Shimadzu ICPE-9800 Series
Shimadzu ·Made in Japan · In Active Production

Shimadzu ICPE-9800 Series

Shimadzu ICPE-9800 Series — Elemental analyzers.

The ICPE-9800 Series is an inductively coupled plasma optical emission spectrometer (ICP-OES) configured with a mini-torch and either axial-only (ICPE-9810) or axial/radial (ICPE-9820) plasma viewing. It covers 167–800 nm with an Echelle CCD detector, achieving ppb-level detection limits and a 5–6 digit dynamic range. The 27 MHz, 1.6 kW RF supply sustains the plasma for multi-element trace analysis in environmental, metallurgical, and petrochemical laboratories.

Full specifications

Core Performance & Functional Specifications
Wavelength range 167 to 800 nm
Spectrometer / detector mfr Echelle semiconductor detector (CCD)
Electrical, Power & Drive Specifications
High-frequency power supply mfr 27 MHz

Description

Construction and series positioning

Integration & Operating Notes

Electrical
Requires stable 1.6 kW RF supply with dedicated circuit

The 27 MHz, 1.6 kW max high-frequency generator draws significant current during plasma ignition and steady-state operation. Verify that the laboratory circuit can deliver the required voltage and current without sharing with other high-draw instruments, and allow for inrush current at startup to avoid breaker trips.

Factory Specification Sheet
Process Limits
Argon supply must be continuous and of high purity

Plasma stability and detection limits depend on a uninterrupted argon flow at the specified purity (typically ≥99.999%). Install a two-stage regulator with sufficient capacity and a backup supply or manifold to prevent plasma extinction during a run, which would require a full restart and recalibration.

Factory Specification Sheet
Mounting
Benchtop placement with exhaust ventilation required

The instrument must be placed on a sturdy bench with adequate clearance for the mini-torch and sample introduction system. Plasma operation generates ozone and heat; connect the exhaust port to a laboratory fume hood or external ventilation system to prevent ozone buildup and thermal drift.

Factory Specification Sheet
Control & I/O
Plasma ignition is interlocked with argon and exhaust sensors

The ICPE-9800 Series will not ignite the plasma unless argon pressure and exhaust flow are confirmed by internal sensors. Ensure that the argon supply and exhaust connections are established before powering up; bypassing these interlocks is not permitted and may damage the torch or detector.

Factory Specification Sheet
Maintenance
Regular torch and nebulizer cleaning extends uptime

The mini-torch and nebulizer are consumable components that require periodic inspection and cleaning to maintain detection limits. Establish a maintenance schedule based on sample load (e.g., weekly for high-throughput labs) and keep spare torches, nebulizers, and peristaltic pump tubing on hand, as these are not included in the base delivery.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose the ICPE-9810 versus the ICPE-9820?

Choose the ICPE-9810 if your work is exclusively trace-level analysis in clean matrices and you want the highest axial sensitivity at lower cost. Choose the ICPE-9820 if you routinely analyze high-matrix samples or need to measure major and trace elements in the same run without dilution, as its radial view reduces matrix effects.

What are the requirements for the argon supply and sample introduction system?

The ICPE-9800 Series requires a continuous argon supply of ≥99.999% purity at a pressure and flow rate specified in the manual (typically 0.5–0.7 MPa). The sample introduction system accepts standard peristaltic pump tubing and a concentric nebulizer; ensure that the argon line is equipped with a two-stage regulator and that the exhaust is vented to a fume hood.

What maintenance items and accessories are not included in the base delivery?

The base delivery includes the spectrometer, software, and basic installation kit. Not included: argon regulator and gas lines, exhaust ducting, peristaltic pump tubing, nebulizers, spray chambers, and replacement mini-torches. Shimadzu offers a consumables kit (part number varies by region) that includes these items; establish a spares inventory based on your sample throughput.