BINDER KBF PRO 720
BINDER KBF PRO 720 — Climate chambers with Temperature -20…100 °C.
The KBF PRO 720 is a 727 L constant climate chamber with a -20…100 °C temperature range and 10…98 % RH humidity control, using a steam humidifier and R600A refrigeration circuit. It is the mid-large standard model in the KBF PRO series, distinguished from smaller siblings by its 40 kg-per-rack load capacity and 1,000 W max heat compensation. It serves as a programmable test environment for steady-state climatic conditioning of bulk samples and assemblies.
BINDER KBF PRO Series — Model Comparison & Selection
| Model | Temperature, °C | Interior volume, L | Load per rack, kg | Max heat compensation, W | Nominal power, kW | Temperature uniformity, K |
|---|---|---|---|---|---|---|
| KBF PRO 720 Current | -20–100 | 727 | 40 | 1000 | 1.8 | 0.3 |
| KBF PRO 1060 | -20–100 | 1070 | 40 | 1000 | 1.9 | 0.3 |
| KBF PRO 130 | -20–100 | 127 | 30 | 450 | 1.9 | 0.2 |
| KBF PRO 1600 | -20–100 | 1610 | 40 | 1000 | 1.9 | 0.3 |
| KBF PRO 260 | -20–100 | 262 | 40 | 800 | 1.6 | 0.2 |
| KBF PRO 470 | -20–100 | 472 | 40 | 800 | 1.6 | 0.3 |
Full specifications
| Temperature | -20...100 °C |
| Interior volume | 727 L |
| Load per rack | 40 kg |
| Maximum heat compensation | 1000 W |
| Nominal power | 1.8 kW |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.3 K |
| Article Number mfr | 9020 |
| Recovery time after door was opened for 30 s at 25°C and 60% RH mfr | 2 min |
| Recovery time after door was opened for 30 s at 40°C and 75% RH mfr | 5 min |
| Net weight of the unit (empty) mfr | 312 kg |
| Power frequency | 50 Hz |
| Rated voltage | 208...240 V |
| Unit fuse | 16.0 A |
| Phase (Nominal voltage) mfr | 1 |
| Humidity fluctuation | 1 % RH |
| Humidity range | 10...98 % RH |
Description
Construction and series positioning
- The KBF PRO 720 (Art. No. 9020-0442, designation 720-230) is a floor-standing chamber with 727 L interior volume and a net empty weight of 312 kg. It uses the signal-triangle instrument panel layout shared by sizes 470, 720, 1060 and 1600, with an ergonomically adjustable control terminal.
- Within the KBF PRO series, the 720 sits between the 470 and 1060 models; all siblings share the -20…100 °C range, but the 720's 40 kg rack load and 1,000 W heat-compensation ceiling define its sample-capacity envelope. The standard version is rated 208…240 V, 1~, 50 Hz, 1.8 kW nominal, with a 16.0 A unit fuse.
- Humidity control is available over 10…98 % RH, but the temperature range with humidity active is restricted to 10…90 °C; the full -20…100 °C span is dry-only.
Behavior under load and in different media
- Temperature fluctuation is 0.1 K and temperature uniformity 0.3 K; humidity fluctuation is 1 % RH. These figures apply to the controlled empty or lightly loaded chamber and should be treated as baseline performance.
- Recovery after a 30 s door opening is 2 min at 25 °C / 60 % RH and 5 min at 40 °C / 75 % RH, indicating faster re-stabilization at moderate conditions than at elevated temperature and humidity.
- The chamber uses a steam humidifier fed by freshwater, with separate wastewater/condensate handling; the humidifying system and wastewater routing differ by serial number (up to 20250000010225 vs. from 20250000010226), so spare-part and connection selection must match the unit's serial.
- Defrosting occurs during refrigerating operation, and anti-condensation protection is provided; both affect usable duty cycles at low temperatures and high humidity.
Installation, connection and control
- Sizes 260 and larger require a wall spacer to maintain the specified rear distance; the 720 therefore needs rear clearance for its refrigeration and water connections.
- The rear connection panel carries the freshwater and wastewater connections; the chamber requires a water supply and drain for humidity operation. Maximum operating pressure is 11 bar, so supply fittings must be rated accordingly.
- Control is via the MB2 chamber controller with fixed-value, time-program and week-program modes, chart-recorder display, Ethernet/E-mail communication and USB data transfer. Temperature safety is provided by a DIN 12880 class 2/3.3 overtemperature safety controller (selectable as temperature limiter or temperature protection) plus an undertemperature protection class 3.2.
- The chamber is IP 20 per EN 60529 and must not be sprayed with water.
Integration & Operating Notes
Although the chamber reaches -20…100 °C dry, humidity is limited to 10…98 % RH and the temperature range with humidity active is 10…90 °C. Programs that combine humidity with temperatures outside this band will not be controllable and may trigger humidity-system alarms.
The humidity sensor has a ±2 % RH tolerance and must not be exposed to substances above the manual's tolerated concentrations — for example ammonia 5,500 ppm, formaldehyde 2,400 ppm, hydrochloric acid 300 ppm, ozone 0.5 ppm under permanent load. Samples outgassing solvents or acids can degrade the sensor and invalidate humidity readings.
From chamber size 260 onward a wall spacer is required to hold the specified distance from the wall. The 720 needs this rear clearance for its refrigeration circuit and rear water connections; blocking it compromises heat rejection and access to the connection panel.
The chamber's maximum operating pressure is 11 bar, and the rear panel carries freshwater and wastewater connections for the steam humidifier. Supply and drain fittings, hoses and valves must be rated for at least this pressure, and the humidifier requires an annual service independent of chamber maintenance.
After 8,760 operating hours or two years the controller displays a maintenance message; confirming it only postpones it, and the window reappears every two weeks until BINDER Service resets it. Plan the service visit before the threshold to avoid recurring operator prompts.
A qualified technician must clean the condensers at least twice yearly; in dusty ambient air, suction cleaning several times a year is advised. Refrigeration work is restricted to personnel trained per EN 13313:2010 because the circuit contains flammable R600A refrigerant.
The overtemperature safety controller per DIN 12880 class 2/3.3 can be configured as temperature limiter or temperature protection, with limit and offset values and an undertemperature protection class 3.2. The chosen mode determines alarm behavior and whether the chamber shuts down or limits heating on deviation, so it must be agreed before validation runs.
Selection & Integration Q&A
Pick the 720 when 727 L interior volume and 40 kg per rack with 1,000 W heat compensation match your batch size. The 470 is smaller; the 1060 offers more volume but the same -20…100 °C range, so the decision is sample capacity and floor space, not temperature performance.
The rear panel has freshwater and wastewater connections for the steam humidifier; fittings must be rated for at least 11 bar maximum operating pressure. Connection method and humidifying-system layout depend on serial number (up to or from 20250000010225), so confirm the unit's serial before ordering plumbing parts.
The chamber is IP 20 and must not be sprayed with water. As a size-260-or-larger unit it requires the wall spacer to hold rear distance for refrigeration and water connections. Defrosting occurs during refrigerating operation, and anti-condensation protection limits continuous low-temperature/high-humidity duty.
Annual maintenance (including steam humidifier service), twice-yearly condenser cleaning and the 8,760 h/two-year service reset are not included. Optional items such as APT-COM 4 software, analog outputs, zero-voltage alarm contacts, Pt 100 object sensor, door lock, LED light and external water cans must be ordered separately.