Eppendorf CryoCube ® F440 Series - ULT Freezer
Eppendorf ·Made in Germany · In Active Production

Eppendorf CryoCube ® F440 Series - ULT Freezer

Eppendorf CryoCube ® F440 Series - ULT Freezer — Ultra-low temperature freezers with Capacity 440 L.

The CryoCube F440h is a 440 L ultra-low temperature freezer using vacuum insulation paneling (Type C) and air cooling, with a −50 to −86 °C working range and 5 inner compartments. It is the high-efficiency VIP variant of the F440 series, distinguished from the foam-insulated F440n by its insulation construction. It serves as a storage stage for high-density sample archives at −80 °C.

Capacity
440 L

Full specifications

Core Performance & Functional Specifications
Capacity 440 L
Act Label yes
Average Daily Consumption (115 V) 6.5 kWh/day
Average Daily Consumption (208 V) −
Average Daily Consumption (230 V) 6.8 kWh/day
Chart Recorder no
Co 2 Backup System no
High-efficiency yes
Insulation vacuum insulation paneling
Ln 2 Backup System no
Nominal power 283 W
Remote Device Monitoring And Notifications (visionize ® ) VisioNize ® box enabled
Temperature range -50...-86 °C
No. of compartments mfr 5
No. of internal doors mfr 5
Max. racks per freezer mfr 20
Dimensions (W × D × H) mfr 95,5 × 89,8 × 195,0 cm / 37,6 × 35,4 × 76,8 in 1
Pull-down time to -80 °C mfr 3 h 20 min 4)
53 mm (2 in) tall boxes mfr 320
64 mm (2.5 in) tall boxes mfr 240
76 mm (3 in) tall boxes mfr 160
89 mm (3.5 in) tall boxes mfr 160
102 mm (4 in) tall boxes mfr 160
102 mm (4 in) tall boxes mfr 480
53 mm (2 in) tall boxes mfr 16
64 mm (2.5 in) tall boxes mfr 12
76 mm (3 in) tall boxes mfr 8
89 mm (3.5 in) tall boxes mfr 8
102 mm (4 in) tall boxes mfr 8
DWP mfr 24
Dimensions, Weight & Mechanical Design
Footprint (wxd) 95.5 cm
Product Weight 286 kg
Shipping Weight 327 kg
Electrical, Power & Drive Specifications
Max. Power Consumption 6800 W
Power supply mfr 230 V, 50 Hz
Interfaces, Connections & Control
Power Plug type F/type G
Access port mfr 2
Operating Conditions, Environment & Protection
Cooling Liquid 290
Noise level 51.5 dBA
Cooling type air cooling

Description

Construction and series positioning

Capacity and storage configuration

Thermal and electrical behavior

Monitoring and options

Integration & Operating Notes

Electrical
Supply circuit must absorb a 6.8 kW maximum draw

Nominal power is 283 W, but maximum power consumption is 6800 W, so the branch circuit, breaker and upstream distribution must be sized for the peak figure rather than the nominal one. The 230 V, 50 Hz supply and type F/type G plug define the socket and cord set required at the installation point.

Factory Specification Sheet
Mounting
Floor and route must carry 286 kg operating weight

The unit weighs 286 kg in operation and 327 kg when shipped, with a 95.5 cm width and 195.0 cm height. Plan the transport route, door clearances and floor loading for the shipping weight before moving the cabinet into place, and confirm the final floor can support the installed mass.

Factory Specification Sheet
Process Limits
Pull-down to −80 °C takes 3 h 20 min before loading

The cabinet requires 3 h 20 min to reach −80 °C, so loading warm samples immediately after start-up exposes them to a temperature above the setpoint for that period. Sequence the pull-down before transferring inventory, and account for recovery time after door openings in the storage protocol.

Factory Specification Sheet
Control & I/O
Remote monitoring needs a VisioNize box, not included as a sensor

The freezer is VisioNize box enabled, which provides remote device monitoring and notifications, but no chart recorder, CO2 backup or LN2 backup is fitted. If continuous local records or backup cooling are required, those must be specified as separate equipment rather than assumed present in the base unit.

Freezers Brochure CryoCube F440 series
Maintenance
Air-cooled condenser needs a defined cleaning interval

Cooling is by air, so the condenser and air path require periodic cleaning to maintain heat rejection; a clogged path raises duty cycle and power draw toward the 6800 W maximum. Two access ports are available for routing independent sensors or cables without disturbing the door seal.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose the F440h over the F440n, and when is the foam-insulated version the better fit?

The F440h uses vacuum insulation paneling (Type C) and is marked high-efficiency; the F440n uses polyurethane foam (Type A). Choose the h variant where energy use and insulation thickness matter; choose the n variant where a foam cabinet is preferred. Both share 440 L, 5 compartments and the −50 to −86 °C range.

What supply and protection must be provided for the 230 V, 50 Hz F440h?

Provide a 230 V, 50 Hz supply with a type F or type G socket. Size the circuit for the 6800 W maximum power consumption, not the 283 W nominal figure, and allow for the 6.8 kWh/day average consumption at 230 V when planning energy and backup capacity.

What installation conditions and duty limits follow from the air-cooled design and cabinet mass?

The cabinet weighs 286 kg installed and 327 kg shipped at 95.5 × 89.8 × 195.0 cm, so the floor and route must be rated accordingly. Air cooling requires unobstructed airflow around the cabinet; noise is 51.5 dBA. Pull-down to −80 °C takes 3 h 20 min, which sets the minimum lead time before loading.

What monitoring and backup equipment is not part of the base delivery?

No chart recorder, no CO2 backup system and no LN2 backup system are fitted. Remote monitoring and notifications require the VisioNize box, for which the unit is enabled. Two access ports are built in for routing external sensor or cable runs; racks and boxes are separate items.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Freezers Brochure CryoCube F440 series Zero Compromise
Brochure ·934,9 КБ ·6 pp.