Shimadzu Aimsight
Shimadzu ·Made in Japan · In Active Production

Shimadzu Aimsight

Shimadzu Aimsight — Laboratory balances.

The AIMsight is an automatic contaminant analysis system integrated with the IRXross Fourier-transform infrared microscope, forming a turnkey IRXross + AIMsight configuration. It combines a standard T2SL detector, ATR objective mirror, pressure sensor, and wide-field camera to automate particle identification. This modification is distinguished by its 40 mm maximum sample thickness for reflectance/ATR and spectral coverage from 4000–400 cm⁻¹ in ultra-micro analysis, serving as a dedicated defect-review station in materials and electronics failure analysis.

Full specifications

Core Performance & Functional Specifications
Maximum sample thickness for reflectance/ATR measurement (lower condenser mirror removed) mfr 40 mm (lower condenser mirror removed)
Spectral range (ultra micro analysis) mfr 4000-400 cm-1 (ultra micro analysis)
Spectral range (high-sensitivity ATR measurement) mfr 4000-1000 cm-1 (high-sensitivity ATR measurement)
Detector mfr standard T2SL detector
ATR objective mfr ATR Objective Mirror
Camera mfr Wide-field camera
Software mfr AMsolution
System configuration mfr IRTracer-100 + AIMsight
Standards, Safety & General Data
Spectral range (standard T2SL detector) mfr 4000-500 cm-1 (standard T2SL detector)

Description

Construction and series positioning

The AIMsight is not a standalone balance or generic IR microscope; it is a factory-integrated automation layer for the IRXross platform. The verified configuration is explicitly "IRXross + AIMsight," meaning the IRXross base spectrometer and the AIMsight automated analysis head are delivered as a matched pair. Within the broader AIMsight family, the same automation concept is also offered on the IRTracer-100 base, but this card covers only the IRXross-based variant. The optical train includes a standard T2SL detector, an ATR objective mirror, an ATR pressure sensor, a grazing-angle objective, and a wide-field camera, all under software control via AMsolution (with LabSolutions IR and AIMsolution also referenced in the system documentation).

Working range and optical behavior

Spectral coverage depends on the measurement mode: 4000–400 cm⁻¹ for ultra-micro analysis, 4000–1000 cm⁻¹ for high-sensitivity ATR, and 4000–500 cm⁻¹ with the standard T2SL detector. These ranges are not interchangeable; the choice of objective and detector path determines the usable window. The 40 mm maximum sample thickness for reflectance/ATR is achieved only with the lower condenser mirror removed, which is a mechanical reconfiguration that must be planned before mounting thick or irregular specimens. No digitized curve points were supplied, so no load- or wavelength-dependent response curves are asserted here.

Installation, control, and operation

Operation is software-centric: AMsolution drives automated particle location and analysis, while the wide-field camera provides visible-light context for the IR measurement. The ATR pressure sensor is part of the hardware set, indicating that contact pressure during ATR measurement is monitored rather than left to operator feel. The documentation digests do not contain wiring, power, or I/O schedules, so electrical and interface details must be taken from the IRXross base instrument manual and the AIMsight installation guide rather than inferred here.

Integration & Operating Notes

Mounting
Thick samples require condenser mirror removal

The 40 mm maximum sample thickness for reflectance/ATR measurement is valid only when the lower condenser mirror is removed. Attempting to load a sample thicker than the clearance with the mirror in place risks collision with the objective and misalignment of the optical path. Verify clearance before every thick or mounted specimen.

Factory Specification Sheet, extra_specs: Maximum sample thickness for reflectance/ATR measurement
Process Limits
Spectral range depends on selected measurement mode

Ultra-micro analysis covers 4000–400 cm⁻¹, high-sensitivity ATR covers 4000–1000 cm⁻¹, and the standard T2SL detector path covers 4000–500 cm⁻¹. A feature that is visible in one mode may fall outside another, so the measurement mode must be chosen before interpreting spectra; do not assume full 4000–400 cm⁻¹ coverage in ATR.

Factory Specification Sheet, extra_specs: Spectral range conditions
Control & I/O
Automated analysis is driven by AMsolution software

Particle location, camera imaging, and IR acquisition are coordinated through AMsolution, with LabSolutions IR and AIMsolution also named in the system documentation. The software stack is therefore part of the integration scope; plan for compatible PC hardware and software licensing in addition to the instrument itself.

Document digest c103-e142, scope_and_variants
Maintenance
Particle Filter Holder is a separate option

The Particle Filter (PF) Holder and the Particle Analysis Program are documented as options for the AIMsight and AIRsight platforms, not as base-delivery items. Sites intending to capture and analyze particles on filters must order these accessories explicitly; they are not included with the IRXross + AIMsight configuration.

Document digest c103-e154, scope_and_variants

Selection & Integration Q&A

When should I choose the IRXross + AIMsight configuration instead of the IRTracer-100 + AIMsight variant?

Choose the IRXross-based AIMsight when the IRXross spectrometer platform is already your standard or when its optical bench matches your sample types. The IRTracer-100 + AIMsight variant is the alternative base; the automation head and AMsolution software concept are shared, so the decision hinges on the base spectrometer, not the AIMsight layer.

What adjacent system parts and accessories are required beyond the base IRXross + AIMsight delivery?

The base configuration includes the T2SL detector, ATR objective mirror, ATR pressure sensor, grazing-angle objective, wide-field camera, and AMsolution software. The Particle Filter (PF) Holder and Particle Analysis Program are separate options (document c103-e154) and must be ordered if filter-based particle capture is needed.

What mounting and sample-size limits apply when measuring thick or irregular specimens?

Reflectance/ATR measurement accepts samples up to 40 mm thick only with the lower condenser mirror removed. With the mirror installed, clearance is reduced, so thick specimens must be planned around this reconfiguration to avoid objective collision and optical misalignment.

What maintenance and consumable items are not included in the base delivery?

The Particle Filter Holder and Particle Analysis Program are documented options, not base items. No consumable schedules or service intervals are stated in the supplied digests; refer to the AIMsight and IRXross manuals for maintenance intervals and any replacement parts.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
c103-e142
Datasheet ·5,7 МБ ·16 pp.
View PDF
PDF
c103-e154
Datasheet ·212,4 КБ ·2 pp.
View PDF