Labconco Purifier Class I Biosafety Enclosures
Labconco · In Active Production

Labconco Purifier Class I Biosafety Enclosures

Labconco Purifier Class I Biosafety Enclosures — Biological safety cabinets.

The Purifier Class I Filtered Enclosure is a benchtop containment unit with an integral motorized impeller that draws room air past the operator and pushes it through a HEPA filter rated at least 99.99% efficient for 0.3-micron particles. It is available in 2', 3' and 4' widths and is distinguished within the series by the availability of a UV lamp for surface decontamination. It is used for containment of toxic, noxious and biohazardous particulates in laboratory procedures.

Full specifications

Dimensions, Weight & Mechanical Design
Shipping weight mfr 621 lbs.
Electrical, Power & Drive Specifications
Current 12 A
Energy Requirements 150 W

Description

Construction and positioning

Behavior under load and in different modes

Installation, connection and operation

Integration & Operating Notes

Process Limits
Face velocity must be set within 75-105 fpm for biohazardous work

Face velocity for biohazardous operations should be 75-105 fpm; for weighing procedures 60-100 fpm. Set speed control within these ranges; lowest appropriate velocity gives quietest operation. Face velocity is measured across the sash opening in two rows: 3" down from the upper sash foil and 3" up from the work surface, with readings 6" from the ends and every 12". Four readings are taken for the 2' enclosure, six for the 3' and eight for the 4', then averaged.

User Manual, Chapter 3 Validating the Vented Enclosure
Maintenance
HEPA filter leak check required before use and at least annually

HEPA filters ship installed with the gasket on the downstream side and are leak-checked at Labconco; a second leak check is required before use and at least annually thereafter. Schedule leak testing as part of commissioning and routine maintenance. Chapter 6 covers the routine maintenance schedule, decontamination, determination of when to replace HEPA and carbon filters, HEPA filter leak testing, setting inflow face velocity with the speed control, calibrating and operating the airflow monitor, calculating carbon filter life, initial certification and re-certification, and replacement of fluorescent/UV lamps, motorized impeller and speed control.

User Manual, Chapter 6 Maintaining Your Filtered Enclosure

Selection & Integration Q&A

When should I choose the Purifier Class I Filtered Enclosure over other models in the Purifier series?

Choose the Class I when you need UV surface decontamination, as the UV lamp is available only on this model. For applications requiring only HEPA filtration without UV, the Purifier HEPA Filtered Enclosure is sufficient. The Class I is available in 2', 3' and 4' widths.

What are the requirements for connecting the enclosure to a house exhaust system?

Use a thimble exhaust connection (part numbers 3924400 for 2', 3924401 for 3', 3924402 for 4') and install an adjustable damper or valve to control airflow. The ductwork weight must be supported independently of the enclosure, and the duct material must match the enclosure, procedures and chemical applications.

What maintenance schedule and accessories are required that are not included in the base delivery?

HEPA filter leak check before use and at least annually. Carbon filters (if used) are gravity-installed and must be replaced when exhausted; calculate life per Chapter 6. Thimble exhaust connections, remote blowers, adjustable dampers, UV lamp (Class I only), work surfaces, utility shelves, and Guardian Airflow Monitor are optional accessories not included in base delivery.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Logic and Axiom Brochure
Brochure ·14,3 МБ ·48 pp.
View PDF
PDF
Purifier Class I and HEPA Filtered Safety Enclosures Manual (2012 to Present)
Manual ·3,2 МБ ·77 pp.
View PDF