μW accessible) and requires dry purge gas at 10 L/min. It serves as the optical engine for routine molecular spectroscopy, with MCT/InSb detectors needing liquid-nitrogen cooling.
Full specifications
| Flow rate | 10 L/min |
| Wavelength range | 632.8 nm |
| Dewar Equilibrium Time | 20 minutes |
| Maximum Accessible Laser Power | 600 μW |
| Maximum Beam Divergence (embedded Laser) | 1.3 milliradians |
| Minimum Side Clearance | 50 mm |
| Warm-up Time To Operating Temperature | 30 minutes |
| Wavelength (embedded Laser) | 632.8 nm |
| Regulator required above source pressure mfr | 60 psi |
| Laser pulse duration mfr | Continuous Continuous |
| Pulse duration (embedded laser) mfr | Continuous Continuous |
| Initial Dewar fill volume mfr | 500 milliliters |
| Dewar top-off volume mfr | 200 milliliters |
| Minimum rear clearance mfr | 150 mm |
| Purge gas source pressure regulator threshold mfr | 60 psi |
| Pulse duration mfr | Continuous Continuous |
| Agilent Cary 600 Series FTIR spectrometer mfr | 4 equivalent |
| Agilent Cary 610 FTIR microscope (typical) mfr | 1 equivalent |
| High-sensitivity MCT mfr | Liquid nitrogen nitrogen |
| Linearized high-sensitivity MCT mfr | Liquid nitrogen nitrogen |
| Linearized broadband MCT mfr | Liquid nitrogen nitrogen |
| Cooled DLaTGS mfr | Peltier Peltier |
| Top — Power/Scan mfr | Flashing green green |
| Bottom — A/D over-range mfr | Flashing blue blue |
| Recommended purge flow rate for spectrometer and microscope mfr | 10 L/min |
| Regulator threshold pressure mfr | 60 psi |
| Dewar fill volume mfr | 475 milliliters |
| Dewar top-up volume mfr | 200 milliliters |
| Rear clearance mfr | 150 mm |
| Maximum Accessible Power (embedded Laser) | 1 mW |
| Fuse type (spectrometer) mfr | 4 equivalent |
| Fuse type (microscope) mfr | 1 equivalent |
| Fuse type mfr | 1 equivalent |
| Liquid Nitrogen Temperature | -196 °C |
| Purge Gas Dew Point | -70 °C |
Description
Construction and series positioning
- The Cary 660 is a spectrometer-only configuration; the Cary 610 FTIR microscope is a separate unit that connects to the spectrometer via the EXT DET 1 or EXT DET 2 penthouse ports. The 670 and 680 are higher-tier spectrometers in the same series, while the 620 microscope is excluded from this guide.
- The spectrometer penthouse provides EXT DET 1, EXT DET 2, Analog Expansion, FPA/TRS, USB 2.0, Accessory, TRIG OUT, USER 1 IN, SIGNAL OUT, and USER 5 IN connectors. It contains no user-serviceable parts and must not be opened.
- The instrument uses a T4 AH 250 V fuse (5×20 mm, Littelfuse 0215004 or equivalent); the 610 microscope uses a lower-rated 1FS1 T2.0 AH 250 V fuse.
Behavior under load and media
- The embedded laser is <1 mW at 632.8 nm, continuous, with a maximum beam divergence of 1.3 milliradians; accessible laser light during operation is <600 μW with a divergence cone of 15 degrees or less. The interferometer compartment interlock automatically turns off laser power when the cover is opened.
- MCT and InSb detectors require liquid-nitrogen cooling at -196 °C. Initial Dewar fill is about 500 mL, followed by about 20 minutes to reach equilibrium, then a 200 mL top-off. Insufficient cooldown can degrade data accuracy.
- Purge gas must be water-free air dried to a dew point of -70 °C or dry nitrogen. Maximum purge flow is 10 L/min; 10 L/min is recommended when purging both spectrometer and microscope, and a second flow meter is recommended for independent microscope control.
Installation, control and operation
- Allow at least 50 mm side clearance and 150 mm rear clearance for free air circulation; insufficient clearance can cause overheating.
- A pressure regulator is required if the purge gas source pressure exceeds 60 psi (420 kPa).
- After an extended off period, the spectrometer takes about 30 minutes to reach operating temperature; collecting data before then may produce inaccurate data.
- The top power/scan LED is solid green when initialized and ready, flashing green when scanning, and solid yellow if initialization failed. A flashing blue A/D over-range LED indicates the A/D exceeds its maximum value; reduce sensitivity or increase beam attenuation.
- The motorized stage controller (accessory) has RS-232, Power, USB (currently not used), Joystick, Z, and Stage connectors; use an RS-232 to USB adapter if the computer has no serial port.
Integration & Operating Notes
Moisture in the purge supply can compromise optical performance and detector cooling. Use water-free air dried to a dew point of -70 °C or dry nitrogen, and monitor flow with a flow meter; a second flow meter is recommended for independent microscope control.
Source pressures above 60 psi (420 kPa) require a regulator to protect the instrument from overpressure. Maximum purge flow is 10 L/min; 10 L/min is recommended when purging both spectrometer and microscope.
Insufficient clearance can cause overheating and impair instrument performance. Ensure at least 50 mm (2 in.) on the sides and 150 mm (6 in.) at the rear of the system.
The spectrometer uses a T4 AH 250 V fuse (IEC601272-2 Sheet 5, 5×20 mm, Littelfuse 0215004 or equivalent). Do not substitute the lower-rated microscope fuse (1FS1 T2.0 AH 250 V) in the spectrometer.
Solid green power/scan LED indicates initialized and ready; flashing green indicates scanning; solid yellow indicates initialization failure. A flashing blue A/D over-range LED means the A/D exceeds its maximum value; reduce sensitivity or increase beam attenuation.
Use a soft cloth dampened with water or mild detergent; do not use organic solvents or abrasive cleaning agents, which can damage instrument surfaces and markings. Routine maintenance procedures are in the Resolutions Pro Help.
Selection & Integration Q&A
The Cary 660 is the base spectrometer in the 600 Series; the 670 and 680 are higher-tier siblings. Choose the 660 for routine FTIR without the additional capabilities of the 670/680. The 610 microscope is a separate accessory that connects via the penthouse EXT DET ports.
The microscope detector output connects to the spectrometer EXT DET 1 or EXT DET 2 penthouse connector. A second purge flow meter is recommended for independent microscope control. The microscope uses its own 1FS1 T2.0 AH 250 V fuse, not the spectrometer fuse.
Liquid nitrogen (500 mL initial plus 200 mL top-off) is required for MCT/InSb detectors but not supplied. Purge gas (dry air or nitrogen) and a pressure regulator (if source >60 psi) are user-supplied. Clean exterior with water or mild detergent only; routine maintenance is in the Resolutions Pro Help.