Agilent Technologies Cary 8454 UV Vis
Agilent Technologies Cary 8454 UV Vis — Molecular spectroscopy with Temperature 40 °C.
The Agilent Cary 8454 is a single-beam diode-array UV-Visible spectrophotometer covering 190–1100 nm with quartz cells, 350–1100 nm with glass, and 400 nm upward with disposable plastic. Its open sample area accepts quick-exchange cell holders and accessories, and it runs from 100–240 V AC at 50/60 Hz with 220 VA maximum consumption and no voltage selector. It is intended for routine and research molecular spectroscopy where fast, reproducible absorbance measurements and flexible cell configurations are required.
Full specifications
| Temperature | 40 °C |
| RAM | 4 GB |
| Warm-up Time | 1 hour |
| Clearance required at rear for electrical connections mfr | 8 cm |
| Wavelength range with quartz cells mfr | 190...1100 nm |
| Wavelength range with glass cells mfr | 350...1100 nm |
| Wavelength range with disposable plastic cells mfr | 400 nm |
| Turn-on and self-test duration mfr | 1...2 minutes |
| Electronic self-test duration before lamp ignition mfr | 20 seconds |
| Side clearance for air circulation mfr | 2.5 cm |
| Rear clearance for electrical connections mfr | 8 cm |
| Maximum cell wall angle mfr | 0.1 arc |
| Processor mfr | 3 GHz |
| A (in), B (out) mfr | Without stirring stirring |
| C (in), D (out) mfr | Without cell thermostatting thermostatting |
| A (in), B to D, C (out) mfr | In series with cell thermostatting thermostattin |
| A and C (in), B and D (out) mfr | In parallel with cell thermostatting thermostatt |
| Side clearance mfr | 2.5 cm |
| Rear clearance mfr | 8 cm |
| Hard disk mfr | 40 GB |
| Depth | 56.0 cm |
| Height | 18.5 cm |
| Weight | 14 kg |
| Width | 34.4 cm |
| Recommended cell path length mfr | 10 mm |
| Mass storage mfr | 40 ROM |
| Power consumption | 220 VA |
| Line voltage mfr | 100...240 AC |
| Line frequency mfr | 50 Hz |
| LAN interface card buffer handling settings mfr | 145 |
| Screen resolution mfr | Super VGA VGA |
| Operating Altitude | 2000 m |
| Operating temperature | 5 °C |
| Clearance required at sides for cooling mfr | 2.5 cm |
| Operating humidity mfr | 95 |
| Non-operating altitude mfr | 4600 m |
| Non-operating humidity mfr | 90 %RH |
| Humidity (%RH) non-condensing mfr | 90 |
Description
Construction and positioning
- Single-beam optical architecture with an open sample area that does not suffer from ambient stray light and permits rapid exchange of cell holders.
- Physical envelope: 34.4 cm wide × 56.0 cm deep × 18.5 cm high, 14 kg; bench planning must add 2.5 cm side clearance for air circulation and about 8 cm at the rear for electrical connections.
- No sibling models are defined in the verified facts, so this card describes the Cary 8454 only; do not infer variants from other series members.
Optical behavior and media
- Wavelength coverage depends on cell material: quartz cells 190–1100 nm, glass cells 350–1100 nm, disposable plastic cells from 400 nm (generally not recommended).
- Recommended cell path length is 10 mm; cell walls must be parallel to below 0.1 degrees of arc, because wedge-shaped cells cause spatial shift of the light beam and degrade photometric accuracy in diode-array instruments.
- Quartz cells with black apertures smaller than 2 mm, used with a multicell transport, can give poor reproducibility; avoid sub-2 mm black apertures in that configuration.
- As a single-beam instrument, a blank measurement should be taken every half hour even in a thermally stable environment.
Installation, control and operation
- Line power: 100–240 V AC, 50/60 Hz, 220 VA maximum; no voltage selector and no externally accessible fuses — automatic electronic fuses protect the supply.
- Turn-on: electronic self-tests take about 20 seconds before lamps ignite; the entire turn-on and self-test process takes about 1–2 minutes. Wait for the front-panel indicator to turn green before starting ChemStation, otherwise the software may not recognize the instrument.
- Recommended warm-up is at least 1 hour after self-tests; the automated self-test in the diagnostic software can only run after that warm-up.
- ChemStation allows a maximum of two parallel sessions per instrument; the online session must be launched first.
- LAN communication requires correct buffer handling settings (T145=64:T146=01:T147=01) and no DHCP server in the same subnet; use fixed IP addresses reserved for instruments.
- The RS 232 connector is not active by default; it must be enabled by the configuration switch module next to the GPIB connector.
- Waste tubing, where used, must have a constant down slope from the instrument to the waste container to prevent liquid accumulation.
- Two independent safety light switches turn off the lamps when the sheet-metal lamp door is opened; a security lever prevents cover removal while line power is connected.
Integration & Operating Notes
The instrument accepts 100–240 V AC at 50/60 Hz and uses automatic electronic fuses; there is no voltage selector to set and no externally accessible fuse to replace. Site power must simply fall within this range and provide 220 VA maximum.
Bench space must accommodate the 34.4 × 56.0 × 18.5 cm envelope plus 2.5 cm clearance on either side for air circulation and approximately 8 cm at the rear for electrical connections. Blocking these clearances risks overheating or inaccessible cabling.
Quartz cells are required for the full 190–1100 nm range; glass cells limit operation to 350–1100 nm; disposable plastic cells are usable only from 400 nm and are generally not recommended. Selecting the wrong cell material invalidates measurements below its cutoff.
Use 10 mm path length cells whose opposite walls are parallel to below 0.1 degrees of arc. Non-parallel or wedge-shaped cells cause spatial shift of the light beam and degrade photometric accuracy in this diode-array instrument.
After power-on, electronic self-tests run about 20 seconds before lamps ignite, and the full turn-on and self-test takes about 1–2 minutes. Start ChemStation only after the front-panel indicator turns green, otherwise the software may not recognize the spectrophotometer.
A maximum of two ChemStation sessions per instrument can run in parallel, and the online session must always be the first launched. Launching offline first can cause connection conflicts.
A DHCP server in the same subnet may respond to bootp requests, assign different IP addresses each time, and fail to set buffer handling correctly. Use fixed IP addresses reserved for instruments and set LAN card buffers to T145=64:T146=01:T147=01.
Because the Cary 8454 is a single-beam spectrophotometer, a blank measurement should be taken every half hour even in a thermally stable environment. Skipping this degrades accuracy as lamp and detector drift accumulate.
For best results, let the spectrophotometer warm up at least 1 hour after passing self-tests. The automated self-test in the diagnostic software can only be executed after this warm-up period.
Where waste tubing is connected, it must have a constant down slope from the spectrophotometer to the waste container. Improper slope can cause liquid accumulation and potential spillage or instrument damage.
Selection & Integration Q&A
The verified facts define no sibling models, so selection is against application needs rather than variants: choose the Cary 8454 when you need 190–1100 nm coverage with quartz cells, an open sample area for quick cell-holder exchange, and single-beam operation with frequent blanking. For work confined above 350 nm, glass cells suffice; below 400 nm, quartz is mandatory.
The PC runs Agilent ChemStation (software revision B.05.xx) and connects over LAN using a LAN interface card in the spectrophotometer. Use fixed IP addresses reserved for instruments, run no DHCP server in the same subnet, and set buffer handling to T145=64:T146=01:T147=01. RS 232 is inactive unless enabled by the configuration switch module next to the GPIB connector. Cells must be 10 mm path length with walls parallel below 0.1 degrees of arc.
Install below 2000 m altitude, within 5–40 °C and <95 %RH non-condensing, with 2.5 cm side and 8 cm rear clearance. For optimum performance hold the lab at 20–25 °C ±2 °C through the working day, away from direct sunlight and variable air circulation. Allow 1 hour warm-up after the 1–2 minute turn-on and self-test before running the automated self-test.
Take a blank measurement every half hour. Optional accessories installed separately include the thermostattable cell holder, cell stirring module, long path cell holder and optical filter wheel. The computer, printer, LAN cabling and ChemStation software are not supplied with non-preconfigured systems; quartz, glass or plastic cells and waste tubing are also user-supplied. Lamp replacement is protected by safety light switches but still requires disconnecting line power.