Eppendorf DASGIP ® Parallel Bioreactor System
Eppendorf ·Made in Germany · In Active Production

Eppendorf DASGIP ® Parallel Bioreactor System

Eppendorf DASGIP ® Parallel Bioreactor System — Bioreactors & fermenters.

The DASGIP Parallel Bioreactor System is a 16-unit parallel cell-culture and microbial bioreactor platform built around autoclavable glass or pre-sterilized single-use vessels of 1.5–2.3 L total volume. Each vessel is driven by a direct/magnetic overhead drive at 30–1250 rpm with pitched-blade impellers, TMFC gas mixing (overlay and/or sparger), and integrated heating/cooling in the DASGIP Bioblock. It is intended for parallel process development and scale-down studies where up to 16 independent runs must be controlled from one process computer.

Full specifications

Core Performance & Functional Specifications
Application Cell culture
Bacteria/yeast/fungi yes
Drive Direct/magnetic overhead drive
Exhaust Condensation Water-cooled or liquid-free
Gas Supply TMFC; overlay and/or sparger
Impellers Pitched-blade
Insect yes
Interchangeable Vessels yes
Mammalian/animal yes
Number Of Parallel Units 16 unit
Software DASware control, other DASware optional
Speed Ranges 30...1250 rpm
Stem Cells yes
Sterilization Autoclavable/ pre-sterilized
Total Volume 1.5...2.3 L
Validatable no
Vessels Glass/single-use vessels
Dimensions (W × D × H, 4-fold system without process computer) mfr 140 x 70 x 53 cm (55 x 28 x 21 in)
Feed lines per vessel mfr Up to 8
pH control mfr Acid and/or base, and other set-ups
DO control mfr Cascade (O 2 concentration, gas flow rate), and
Level/foam mfr Optional
OD measurement mfr Optional (DASGIP OD4)
Exhaust analysis mfr Optional (DASGIP GA4)
Dimensions, Weight & Mechanical Design
Typical Weight 78 kg
Electrical, Power & Drive Specifications
Typical power consumption (4-fold system without process computer) mfr 480 W (at 230 V)/ 432 W (at 115 V) (595 W/552 W
Interfaces, Connections & Control
User Interface Process computer with monitor
Standards, Safety & General Data
Standard Feed Rates (depending On Tube Diameter) 0.3...9.5 L/h
Standard Gas Flow Rates 0.1...50
Standard Gas Mixing 2
Standard temperature range (at 25° C room temp.) mfr 5 K above cooling agent temperature – 99°C (DASG

Description

Architecture and positioning

Behavior under load and in different media

Control, installation and operation

Integration & Operating Notes

Electrical
Plan 480 W nominal power per system, not per vessel

The nominal power of 480 W is stated for the system. When sizing the laboratory circuit and UPS, account for the process computer and monitor in addition to the bioreactor frame, and confirm the supply voltage from the nameplate since the supplied facts do not state it.

Factory Specification Sheet, Nominal power
Process Limits
Feed rate is fixed by the installed tube diameter

Standard feed rates of 0.3–9.5 L/h depend on the tube diameter, so the peristaltic pump tubing must be selected against the required feed rate before the run. Up to 8 feed lines per vessel are available, allowing separate acid, base and nutrient streams.

Factory Specification Sheet, Standard feed rates depending on tube diameter
Control & I/O
OD, exhaust gas and level/foam need optional modules

OD measurement (DASGIP OD4), exhaust analysis (DASGIP GA4) and level/foam detection are optional and not part of the base delivery. Specify them at order time if biomass or off-gas data are required, because they add sensors and interfaces to each vessel.

Factory Specification Sheet, Optional measurements
Maintenance
Vessel preparation differs for glass and single-use formats

Glass vessels are autoclavable and single-use vessels are pre-sterilized; the choice determines the sterilization workflow and the consumable list. The supplied facts contain no calibration or service-interval schedule, so intervals must be taken from the DASGIP operating manual.

Factory Specification Sheet, Sterilization; Brochure scope and variants

Selection & Integration Q&A

When should I choose the 16-unit DASGIP system rather than a 4- or 8-fold DASGIP or a DASbox system?

Choose 16-fold DASGIP when 1.5–2.3 L vessels and up to 16 parallel runs are needed for process development. DASbox covers 60–250 mL working volume for smaller scale-down; 4- and 8-fold DASGIP frames suit lower throughput with the same vessel size.

What must be provided on the utility side for the gas, feed and cooling connections?

TMFC gas supply with overlay and/or sparger at 0.1–50 standard gas flow, up to 8 feed lines per vessel with tubing sized for 0.3–9.5 L/h, and a cooling agent cold enough that the setpoint stays at least 5 K above it up to 99 °C.

Which items are optional or consumable and not included in the base delivery?

OD4 measurement, GA4 exhaust analysis, level/foam detection and additional cooling fingers are optional. Glass versus single-use vessels are a configuration choice, and the supplied facts list no cables or calibration kits, so these must be confirmed against the order configuration.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Fermentors Bioreactors Brochure Parallel Bioreactor Systems Cell Cultu
Brochure ·2,3 МБ ·4 pp.
View PDF