Agilent Technologies IMG 300 Inverted Magnetron Gauge
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies IMG 300 Inverted Magnetron Gauge

Agilent Technologies IMG 300 Inverted Magnetron Gauge — Vacuum gauges.

The IMG-300 is a 1.6-inch inverted magnetron cold cathode ionization gauge with an all-welded 304L stainless steel shell, ceramic-brazed feedthrough, and ConFlat flange, covering 10⁻³ to 10⁻¹¹ Torr. Its 3 kV DC high-voltage discharge, 2.5 A/Torr sensitivity, and <10 ms response suit fast UHV measurement where no hot filament is acceptable. It is intended for rugged UHV service and environments where stray electrons and photons disturb hot-filament gauges.

Full specifications

Core Performance & Functional Specifications
Measurement range -3...10 Torr
Sensitivity 2.5 A/Torr
Response Time 10 milliseconds
Feedthrough/UHV mfr Ceramic brazed brazed
Ignition Time mfr 5 Torr
Electrical, Power & Drive Specifications
Operating voltage (HV range) mfr 3 DC
Operating Voltage mfr 3 DC
Interfaces, Connections & Control
Electrical Connector mfr 5 rated
Materials, Media & Process Compatibility
Material Exposed to the Vacuum mfr 304 ceramic
Operating Conditions, Environment & Protection
Bakeout Temperature 250 °C
Operating temperature 250 °C

Description

Construction and series positioning

Behavior and measurement characteristics

Installation, connection, and operation

Integration & Operating Notes

Mounting
Keep magnetic material a few inches away from the gauge body

The inverted magnetron's external field interacts with nearby magnetic parts. Mount the gauge so that no magnetic force is felt at the chosen distance; contact with steel tools and rough handling can weaken the magnet and change calibration, so plan the flange location and tool access accordingly.

User Manual, Installation Guidelines, page 14
Electrical
Use SHV cable and connectors rated for at least 5 kV

The gauge operates at 3 kV DC on the high-voltage range and terminates in a standard SHV connector rated 5 kV. The mating cable and connector must carry that rating; a lower-rated assembly is not a safe substitute for the high-voltage discharge path.

User Manual, Table 1-1, page 15
Process Limits
Never energize the gauge with combustible gases present

The cold cathode high-voltage discharge can act as an ignition source, and the device is not intrinsically safe. The gauge must be disconnected from the controller and never turned on when combustible gases or mixtures are present in the system.

User Manual, Use with Combustibles and Mixtures, page 9
Process Limits
Allow up to 20 minutes for ignition at 1×10⁻¹⁰ Torr

Ignition is faster than 5 seconds above 1×10⁻⁶ Torr, but at 1×10⁻¹⁰ Torr it typically takes about 20 minutes. Control logic and operator procedures should tolerate this startup delay at very high vacuum rather than treating a slow start as a fault.

User Manual, Table 1-1, page 15; Application, page 14
Cooling & Media
Unplug the cable before bakeout at up to 250 °C

With the ConFlat metal gasket and magnet, the gauge may be baked to 250 °C, but the cable must be unplugged during bakeout. Plan the bakeout cycle and cable routing so the connector can be disconnected without disturbing the flange or the gauge alignment.

User Manual, Table 1-1, page 15; Application, page 13
Maintenance
Use Apiezon L grease only; never silicone oil or grease

Silicone oil or grease must not be used on the gauge or vacuum surfaces. If grease is necessary, apply Apiezon L grease (Agilent part number 695400004) sparingly. Do not clean aluminum parts with Alconox, and wear powder-free butyl or polycarbonate gloves when handling vacuum surfaces.

User Manual, Vacuum Equipment and Cleanliness, page 10
Maintenance
Calibration can be maintained under the STARRS service program

Periodic service and calibration are available from Agilent under the STARRS service program. Because mechanical shock or prolonged contact with the metal surface can change magnet strength and shift calibration, a recalibration path should be planned for after any rough handling.

User Manual, Application, page 14

Selection & Integration Q&A

When should I choose the IMG-300 instead of a hot-filament Bayard-Alpert gauge?

Choose the IMG-300 for rugged UHV service, fast response (<10 ms), and where stray electrons or photons disturb hot-filament gauges. The manual notes it is generally not as stable or linear as a well-maintained Bayard-Alpert gauge, so prefer the Bayard-Alpert when linearity and stability dominate over ruggedness.

What controller, cable, and connector does the IMG-300 require?

It connects to a Multi-Gauge controller via a standard SHV connector rated 5 kV, and the operating high-voltage range is 3 kV DC. The mating coaxial cable and connector must be rated for at least 5 kV. The controller and vacuum system must be separately grounded to a common ground tested for at least 10 A.

What mounting and environment limits apply to the IMG-300?

Mount at least a few inches from any magnetic material; if no magnetic force is felt, the distance is sufficient. Avoid steel tools and rough handling, which can weaken the magnet and shift calibration. Bakeout is allowed to 250 °C with the ConFlat metal gasket and magnet, with the cable unplugged. The gauge is Class A and may cause radio interference in domestic environments.

What maintenance, service, and accessories are needed beyond the base gauge?

The base gauge is part R0343301 (IMG-300 2.75 CFF); a maintenance kit is listed separately as R0342301. Calibration can be maintained under the STARRS service program. Use Apiezon L grease (part 695400004) sparingly if grease is needed, never silicone oil or grease, and wear powder-free butyl or polycarbonate gloves when handling vacuum surfaces.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
IMG-300 Inverted Magnetron Gauge
Manual ·24 pp.
View PDF