Shimadzu Airsight
Shimadzu ·Made in Japan · In Active Production

Shimadzu Airsight

Shimadzu Airsight — Laboratory balances.

The Shimadzu AIRsight is a combined infrared and Raman microscope system built around a germanium ATR objective (P/N 206-32600-41) and a 532 nm Raman laser, delivering 4 cm⁻¹ wavenumber resolution for both IR and Raman and a minimum spatial resolution of 3 µm. It is configured with either a TEC MCT or DLATGS detector and supports microplastic analysis from a few to several tens of micrometres in diameter, including 1 µm polystyrene beads by Raman. The instrument is intended for materials and particle identification in laboratory microscopy workflows where co-located IR and Raman data are required from the same sample area.

Full specifications

Core Performance & Functional Specifications
Wavenumber resolution (IR and Raman) mfr 4 cm-1 (IR and Raman)
Spatial resolution (Minimum) mfr 3 μm (Minimum)
Laser wavelength (Raman) mfr 532 nm (Raman)
Raman spectral range mfr 1750–750 cm-1
Detector type mfr TEC MCT (IR)
ATR cell fiber diameter (C II cell) mfr 13 mm (C II cell)
Microplastic measurement size range (diameter) mfr a few to several tens μm (diameter)
Microbead Raman measurement diameter (polystyrene) mfr 1 μm (polystyrene)
DLC I(D)/I(G) ratio (CH4_center) mfr 0.32 (CH4_center)
DLC I(D)/I(G) ratio (CH4_periphery) mfr 0.32 (CH4_periphery)
DLC I(D)/I(G) ratio (C2H2_center) mfr 0.34 (C2H2_center)
DLC I(D)/I(G) ratio (C2H2_periphery) mfr 0.34 (C2H2_periphery)
DLC FWHM(G) (CH4_center) mfr 182.17 (CH4_center)
DLC FWHM(G) (CH4_periphery) mfr 181.40 (CH4_periphery)
DLC FWHM(G) (C2H2_center) mfr 190.85 (C2H2_center)
DLC FWHM(G) (C2H2_periphery) mfr 190.25 (C2H2_periphery)
DLC log(N(G)/I(G)) (CH4_center) mfr -0.29 (CH4_center)
DLC log(N(G)/I(G)) (CH4_periphery) mfr -0.28 (CH4_periphery)
DLC log(N(G)/I(G)) (C2H2_center) mfr -0.44 (C2H2_center)
DLC log(N(G)/I(G)) (C2H2_periphery) mfr -0.44 (C2H2_periphery)
Dimensions, Weight & Mechanical Design
Microplastic IR sample length (major axis) mfr 115 μm (major axis)
Microplastic IR sample width (minor axis) mfr 53 μm (minor axis)
Materials, Media & Process Compatibility
ATR objective prism material mfr Ge (P/N 206-32600-41)

Description

Construction and positioning

Behaviour across sample types

Installation and operation

Integration & Operating Notes

Configuration
Detector choice fixes the IR configuration at order time

The AIRsight is documented in two configurations, IRXross + AIRsight and IRTracer-100 + AIRsight, and with either a TEC MCT or a DLATGS room-temperature detector. The detector and spectrometer pairing must be decided before order because they determine the IR sensitivity and cooling requirements of the installed system.

c103-e139 system configuration document
Process Limits
Raman coverage stops at 750 cm⁻¹ on the low-wavenumber side

The 532 nm Raman channel covers 1750–750 cm⁻¹, so bands below 750 cm⁻¹ are outside the Raman range and must be obtained by the IR channel. Method development should assign low-wavenumber features to IR rather than expecting Raman data there.

Factory Specification Sheet
Maintenance
Ge ATR objective is a replaceable part with a defined part number

The ATR objective uses a germanium prism identified as P/N 206-32600-41. Because the prism is the contact element, its condition governs IR data quality; replacement and handling should be planned against this part number rather than as a generic objective service.

c103-e139 accessories list

Selection & Integration Q&A

When should I choose the AIRsight rather than a separate IR or Raman instrument?

Choose the AIRsight when the same sample area must be measured by both IR and Raman at 4 cm⁻¹ resolution and 3 µm minimum spatial resolution. No sibling variants are listed, so the main selection decision is the detector (TEC MCT vs DLATGS) and the host spectrometer (IRXross or IRTracer-100).

What host equipment and accessories must be specified alongside the microscope?

The AIRsight is configured with either an IRXross or an IRTracer-100 FTIR spectrometer, one detector type (TEC MCT or DLATGS), the Ge ATR objective P/N 206-32600-41, and optionally Sadtler spectral libraries. These are ordered as a matched set rather than added later.

What sample-size and spectral-range limits should be checked before committing a method?

Raman covers 1750–750 cm⁻¹ at 532 nm, so lower-wavenumber bands need the IR channel. Spatial resolution is 3 µm minimum; microplastics are quoted from a few to several tens of micrometres in diameter, and 1 µm polystyrene beads are measurable by Raman.

What is supplied with the base system and what must be added separately?

The base delivery is the microscope with its Ge ATR objective and the chosen detector and spectrometer. Sadtler spectral libraries are an accessory listed separately. The supplied documents contain no electrical, environmental or maintenance schedules, so service intervals and cabling must be confirmed against the instrument manual.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Learn more
Datasheet ·10,1 МБ ·1 pp.
View PDF
PDF
c103-e139
Datasheet ·9,5 МБ ·16 pp.
View PDF