Eppendorf Concentrator plus - a Fast, yet Gentle Vacuum Centrifuge
Eppendorf Concentrator plus - a Fast, yet Gentle Vacuum Centrifuge — Vacuum concentrators.
The Eppendorf Concentrator plus is a benchtop vacuum concentrator supplied without a pump, with a fixed 1400 rpm rotor drive and a 144-sample capacity for 0.2–50 mL tubes. It is the base configuration of a three-variant line (Basic, Complete system, Complete system w/port) and requires an external vacuum source for solvent removal.
Full specifications
| Dimensions (w × D × H) | 33 × 37 × 23 cm |
| Sample Capacity | 144 |
| Vacuum | − |
| Fixed speed mfr | 1400 rpm |
| Tube volume mfr | 0.2 – 50 mL |
| Pump mfr | w/o pump |
| Product Weight | 16.5 kg |
| Voltage | 230 V |
| Max. power consumption mfr | 150 W |
| Port to connect external devices to vacuum pump mfr | no |
Description
Construction and series positioning
- The Concentrator plus is a single benchtop unit measuring 33 × 37 × 23 cm and weighing 16.5 kg, powered from 230 V and drawing up to 150 W.
- This card covers the **Basic device** configuration: no vacuum pump is included and no port for connecting external devices to the vacuum pump is provided. The sibling configurations — Complete system (with pump) and Complete system w/port (with vacuum port) — are separate variants and are not described here.
- The drive runs at a fixed 1400 rpm; there is no variable-speed control. Sample capacity is 144 tubes across a 0.2–50 mL tube volume range, using one of 15 rotor options that are ordered separately.
Working behavior
- Because the unit is supplied without a pump, vacuum generation is entirely the responsibility of the user-supplied vacuum source; the device itself provides only centrifugation at fixed speed.
- No performance curves or vacuum-level data are available for this configuration, so evaporation rates cannot be stated from the verified facts.
Installation and operation
- The brochure identifies three system configurations at 230 V, 50/60 Hz; the Basic device is the pump-less entry point of that line.
- Rotors are not part of the base delivery and must be selected for the tube format in use (0.2–50 mL).
Integration & Operating Notes
The Basic device ships without a pump and has no port for connecting external devices to the vacuum pump, so the user must provide and connect a separate vacuum source. Without it the unit can centrifuge but cannot concentrate.
Fifteen rotor options exist for the 0.2–50 mL tube range; the base delivery includes no rotor. Select the rotor for the actual tube size before commissioning, since capacity of 144 samples depends on the rotor fitted.
Selection & Integration Q&A
Choose the Basic device when a suitable vacuum pump and plumbing already exist in the lab, since it ships without a pump and has no vacuum-port connection. Pick the Complete system if you need the pump included, or the w/port variant if you must connect external devices to the pump.
A separate vacuum pump and its tubing, plus at least one of the 15 rotors sized for your 0.2–50 mL tubes. The device provides only the 1400 rpm centrifugation; vacuum generation and sample holders are outside the base delivery.
The vacuum pump, the rotor (one of 15 options for 0.2–50 mL tubes) and any tubing to the vacuum source. The Basic device has no port for connecting external devices to the pump, so pump interfacing must be arranged separately.