BINDER VD 115
BINDER VD 115 — Drying ovens with Temperature 9…220 °C.
The BINDER VD 115 is a 119 L vacuum drying chamber with a 16 mm small-flange vacuum connection, operating from +9 to 220 °C with 0.1 K temperature fluctuation and 2.9 K uniformity. It is the largest model in the VD series, distinguished from the VD 23 and VD 56 by its 119 L interior volume and 70 kg permitted load. It is used for gentle drying of non-flammable solvents under vacuum in laboratory and production processes.
BINDER VD Series — Model Comparison & Selection
Full specifications
| Temperature | 9...220 °C |
| Heating time to 100°C | 130 min |
| Interior volume | 119 L (VD115-230V) |
| Leak rate | 0.01 bar/h |
| Load per rack | 20 kg (VD115-230V) |
| Nominal power | 1.6 kW |
| Permitted load | 70 kg (VD115-230V) |
| Pressure fluctuation over time | 1 mbar |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 2.9 K |
| Rear wall clearance | 100 mm (VD115-230V) |
| Side wall clearance | 70 mm (VD115-230V) |
| Article Number mfr | 9630 |
| Pressure fluctuation over time at 400 mbar mfr | 70 mbar |
| Depth | 581 mm (VD115-230V) |
| Height | 942 mm (VD115-230V) |
| Width | 743 mm (VD115-230V) |
| Power frequency | 50 Hz |
| Rated voltage | 200...230 V |
| Unit fuse | 10.0 A |
| Phase (Nominal voltage) mfr | 1 |
| Vacuum connection with small flange mfr | 16 mm |
Description
Construction and Series Positioning
- The VD 115 is a vacuum drying oven with a corrosion-resistant stainless-steel inner chamber and a 16 mm small-flange vacuum connection.
- Within the VD series, it is the largest model: VD 23 and VD 56 share the same +9 to 220 °C temperature range but have smaller interior volumes.
- The chamber is available in two voltage variants: VD115-230V (Art. 9630-0003, 200...230 V 1~ 50/60 Hz, CEE 7/7 plug) and VD115UL-120V (Art. 9630-0007, 100...120 V 1~ 50/60 Hz, NEMA 5-15 plug).
- Standard controller is the RD4 microprocessor program controller; the MB2 controller is optional.
Behavior Under Load and in Different Media
- Temperature control: set-point range +9 to 220 °C, fluctuation 0.1 K, uniformity 2.9 K (at 100 °C per DIN 12880).
- Heating-up time to 100 °C is 130 min.
- Vacuum system: leak rate 0.01 bar/h; pressure fluctuation over time 1 mbar (at 400 mbar: 70 mbar per extra spec).
- Permitted load: 70 kg total, 20 kg per rack.
- The chamber is designed for non-flammable solvents; inert gas operation and ventilation after power failure are supported.
Installation, Connection, Control and Operation
- Wall clearances: 100 mm at the back, 70 mm at the sides.
- Electrical supply: 200...230 V, 1~, 50 Hz, nominal power 1.6 kW, unit fuse 10.0 A.
- Vacuum connection: 16 mm small flange; pump module with pump stand and connection kit available as accessory.
- Controller: RD4 with Ethernet and USB interfaces; APT-COM 4 software for networking; optional analog outputs for temperature and pressure.
- Safety: overtemperature protective device class 2 (thermal switch) and overtemperature safety controller class 2 per DIN 12880.
- Operation includes evacuation, ventilation with ambient air or inert gas, and drying monitoring.
Integration & Operating Notes
The VD 115 is available as a 230 V version (200...230 V, 1~, 50 Hz) or a UL 120 V version (100...120 V, 1~, 50/60 Hz). Connecting the wrong voltage will damage the chamber. Check the type plate and use the appropriate plug (CEE 7/7 or NEMA 5-15).
Exceeding the permitted load can deform the expansion racks or damage the chamber. Distribute the load evenly and use only the delivered expansion racks; do not mix aluminum and stainless-steel racks to avoid calibration errors.
Maintenance must be performed at least once a year by authorized personnel. An annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended. Replace the door gasket only when cold to avoid damage.
The safety controller class 2 must be set to a value slightly above the maximum process temperature. If the alarm 'Safety controller' appears, the chamber has reached the safety limit; check set-points and heating system. Acknowledge alarms and allow cooling before restart.
Selection & Integration Q&A
Choose the VD 115 when you need a larger interior volume (119 L) and higher permitted load (70 kg) for drying larger batches or bulky items. The VD 56 and VD 23 offer the same temperature range but smaller volumes, suitable for smaller loads or space-constrained labs.
The chamber has a 16 mm small-flange vacuum connection. Use a pump module with pump stand and connection kit (available as accessory). For inert gas, connect to the designated inlet; ensure the gas is non-flammable and compatible with the chamber materials. Ventilation after power failure is possible.
Annual maintenance by authorized personnel and annual electrical safety test are required. Base delivery includes the chamber and factory calibration; not included: vacuum pump, pump module, connection kits, APT-COM 4 software, analog outputs, Pt 100 sensor, LED lighting, and expansion racks beyond the delivered set.