Agilent Technologies FRG 720
Agilent Technologies FRG 720 — Vacuum gauges with Max pressure 0.2 megapascal.
The FRG-720 is a Bayard-Alpert Pirani combination full-range vacuum gauge covering 5×10⁻¹⁰ to 1000 mbar in a single sensor. It is available in NW 25 KF or DN 40 CF flange variants, with or without a 32×16 LCD display, and with an optional Profibus interface plus two relay switching functions. It serves as the primary pressure sensor in high- and ultra-high-vacuum systems where continuous atmosphere-to-UHV coverage is required.
Full specifications
| Max pressure | 2 bar |
| Measurement range | 0.774 V |
| Electrical connection | 34 |
| Internal Volume | 24 microliter |
| Operation | 43 |
| Vacuum connection | 26 millimeter |
| Measurement range (air, O2, CO, N2) mfr | 5 mbar |
| Accuracy (after 10 min. stabilization) mfr | 15 |
| Repeatability (after 10 min. stabilization) mfr | 5 |
| Switching ON threshold mfr | 2.4 mbar |
| Switching OFF threshold mfr | 3.2 mbar |
| Degas duration mfr | 3 min. |
| Minimum loaded impedance mfr | 10 |
| Pressure unit mfr | mbar , Torr, Pa Pa |
| Gauge identification mfr | 42 GND |
| Adjustment range mfr | 1 mbar |
| RS232C data rate mfr | 9600 Baud |
| RS232C data format mfr | 8 handshake |
| RS232C TxD mfr | 13 |
| RS232C RxD mfr | 14 |
| RS232C GND mfr | 5 |
| Materials exposed to vacuum - housing, supports, screens mfr | stainless steel steel |
| Materials exposed to vacuum - feedthroughs mfr | NiFe, nickel plated plated |
| Materials exposed to vacuum - insulator mfr | glass glass |
| Materials exposed to vacuum - cathode mfr | iridium, yttrium oxide oxide |
| Materials exposed to vacuum - cathode holder mfr | molybdenum molybdenum |
| Materials exposed to vacuum - Pirani element mfr | tungsten, copper copper |
| Use mfr | 2000 NN |
| Type of protection mfr | 30 |
| Switching functions mfr | 56 |
| Materials exposed to vacuum - Housing, supports, screens mfr | stainless steel steel |
| Materials exposed to vacuum - Feedthroughs mfr | NiFe, nickel plated plated |
| Materials exposed to vacuum - Insulator mfr | glass glass |
| Materials exposed to vacuum - Cathode mfr | iridium, yttrium oxide oxide |
| Materials exposed to vacuum - Cathode holder mfr | molybdenum molybdenum |
| FRG720KF25 mfr | -720 flange. |
| FRG720CF35 mfr | -720 flange. |
| FRG720KF25D mfr | -720 flange. |
| FRG720CF35D mfr | -720 flange. |
| FRG720KF25SP mfr | 430 g |
| FRG720CF35SP mfr | 695 g |
| FRG720BKE mfr | -720 mm. |
| FRG720KF25RS mfr | -720 KF. |
| FRG720CF35RS mfr | -720 Flange. |
| FRG720BFL mfr | -720 gauge. |
| FRG720PWR mfr | -720 C. |
| Product Description mfr | 15 |
| Technical Specifications mfr | 17 |
| FRG-720 Outline mfr | 21 |
| Measurement Signal vs. Pressure mfr | 22 |
| Installation mfr | 26 |
| Removing and Installing the Electronics Unit mfr | 28 |
| Installing the Optional Extension mfr | 29 |
| Using the Optional Baffle mfr | 31 |
| Use With Other Controllers mfr | 35 |
| Measuring Principle, Measuring Behavior mfr | 43 |
| Putting the Gauge Into Operation mfr | 46 |
| Description of the Functions mfr | 53 |
| Operating Parameters mfr | 54 |
| Setting the Switching Functions mfr | 57 |
| Deinstallation mfr | 59 |
| Maintenance, Repair mfr | 61 |
| Cleaning the Gauge mfr | 61 |
| Adjusting the Gauge mfr | 62 |
| Adjustment at Atmospheric Pressure mfr | 62 |
| Zero Point Adjustment mfr | 64 |
| Troubleshooting mfr | 65 |
| What to Do in Case of Problems mfr | 65 |
| Replacing the Sensor mfr | 67 |
| Accessories and Spare Parts mfr | 68 |
| Analog values only Without degas function mfr | 4 shielding |
| Analog values with degas function mfr | 5 shielding |
| RS232C operation mfr | 30 m |
| FRG720KF25, FRG720KF25D mfr | 285 g |
| FRG720CF35, FRG720CF35D mfr | 550 g |
| Cable mfr | 9 female |
| [V] mfr | [mbar] |
| Accuracy mfr | 15 mbar |
| Repeatability mfr | 5 mbar |
| Output slope mfr | 0.75 V/decade |
| Weight | 285 g |
| Max. cable length RS232C operation mfr | 30 m |
| Weight FRG720KF25, FRG720KF25D mfr | 285 g |
| Weight FRG720CF35, FRG720CF35D mfr | 550 g |
| Weight FRG720KF25SP mfr | 430 g |
| Weight FRG720CF35SP mfr | 695 g |
| Max. cable length RS232C mfr | 30 m |
| Weight FRG720CF35 mfr | 550 g |
| Current | 0.5 ADC |
| Emission Current | 5 mA |
| Power consumption | 16 W |
| Voltage | 60 V |
| Degas emission current mfr | 16 mA |
| Voltage vs. pressure mfr | 0.75 V/decade |
| Error signal mfr | 0.3 V |
| Operating voltage at the gauge mfr | 24 Vpp |
| Power consumption degas mfr | 0.8 A |
| Power consumption emission start (<200 ms) mfr | 1.4 A |
| Fuse necessary mfr | 1.25 AT |
| Max. cable length (supply voltage 24 V) mfr | 30 m |
| Relay contact rating voltage mfr | 60 V |
| Relay contact rating current mfr | 0.5 ADC |
| Optional power supply mains voltage mfr | 90 VAC |
| Optional power supply output voltage mfr | 21 VDC |
| Optional power supply output current mfr | 1.5 A |
| Voltage vs. Pressure mfr | 0.75 V/decade |
| Degas mfr | 0.8 A |
| Emission start (<200 ms) mfr | 1.4 A |
| Mains cable mfr | 1.8 meter |
| Operating voltage mfr | 24 VDC |
| Fuse mfr | 1.25 AT |
| Relay contact voltage mfr | 60 V |
| Relay contact current mfr | 0.5 ADC |
| Output voltage (optional power supply) mfr | 21 VDC |
| Output current (optional power supply) mfr | 1.5 A |
| Profibus Data Rate | 12 MBaud |
| Display panel mfr | 32 illumination |
| Display dimensions mfr | 16.0 mm |
| Electrical connector mfr | 15 male |
| Cable for Profibus gauge mfr | 15 shielding |
| Switching functions Profibus gauge mfr | 2 |
| Profibus fieldbus name mfr | Profibus Profibus |
| Profibus connection mfr | 9 female |
| Profibus cable mfr | shielded, special Profibus cable according to Pr |
| Use With Agilent Vacuum Gauge Controller mfr | 34 |
| Display mfr | 47 |
| RS232C Interface mfr | 48 |
| Profibus Interface mfr | 53 |
| Profibus gauge mfr | 2 |
| Analog values with degas function and RS232C interface mfr | 7 shielding |
| Analog and fieldbus operation mfr | 35 mm2 |
| Connector mfr | 15 female |
| Output signal range mfr | 0.774 V |
| Power consumption Profibus mfr | 18 W |
| Max. cable length analog/fieldbus mfr | 100 mm2 |
| Gas type dependence mfr | gas type dependent dependent |
| Gas Type Dependence mfr | 23 |
| Bakeout Temperature | 150 °C |
| Storage temperature | 20 °C |
| Relative humidity mfr | 65 |
| Relative humidity (year's mean / during 60 days) mfr | 65 |
| Cable for standard gauge mfr | 7 shielding |
| Profibus standard applied mfr | 485 |
| Standard mfr | 0.5 A |
| Standard gauge mfr | 16 W |
Description
Construction and series positioning
- The FRG-720 combines a Bayard-Alpert hot-cathode ionization system (active for p < 2.0×10⁻² mbar) with a Pirani thermal-conductance system (active for p > 5.5×10⁻³ mbar); in the overlap 2.0×10⁻² … 5.5×10⁻³ mbar a mixed signal is output.
- Six configurations exist: FRG720KF25 and FRG720CF35 (no display), FRG720KF25D and FRG720CF35D (32×16 LCD), and FRG720KF25SP and FRG720CF35SP (Profibus interface with two switching functions).
- Internal volume is ≤24 cm³ for DN 25 ISO-KF and ≤34 cm³ for DN 40 CF-R; maximum pressure is 2 bar absolute. Weight is 285 g.
Measurement behavior
- The hot cathode switches on only below 2.4×10⁻² mbar and off above 3.2×10⁻² mbar to prevent filament burn-out; Pirani measurement covers the higher-pressure region.
- Emission current is 5 mA at p ≤7.2×10⁻⁶ mbar and 25 µA between 7.2×10⁻⁶ and 3.2×10⁻² mbar, switching automatically.
- Output is 0 … +10 V logarithmic at 0.75 V/decade; minimum load impedance is 10 kΩ. Accuracy is 15% of reading and repeatability 5% of reading after 10 minutes stabilization.
- Degas heats the collector grid to approximately 700 °C by electron bombardment at ~16 mA emission current; it should be run below 7.2×10⁻⁶ mbar and terminates automatically after 3 minutes.
Installation, control and operation
- The gauge may be mounted in any orientation, preferably horizontal to upright to keep condensates and particles out. The sensor can be baked to 150 °C; above 50 °C the electronics unit must be removed or an optional extension installed.
- Operating voltage is +24 VDC (+20 … +28 VDC) with ripple ≤2 Vpp; a 1.25 AT fuse is required. Power consumption is 0.5 A standard, 0.8 A during degas, and 1.4 A for <200 ms at emission start.
- The gauge can be run from an Agilent AGC-100 or Turbo AG Rack Controller, or from another controller via the 15-pin male connector. RS232C output is sent approximately every 20 ms; Profibus variants support ≤12 MBaud and node addresses 00 … 7Dhex.
- The gauge must not be operated with an Agilent controller and Profibus simultaneously, nor with RS232 and Profibus simultaneously.
Integration & Operating Notes
The gauge draws 0.5 A continuously, 0.8 A during degas, and up to 1.4 A for <200 ms at emission start. A 1.25 AT fuse protects the wiring, and the supply must be SELV with ripple ≤2 Vpp to avoid measurement noise or electronics damage.
The sensor's maximum pressure is 2 bar absolute. Exceeding this risks mechanical damage to the electrode system and loss of calibration. Ensure upstream valves and process sequences cannot pressurize the gauge beyond this limit.
The sensor may be baked to 150 °C, but the electronics unit must be removed or an optional bakeout extension installed whenever temperature exceeds 50 °C. With the extension fitted, in-operation bakeout is limited to p < 10⁻² mbar because Pirani accuracy degrades at high temperature.
Degas heats the collector grid to ~700 °C by electron bombardment at ~16 mA. It must be initiated only at pressures below 7.2×10⁻⁶ mbar; the gauge stops degas automatically after 3 minutes to prevent overheating.
Maximum cable length is ≤35 m with 0.25 mm² conductors, ≤50 m with 0.34 mm², and ≤100 m with 1.0 mm². RS232C operation is limited to ≤30 m regardless. For runs above 5 m, measure the output differentially between Pin 2 and Pin 12 to preserve accuracy.
After replacing a contaminated or damaged sensor, perform the atmospheric adjustment at 1000 mbar and the Pirani zero-point adjustment. The zero-point correction is required if the gauge fails to switch over to BA at low pressures.
Selection & Integration Q&A
Choose the FRG-720 when one sensor must cover 5×10⁻¹⁰ to 1000 mbar continuously, avoiding a separate Pirani and BA gauge. Select NW 25 KF for elastomer-sealed systems or DN 40 CF for UHV; add the D suffix if local pressure readout is needed, or the SP suffix for Profibus control with two relay setpoints.
It runs on +24 VDC (+20…+28 VDC, ripple ≤2 Vpp, 1.25 AT fuse) and can be driven by an Agilent AGC-100 or Turbo AG Rack Controller, or by a third-party controller via the 15-pin connector. A 42 kΩ resistor between Pin 10 and GND identifies the gauge to Agilent controllers. RS232C or Profibus can be used, but not both simultaneously.
Mount in any orientation, preferably horizontal to upright to keep condensates and particles out. Operating ambient is 0…+50 °C, IP 30, indoor use only up to 2000 m altitude. The sensor can be baked to 150 °C only if the electronics unit is removed or a bakeout extension is fitted; in-operation bakeout with the extension is limited to p < 10⁻² mbar.
The base delivery does not include the 24 V power supply, sensor cable, bakeout extension, baffle, or replacement sensor. Plan for periodic atmospheric and zero-point adjustment, and sensor replacement if the filament or Pirani element fails or contamination is severe. Degas can be run in situ below 7.2×10⁻⁶ mbar for 3 minutes to clean the electrodes.