Peter Huber Kältemaschinenbau CC-304B Wärme-Badumwälzthermostat
Peter Huber Kältemaschinenbau CC-304B Wärme-Badumwälzthermostat — Thermostatic baths.
The CC-304B is a 5-litre stainless-steel bath circulator with a variable-speed pressure/suction pump and Pilot ONE controller, rated for 28–300 °C without cooling (20–300 °C with water cooling, -20–300 °C with refrigerator). Its 0.02 K stability and up to 3 kW heating distinguish it from the larger CC-308B and CC-315B siblings. It serves as a precise temperature source for external circuits or direct bath immersion.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Mit Edelstahlbaedern Bis 300C Series — Model Comparison & Selection
| Model | Current consumption, A | Dimensions w d h | Fuse rating, A | Weight, kg | Ambient temperature limit, °C | Controller |
|---|---|---|---|---|---|---|
| CC-304B Wärme-Badumwälzthermostat Current | 14 | 210 x 335 x 392 | 16 | 14 | 5–40 | Pilot ONE |
| CC-308B Wärme-Badumwälzthermostat | 14 | 242 x 404 x 392 | 16 | 18 | 5–40 | Pilot ONE |
| CC-315B Wärme-Badumwälzthermostat | 16 | 335 x 382 x 433 | 20 | 22 | 5–40 | Pilot ONE |
Full specifications
| Bath volume | 5 l |
| Dimensions (w × D × H) | 210 x 335 x 392 mm |
| Maximum delivery flow rate | 25 l/min (pressure pump; 0, 7 bar) |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | 28...300 °C with cooling unit: -20...300 °C · with water cooling: 20...300 °C |
| Temperature stability mfr | ±0, 02 K |
| Heating capacity mfr | 2, 2 - 3 kW |
| Delivery capacity suction pump max. mfr | 18, 5 l/min (suction pump; 0, 4 bar) |
| Bath opening WxD/BD mfr | 130 x 100 / 155 mm |
| Filling volume min. mfr | 3, 2 l (min) |
| Weight | 14 kg |
| Current consumption | 14 A (max) |
| Fuse rating | 16 A |
| Power supply | 200-240V 1~/2~ 50/60Hz V; Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Stainless-steel bath with 5 l volume and 130 x 100 mm opening (155 mm depth); minimum filling 3.2 l.
- Variable-speed pressure and suction pump: 25 l/min at 0.7 bar delivery, 18.5 l/min at 0.4 bar suction; M16x1 male connections.
- Pilot ONE controller with Ethernet, USB host/device and RS232; Pt100 external sensor input.
- Within the series, the CC-304B is the 5-litre entry size; the CC-308B and CC-315B offer larger bath volumes for greater thermal inertia or immersion capacity.
Thermal behaviour and media
- Base range 28–300 °C; water cooling extends the low end to 20 °C, and a refrigerator to -20 °C. The cooling method must be chosen at selection because it sets the achievable setpoint envelope.
- Heating power is voltage-dependent: 3 kW at 230–240 V, 2.8 kW at 220 V, 2.5 kW at 208 V, 2.2 kW at 200 V. Stability is 0.02 K at 70 °C.
- Bath fluid kinematic viscosity must not exceed 50 mm²/s at operating temperature; this constrains thermofluid choice at the low end of the range.
Installation and control
- Supply is 200–240 V, 1~ or 2~, 50/60 Hz, max 14 A, 16 A fuse; external UL 489 branch protection required. Voltage and frequency tolerances are ±10% only if not opposing.
- IP20 enclosure requires a dry location; ambient 5–40 °C. Country-specific plug must be specified at order.
- Adjustable overtemperature protection per DIN 12876 must be set for the fluid and application during commissioning.
Integration & Operating Notes
The unit draws up to 14 A at 200–240 V; the supply circuit must be protected by a 16 A fuse and, for UL markets, a UL 489-listed breaker. Undersized wiring or a lower-rated breaker will trip during heat-up at full heating power.
Without cooling the bath cannot go below 28 °C; water cooling lowers the limit to 20 °C, and only a refrigerator reaches -20 °C. The cooling utility must be decided before specifying the unit, because it changes the usable setpoint range and the required site services.
The pump is rated for a maximum kinematic viscosity of 50 mm²/s. At low temperatures many thermofluids become too viscous, causing cavitation, poor flow and possible pump overload; select a fluid whose viscosity at the lowest operating temperature remains within this limit.
Single- and two-phase units are delivered with a power cable and a country-specific plug that must be specified at ordering. Without this, the unit may arrive with a plug incompatible with the site socket, delaying commissioning.
The unit includes adjustable overtemperature protection per DIN 12876. It must be configured for the specific bath fluid and application during commissioning; an incorrect setting either trips prematurely or fails to protect the fluid and system if control is lost.
Selection & Integration Q&A
Pick the CC-304B for a 5-litre bath when external circuit volume and immersion load are small; the CC-308B and CC-315B offer larger baths for greater thermal stability under heavy external loads or larger immersed samples.
Pump ports are M16x1 male, so hoses and adapters must match this thread. The bath fluid must have a kinematic viscosity no higher than 50 mm²/s at the lowest operating temperature to avoid pump overload.
Supply is 200–240 V, 1~ or 2~, 50/60 Hz, drawing up to 14 A; provide a 16 A fuse and, for UL markets, a UL 489-listed breaker. Voltage and frequency tolerances of ±10% apply only if not opposing.
The base unit includes a power cable with country-specific plug only if specified at order. Bath fluid, hoses with M16x1 fittings, and any cooling-water or refrigerator circuit are not included and must be sourced separately; overtemperature protection must be set at commissioning.