Eppendorf New Brunswick Innova ® 42/42R - Stackable Incubator Shaker
Eppendorf New Brunswick Innova ® 42/42R - Stackable Incubator Shaker — Laboratory shakers.
The Innova 42/42R is a stackable incubator shaker with a triple-eccentric counterbalanced drive and solid-state DC brushless motor, offered in non-refrigerated (42) and refrigerated (42R) configurations. It covers 10–80 °C with 0.25 °C uniformity and 25–400 rpm (25–300 rpm for flasks >500 mL when stacked). It serves as a benchtop, under-bench or stacked incubation platform for temperature-controlled shaking.
Full specifications
| Audible And Visual Alarms | yes |
| Available Program Modes | -232 |
| Benchtop (b), Under Bench (u), Floor (f) Or Stackable (s) | B, U, S |
| Door Opening | opening to front |
| Drive Mechanism | triple-eccentric counterbalanced drive |
| Gassing Manifold | yes |
| Internal Convenience Light | yes |
| Memory | non volatile with automatic power failure restart |
| Motor Type | Solid state, DC brushless motor |
| Multi-step Programming | yes |
| Photosynthetic Lighting | no |
| Platform Size (cm/in) | 46 cm |
| Refrigerated | − |
| Slide-out Platform | no |
| Temperature range | 10...80 °C |
| Temperature uniformity | 0.25 °C |
| Timer mfr | 0.1 – 99.9 h, continuous |
| Dimensions (W × D × H) mfr | 63,5 × 74,5 × 81,8 cm / 25 × 29,3 × 32,8 in |
| In chamber power receptacle mfr | yes |
| Speed range mfr | 25 – 400 rpm; 25 – 300 rpm when using flasks > 5 |
| Steps per program mfr | 15 |
| Stored programs mfr | 4 |
| Product Weight | 98.5 kg |
| External depth (with lid/door open) mfr | 131 cm (51.6 in) |
| Power supply mfr | 230 V ± 10 %, 50 Hz |
| Power Plug | type F/type I |
| Water/humidity Monitoring | yes |
Description
Construction and series positioning
- The Innova 42/42R is a stackable incubator shaker (B/U/S: benchtop, under-bench or stacked) with a front-opening door and a 46 cm platform.
- The drive is a triple-eccentric counterbalanced design driven by a solid-state, DC brushless motor; the unit weighs 98.5 kg and measures 63.5 × 74.5 × 81.8 cm (W × D × H), with 131 cm external depth when the lid/door is open.
- The series includes refrigerated (R) and non-refrigerated variants; this model is listed as non-refrigerated in the matched specs, while the brochure portfolio covers both 42 and 42R. No sibling models were supplied for direct comparison.
Behavior and operating envelope
- Temperature range is 10–80 °C with 0.25 °C uniformity; no photosynthetic lighting is provided.
- Speed range is 25–400 rpm, reduced to 25–300 rpm when using flasks >500 mL in stacked devices.
- The timer spans 0.1–99.9 h plus continuous operation; multi-step programming supports 15 steps per program and 4 stored programs.
- Memory is non-volatile with automatic power-failure restart.
Control, connection and operation
- Power supply is 230 V ± 10 %, 50 Hz; power plug type F or type I.
- Alarms are audible and visual; an internal convenience light and an in-chamber power receptacle are provided.
- A gassing manifold and water/humidity monitoring are available; the platform does not slide out.
- The supplied brochure digest contains no additional actionable engineering facts beyond the portfolio scope.
Integration & Operating Notes
The external depth with lid/door open is 131 cm, so the installation envelope must accommodate this swing plus service access; benchtop, under-bench and stacked configurations are supported, but the 98.5 kg unit weight requires a load-rated surface or stacking frame.
When the device is stacked and flasks larger than 500 mL are used, the speed range is limited to 25–300 rpm instead of 25–400 rpm; program setpoints above 300 rpm must be avoided in that configuration to prevent exceeding the stated envelope.
The controller stores programs in non-volatile memory and performs an automatic power-failure restart; engineers should account for this behavior in protocols where an unintended restart after mains interruption is not acceptable, and use the audible/visual alarms to detect fault states.
The supplied factory specification and brochure digest do not define service intervals, consumables or accessory cables; these must be obtained from the full instruction manual before commissioning, and the in-chamber power receptacle and gassing manifold should be checked as part of site acceptance.
Selection & Integration Q&A
Choose this model when you need a stackable, front-door incubator shaker covering 10–80 °C with 0.25 °C uniformity and up to 400 rpm. The supplied facts list it as non-refrigerated; the portfolio also covers refrigerated (R) variants, so confirm the R configuration if sub-ambient control is required.
It requires 230 V ± 10 %, 50 Hz with a type F or type I plug. An in-chamber power receptacle and gassing manifold are provided for ancillary devices; no external drive or controller is needed because the solid-state DC brushless motor and non-volatile program memory are integral.
The supplied documents state no service intervals, consumable list or included cable set; only the type F/type I power plug is specified. Request the full instruction manual for maintenance scheduling and confirm accessory scope (platform, gassing manifold connections) before ordering.