Agilent Technologies FRG 720

Agilent Technologies FRG 720 — Vacuum gauges with Max pressure 0.2 megapascal.

The FRG-720 is a Bayard-Alpert Pirani combination full-range vacuum gauge covering 5×10⁻¹⁰ to 1000 mbar in a single sensor. It is available in NW 25 KF or DN 40 CF flange variants, with or without a 32×16 LCD display, and with an optional Profibus interface plus two relay switching functions. It serves as the primary pressure sensor in high- and ultra-high-vacuum systems where continuous atmosphere-to-UHV coverage is required.

Max pressure
0.2 megapascal

Full specifications

Core Performance & Functional Specifications
Max pressure 2 bar
Measurement range 0.774 V
Electrical connection 34
Internal Volume 24 microliter
Operation 43
Vacuum connection 26 millimeter
Measurement range (air, O2, CO, N2) mfr 5 mbar
Accuracy (after 10 min. stabilization) mfr 15
Repeatability (after 10 min. stabilization) mfr 5
Switching ON threshold mfr 2.4 mbar
Switching OFF threshold mfr 3.2 mbar
Degas duration mfr 3 min.
Minimum loaded impedance mfr 10
Pressure unit mfr mbar , Torr, Pa Pa
Gauge identification mfr 42 GND
Adjustment range mfr 1 mbar
RS232C data rate mfr 9600 Baud
RS232C data format mfr 8 handshake
RS232C TxD mfr 13
RS232C RxD mfr 14
RS232C GND mfr 5
Materials exposed to vacuum - housing, supports, screens mfr stainless steel steel
Materials exposed to vacuum - feedthroughs mfr NiFe, nickel plated plated
Materials exposed to vacuum - insulator mfr glass glass
Materials exposed to vacuum - cathode mfr iridium, yttrium oxide oxide
Materials exposed to vacuum - cathode holder mfr molybdenum molybdenum
Materials exposed to vacuum - Pirani element mfr tungsten, copper copper
Use mfr 2000 NN
Type of protection mfr 30
Switching functions mfr 56
Materials exposed to vacuum - Housing, supports, screens mfr stainless steel steel
Materials exposed to vacuum - Feedthroughs mfr NiFe, nickel plated plated
Materials exposed to vacuum - Insulator mfr glass glass
Materials exposed to vacuum - Cathode mfr iridium, yttrium oxide oxide
Materials exposed to vacuum - Cathode holder mfr molybdenum molybdenum
FRG720KF25 mfr -720 flange.
FRG720CF35 mfr -720 flange.
FRG720KF25D mfr -720 flange.
FRG720CF35D mfr -720 flange.
FRG720KF25SP mfr 430 g
FRG720CF35SP mfr 695 g
FRG720BKE mfr -720 mm.
FRG720KF25RS mfr -720 KF.
FRG720CF35RS mfr -720 Flange.
FRG720BFL mfr -720 gauge.
FRG720PWR mfr -720 C.
Product Description mfr 15
Technical Specifications mfr 17
FRG-720 Outline mfr 21
Measurement Signal vs. Pressure mfr 22
Installation mfr 26
Removing and Installing the Electronics Unit mfr 28
Installing the Optional Extension mfr 29
Using the Optional Baffle mfr 31
Use With Other Controllers mfr 35
Measuring Principle, Measuring Behavior mfr 43
Putting the Gauge Into Operation mfr 46
Description of the Functions mfr 53
Operating Parameters mfr 54
Setting the Switching Functions mfr 57
Deinstallation mfr 59
Maintenance, Repair mfr 61
Cleaning the Gauge mfr 61
Adjusting the Gauge mfr 62
Adjustment at Atmospheric Pressure mfr 62
Zero Point Adjustment mfr 64
Troubleshooting mfr 65
What to Do in Case of Problems mfr 65
Replacing the Sensor mfr 67
Accessories and Spare Parts mfr 68
Analog values only Without degas function mfr 4 shielding
Analog values with degas function mfr 5 shielding
RS232C operation mfr 30 m
FRG720KF25, FRG720KF25D mfr 285 g
FRG720CF35, FRG720CF35D mfr 550 g
Cable mfr 9 female
[V] mfr [mbar]
Accuracy mfr 15 mbar
Repeatability mfr 5 mbar
Output slope mfr 0.75 V/decade
Dimensions, Weight & Mechanical Design
Weight 285 g
Max. cable length RS232C operation mfr 30 m
Weight FRG720KF25, FRG720KF25D mfr 285 g
Weight FRG720CF35, FRG720CF35D mfr 550 g
Weight FRG720KF25SP mfr 430 g
Weight FRG720CF35SP mfr 695 g
Max. cable length RS232C mfr 30 m
Weight FRG720CF35 mfr 550 g
Electrical, Power & Drive Specifications
Current 0.5 ADC
Emission Current 5 mA
Power consumption 16 W
Voltage 60 V
Degas emission current mfr 16 mA
Voltage vs. pressure mfr 0.75 V/decade
Error signal mfr 0.3 V
Operating voltage at the gauge mfr 24 Vpp
Power consumption degas mfr 0.8 A
Power consumption emission start (<200 ms) mfr 1.4 A
Fuse necessary mfr 1.25 AT
Max. cable length (supply voltage 24 V) mfr 30 m
Relay contact rating voltage mfr 60 V
Relay contact rating current mfr 0.5 ADC
Optional power supply mains voltage mfr 90 VAC
Optional power supply output voltage mfr 21 VDC
Optional power supply output current mfr 1.5 A
Voltage vs. Pressure mfr 0.75 V/decade
Degas mfr 0.8 A
Emission start (<200 ms) mfr 1.4 A
Mains cable mfr 1.8 meter
Operating voltage mfr 24 VDC
Fuse mfr 1.25 AT
Relay contact voltage mfr 60 V
Relay contact current mfr 0.5 ADC
Output voltage (optional power supply) mfr 21 VDC
Output current (optional power supply) mfr 1.5 A
Interfaces, Connections & Control
Profibus Data Rate 12 MBaud
Display panel mfr 32 illumination
Display dimensions mfr 16.0 mm
Electrical connector mfr 15 male
Cable for Profibus gauge mfr 15 shielding
Switching functions Profibus gauge mfr 2
Profibus fieldbus name mfr Profibus Profibus
Profibus connection mfr 9 female
Profibus cable mfr shielded, special Profibus cable according to Pr
Use With Agilent Vacuum Gauge Controller mfr 34
Display mfr 47
RS232C Interface mfr 48
Profibus Interface mfr 53
Profibus gauge mfr 2
Analog values with degas function and RS232C interface mfr 7 shielding
Analog and fieldbus operation mfr 35 mm2
Connector mfr 15 female
Output signal range mfr 0.774 V
Power consumption Profibus mfr 18 W
Max. cable length analog/fieldbus mfr 100 mm2
Materials, Media & Process Compatibility
Gas type dependence mfr gas type dependent dependent
Gas Type Dependence mfr 23
Operating Conditions, Environment & Protection
Bakeout Temperature 150 °C
Storage temperature 20 °C
Relative humidity mfr 65
Relative humidity (year's mean / during 60 days) mfr 65
Standards, Safety & General Data
Cable for standard gauge mfr 7 shielding
Profibus standard applied mfr 485
Standard mfr 0.5 A
Standard gauge mfr 16 W

Description

Construction and series positioning

Measurement behavior

Installation, control and operation

Integration & Operating Notes

Electrical
Fuse the 24 VDC supply with 1.25 AT and use SELV

The gauge draws 0.5 A continuously, 0.8 A during degas, and up to 1.4 A for <200 ms at emission start. A 1.25 AT fuse protects the wiring, and the supply must be SELV with ripple ≤2 Vpp to avoid measurement noise or electronics damage.

Technical Specifications, page 18
Process Limits
Never exceed 2 bar absolute at the flange

The sensor's maximum pressure is 2 bar absolute. Exceeding this risks mechanical damage to the electrode system and loss of calibration. Ensure upstream valves and process sequences cannot pressurize the gauge beyond this limit.

Technical Specifications, page 19
Mounting
Remove electronics or fit extension before baking above 50 °C

The sensor may be baked to 150 °C, but the electronics unit must be removed or an optional bakeout extension installed whenever temperature exceeds 50 °C. With the extension fitted, in-operation bakeout is limited to p < 10⁻² mbar because Pirani accuracy degrades at high temperature.

Installation, page 27; Installing the Optional Extension, page 29
Process Limits
Run degas only below 7.2×10⁻⁶ mbar

Degas heats the collector grid to ~700 °C by electron bombardment at ~16 mA. It must be initiated only at pressures below 7.2×10⁻⁶ mbar; the gauge stops degas automatically after 3 minutes to prevent overheating.

Degas, page 46
Electrical
Match cable conductor cross-section to run length

Maximum cable length is ≤35 m with 0.25 mm² conductors, ≤50 m with 0.34 mm², and ≤100 m with 1.0 mm². RS232C operation is limited to ≤30 m regardless. For runs above 5 m, measure the output differentially between Pin 2 and Pin 12 to preserve accuracy.

Technical Specifications, page 18; Electrical Connection, page 38
Maintenance
Adjust at atmosphere and zero after sensor replacement

After replacing a contaminated or damaged sensor, perform the atmospheric adjustment at 1000 mbar and the Pirani zero-point adjustment. The zero-point correction is required if the gauge fails to switch over to BA at low pressures.

Maintenance, Repair, Cleaning and Adjusting the Gauge

Selection & Integration Q&A

When should I choose the FRG-720 over a Pirani-only or BA-only gauge, and which flange/display variant fits my system?

Choose the FRG-720 when one sensor must cover 5×10⁻¹⁰ to 1000 mbar continuously, avoiding a separate Pirani and BA gauge. Select NW 25 KF for elastomer-sealed systems or DN 40 CF for UHV; add the D suffix if local pressure readout is needed, or the SP suffix for Profibus control with two relay setpoints.

What controller, power supply and cabling does the FRG-720 require, and can I use my own PLC?

It runs on +24 VDC (+20…+28 VDC, ripple ≤2 Vpp, 1.25 AT fuse) and can be driven by an Agilent AGC-100 or Turbo AG Rack Controller, or by a third-party controller via the 15-pin connector. A 42 kΩ resistor between Pin 10 and GND identifies the gauge to Agilent controllers. RS232C or Profibus can be used, but not both simultaneously.

What are the mounting orientation, ambient and bakeout limits for the FRG-720?

Mount in any orientation, preferably horizontal to upright to keep condensates and particles out. Operating ambient is 0…+50 °C, IP 30, indoor use only up to 2000 m altitude. The sensor can be baked to 150 °C only if the electronics unit is removed or a bakeout extension is fitted; in-operation bakeout with the extension is limited to p < 10⁻² mbar.

What maintenance and accessories are needed beyond the base gauge delivery?

The base delivery does not include the 24 V power supply, sensor cable, bakeout extension, baffle, or replacement sensor. Plan for periodic atmospheric and zero-point adjustment, and sensor replacement if the filament or Pirani element fails or contamination is severe. Degas can be run in situ below 7.2×10⁻⁶ mbar for 3 minutes to clean the electrodes.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
FRG-720 Vacuum Gauge User Manual
Manual ·73 pp.
View PDF