Agilent Technologies Cary UV-Vis Instruments, Cary UV-Vis-NIR Instruments
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies Cary UV-Vis Instruments, Cary UV-Vis-NIR Instruments

Agilent Technologies Cary UV-Vis Instruments, Cary UV-Vis-NIR Instruments — Molecular spectroscopy.

External diffuse reflectance accessory (DRA) with a 50 mm integrating sphere, configured for the Cary 5000 UV-Vis-NIR instrument. It provides diffuse/total reflection measurement with a standard clamp holder accepting 15–200 mm samples and a filter holder for 12–50 mm filters. Absorbance accuracy is verified against a certified filter (0.3688 Abs) with measured 0.3694 Abs on the Cary 7000.

Full specifications

Core Performance & Functional Specifications
Absorbance Error 0.1894 %
Certified Absorbance Value Of Filter 0.3688 Abs
Integrating Sphere Diameter 50 mm
Measured Absorbance (cary 7000) 0.3694 Abs
Sphere Size 50 mm
Setup time mfr less than five minutes minutes
Minimum sample size (w × h), filter holder mfr 12 mm
Maximum sample size (w × h), filter holder mfr 50 mm
DRA type mfr Internal Internal
Compatible instrument mfr 60 Vis
Measurement capability mfr Diffuse/total reflection reflection
Absorbance accuracy requirement below 1 Abs mfr 0.01
Absorbance accuracy requirement above 1 Abs mfr 1
Certified absorbance tolerance of filter mfr 0.0025 Abs
Standards, Safety & General Data
Maximum Sample Thickness, Standard Clamp Holder 5 mm
Minimum sample size (w × h), standard clamp holder mfr 15 mm

Description

Construction and positioning

Optical behavior and accuracy

Installation and operation

Integration & Operating Notes

Control & I/O
Compatible only with the Cary 5000 instrument

The external DRA is specified for the Cary 5000 UV-Vis-NIR; it is not listed as compatible with the Cary 60 UV-Vis (which uses its own internal 50 mm DRA) or with the 110 mm/150 mm UV-Vis-NIR DRA configurations, so instrument pairing must be confirmed before ordering.

Factory Specification Sheet; Cary 60 UV-Vis DRA technical document
Maintenance
Verify photometric accuracy with the certified reference filter

Absorbance accuracy should be checked against the certified metal-on-quartz filter (certified 0.3688 Abs, tolerance 0.0059 Abs); a measured deviation beyond the stated 0.01 Abs (below 1 Abs) or 1 % (above 1 Abs) requirement indicates the accessory or instrument needs service.

Cary 7000 NIR photometric accuracy technical overview

Selection & Integration Q&A

When should I choose this external 50 mm DRA instead of the internal 50 mm DRA or the larger 110/150 mm UV-Vis-NIR DRAs?

Pick this external 50 mm unit for the Cary 5000 when samples up to 200 mm wide and 5 mm thick must be measured outside the compartment. The internal 50 mm DRA belongs to the Cary 60 UV-Vis; the 110 mm internal and 150 mm external DRAs are the larger UV-Vis-NIR options for bigger collection geometry.

What instrument and sample-holder configuration must be in place before the accessory can be used?

It requires a Cary 5000 UV-Vis-NIR as the host instrument. Fit either the standard clamp holder (15–200 mm samples, max 5 mm thick) or the filter holder (12–50 mm filters); setup is stated at under five minutes, so no dedicated drive or controller is needed beyond the spectrophotometer.

What maintenance and reference materials are needed, and what is not included in base delivery?

Periodic photometric checks use a certified metal-on-quartz filter (certified 0.3688 Abs, tolerance 0.0059 Abs); the filter and any alternative sample holders are separate items, not part of the base accessory. Absorbance error should stay within 0.01 Abs below 1 Abs and 1 % above 1 Abs.

Dimensional Drawings & Technical Documentation

Technical documentation