Eppendorf Eppendorf Top Buret - Bottle-Top Burette
Eppendorf Eppendorf Top Buret - Bottle-Top Burette — Pipettes.
The Eppendorf Top Buret is a mechanical bottle-top burette using positive displacement, with a 0.01–999.9 mL range and 2.5 mL per-turn dispensing increment. It mounts directly on GL 45 laboratory bottles via included adapters (GL 32, GL 38, S 40), and is intended for precise, repetitive liquid dispensing in laboratory workflows.
Full specifications
| Autoclavable | no |
| Open System (third Party Consumables) | yes |
| Operating Mode | mechanical |
| Pipetting Type | Positive displacement system |
| Pipetting volume | 0.01...999.9 mL |
| Random Measurement Error 1) | 2.5 mL |
| Systematic Measurement Error 1) | 2.5 mL |
| Dispensing increment mfr | 2.5 mL per turn |
| Delivery content mfr | Burette, Battery LR03/AAA 1.5V (x2), Operating m |
| Number Of Channels | 1 |
| Battery Type | AAA battery |
| Thread mfr | GL 45 |
| Bottle thread adapter incl. mfr | GL 32, GL 38, S 40 2) |
Description
Construction and positioning
- The Top Buret is a bottle-top burette that threads onto a GL 45 bottle; three adapters (GL 45/S 40, GL 45/GL 38, GL 45/GL 32) are included to fit common laboratory bottle necks.
- It is a single-channel, mechanical (non-electronic) positive-displacement dispenser; the only electrical component is the battery-powered display/control (two LR03/AAA 1.5 V batteries supplied).
- Delivery includes the burette, canula arm, discharge tube with knurled nut, tube holder, telescopic aspirating tube, and the three thread adapters; no sibling models are defined in the verified data, so series positioning cannot be stated.
Dispensing behavior
- Volume range is 0.01–999.9 mL; each full turn of the dispensing control delivers 2.5 mL, so fine increments are achieved by partial turns.
- The stated random and systematic measurement error are both 2.5 mL (as given); this is a large error relative to small volumes, so the unit is suited to larger-volume dispensing rather than microliter work.
- Positive displacement means the dispensed liquid contacts the internal flow path; the unit is not autoclavable, so sterility must be maintained by other means.
Installation and operation
- Mount the burette on a GL 45 bottle using the appropriate adapter; the telescopic aspirating tube adjusts to bottle depth.
- The discharge tube with knurled nut and canula arm direct liquid to the receiving vessel; the tube holder stores the discharge tube when not in use.
- No PDF digest provides installation, control, or maintenance specifics for this model; the only attached document covers pipette calibration services and does not apply to the Top Buret.
Integration & Operating Notes
The burette threads onto GL 45; bottles with GL 32, GL 38, or S 40 necks require the corresponding included adapter. Verify the bottle thread before installation, because an unmatched neck will not seal and will prevent proper aspiration.
The unit is mechanical with a battery-powered display/control; two LR03/AAA 1.5 V batteries are included. Plan for battery replacement as a consumable, and note that no external power supply or controller interface is specified.
Because both error figures are given as 2.5 mL, dispensing small volumes (e.g., below a few milliliters) will carry a large relative uncertainty. Use this burette for larger-volume work where a 2.5 mL error is acceptable, and do not assume microliter-level accuracy.
The autoclavable spec is 'no', so do not autoclave the burette. Cleaning and decontamination must follow the operating manual's chemical-compatibility guidance; the attached services brochure covers pipette calibration only and does not define a service interval for this burette.
Selection & Integration Q&A
Choose it for repetitive dispensing from a GL 45 bottle in the 0.01–999.9 mL range with 2.5 mL per-turn resolution. If you need microliter accuracy or autoclavable wetted parts, this unit is not suitable; no sibling models are defined in the verified data for comparison.
It mounts on a GL 45 bottle; included adapters cover GL 32, GL 38, and S 40 necks. The telescopic aspirating tube adjusts to bottle depth, and the discharge tube with knurled nut connects to the canula arm. No external drive, controller, or power supply is required beyond two AAA batteries.
The unit is not autoclavable and uses a positive-displacement path, so it must be mounted upright on a matching bottle thread and kept within the chemical compatibility of its wetted parts. No duty-cycle or environmental limits are stated in the verified facts or the attached services brochure.
Base delivery includes the burette, two AAA batteries, canula arm, discharge tube with knurled nut, tube holder, telescopic aspirating tube, and three thread adapters. Batteries are consumable; no calibration or service interval for this burette is defined in the attached services brochure, which covers pipette calibration only.