Agilent Technologies Transformer oil gas, transformer oil analysis
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies Transformer oil gas, transformer oil analysis

Agilent Technologies Transformer oil gas, transformer oil analysis — Gas chromatography.

The Agilent TOGA is a transformer oil gas analyzer built on the 8890 GC and 8697 headspace sampler, configured with FID/TCD detection and 2 valves/2 columns per ASTM D3612. It quantifies dissolved acetylene, ethane and methane in a 22.5-minute run, distinguishing the headspace (Method C) variant from vacuum-extraction (Method A) siblings.

Full specifications

Core Performance & Functional Specifications
Analysis time 22.5 min
Run Time 22.5 min
Software OpenLab CDS
GC platform mfr 8890
Headspace sampler mfr 8697
Detectors mfr FID/TCD FID/TCD
Analyzed compounds mfr acetylene, ethane, methane methane
Valve/column configuration mfr 2 columns/TCD/FID
Detector configuration mfr TCD/FID TCD/FID
Enhanced method mfr 3612 C
Pneumatics control mfr microchannel-based electronic pneumatics control
Measured components mfr 2
Maximum sample sequence mfr 120 samples
Sampler mfr 8697 sampler
Standards, Safety & General Data
Standard configuration 1 mfr 3612 A
Standard configuration 2 mfr 3612 columns
Standard configuration 3 mfr 3612 opt
Standard method mfr 3612 A
Compliance standard mfr 3612

Description

Construction and series positioning

Behavior and run characteristics

Control and operation

Integration & Operating Notes

Process Limits
Match the configuration to the extraction method before ordering

The TOGA family spans ASTM D3612 Method A (vacuum extraction, packed columns) and Method C (headspace injection, PLOT columns with methanizer). This FID/TCD headspace configuration is not interchangeable with the Method A packed-column variant; the extraction procedure used in the laboratory must be fixed first, because column set, valving and methanizer presence follow from it.

TOGA GC Analyzer brochure 5994-8705 (EN/DE/FR), scope and variants
Control & I/O
Sequence capacity of 120 samples sets the unattended batch window

At 22.5 min per run, a full 120-sample sequence represents roughly 45 hours of continuous operation. Plan sampler vial capacity, tray loading and any intermediate calibration or check standards around that horizon so that a batch is not interrupted mid-sequence.

Factory Specification Sheet; TOGA GC Analyzer brochure 5994-8705
Maintenance
Consumables and gas supplies are not part of the base delivery

Columns, methanizer, sampler vials and septa, calibration gas standards and carrier/detector gases are separate items; the base system comprises the 8890 GC, 8697 sampler and OpenLab CDS. Budget these consumables and their replacement intervals into commissioning, since the supplied documentation digests give no maintenance schedule.

TOGA GC Analyzer brochure 5994-8705; Factory Specification Sheet
Electrical
Utility and site requirements must be taken from the full manuals

The available document digests list no voltage, power, gas-purity or bench-clearance data for this configuration. Confirm electrical supply, carrier and fuel gas plumbing, and exhaust/vent routing against the 8890 GC and 8697 site-preparation manuals before installation; do not size utilities from this card.

TOGA GC Analyzer brochure 5994-8705 (digests contain no section-level utility facts

Selection & Integration Q&A

When should I choose this headspace FID/TCD TOGA instead of the vacuum-extraction (Method A) or trace-olefin backflush variants?

Pick this configuration when the laboratory follows ASTM D3612 Method C headspace injection and needs acetylene, ethane and methane by FID/TCD in a 22.5-min run. Choose the Method A packed-column variant if vacuum extraction is mandated, or the backflush/PLOT variant with methanizer bypass when trace olefins are the analytical target.

What adjacent system parts and supplies must be specified alongside the 8890 GC and 8697 sampler?

The base system is the 8890 GC, 8697 headspace sampler and OpenLab CDS; columns, methanizer (where fitted), sampler vials and septa, calibration standards and carrier/fuel gases are separate. Confirm gas plumbing and electrical supply against the 8890 and 8697 site-preparation manuals, since the brochure digests give no utility data.

What duty-cycle and batch-size limits should I plan around for unattended operation?

The maximum sample sequence is 120 samples at 22.5 min per run, about 45 hours of continuous operation. Plan vial tray loading, calibration checks and operator intervention within that window; the available digests give no further environmental or duty-cycle limits, so consult the full manuals for bench conditions.

What maintenance items and consumables are not included in the base delivery?

Columns, methanizer, sampler vials and septa, calibration gas standards and carrier/detector gases are not part of the base 8890 GC + 8697 + OpenLab CDS package. The supplied document digests contain no maintenance schedule or service intervals, so replacement intervals must be taken from the full instrument manuals.