Shimadzu Inspexio 7000
Shimadzu ·Made in Japan · In Active Production

Shimadzu Inspexio 7000

Shimadzu Inspexio 7000 — Spectrometers.

The inspeXio 7000 is a microfocus X-ray CT system with a 225 kV/1000 µA tube and a 16-inch flat-panel detector, housed in a shielded cabinet of 2170 × 1350 × 1858 mm and 3100 kg. It supports normal, half, offset, FS, 2DCT, and CBCT scan modes, with SOD and SID axes enabling geometric magnification. The system is used for non-destructive internal inspection and dimensional analysis of industrial components.

Full specifications

Core Performance & Functional Specifications
Detector types Flat panel detector
Input resolution mfr 14 megapixels (inspeXio 7000)
Processing speed improvement (HPCinspeXio ver. 3.0) mfr 50 times faster (HPCinspeXio ver. 3.0)
Sample rotation (CT scanning) mfr 360 degrees (CT scanning)
MPR image FOV (MPR image) mfr ø207.6 mm (MPR image)
Cross-sectional image FOV (GFRP) (GFRP cross-sectional image) mfr ø20 mm (GFRP cross-sectional image)
CR image FOV (CR image) mfr 10 × 20 mm (CR image)
Oblique image FOV (Oblique image) mfr ø30 mm (Oblique image)
MPR image FOV (void analysis) (MPR image) mfr ø42 mm (MPR image)
Module type (CORE Series) mfr Option module for X-ray CT system (CORE Series)
Compatible CT systems (Microfocus X-ray CT System) mfr inspeXio 7000, SMX-225CT FPD HR Plus (Microfocus X-ray CT System)
Module variant (dual-energy scanning, NEW) mfr CORE XDEF (dual-energy scanning, NEW)
Module variant (de-noising technology, NEW) mfr CORE Xtract (de-noising technology, NEW)
Module variant (ultra-fast reconstruction) mfr CORE Boost (ultra-fast reconstruction)
Scanning technique (CORE XDEF) mfr dual-energy scanning (CORE XDEF)
Image processing (CORE Xtract) mfr proprietary algorithm eliminating only image noi (CORE Xtract)
Reconstruction (CORE Boost) mfr high-speed parallel processing (CORE Boost)
Image reconstruction function (CORE Boost) mfr ROI image reconstruction (CORE Boost)
Artifact reduction (CORE XDEF) mfr metal artifacts reduced (CORE XDEF)
Contrast optimization (CORE XDEF) mfr per material (CORE XDEF)
X-Ray Detector mfr Flat panel detector
X-Ray Detector Size mfr 16 inch
X-Ray Detector Shades of Gray mfr 16
Max. Input Resolution (for Offset Scan) mfr 14,000,000 pixels
Two-Dimensional CT mfr 4096
Cone-Beam CT mfr 4096
Version mfr 3.0
For 1024 × 1024 pixel CT images mfr 10 sec
Positioning via an Exterior Camera mfr Yes
Recommend Scanning mfr Yes
SOD Axis *1 mfr 890 mm
SID Axis *2 mfr 3 levels
CT-Z Axis mfr 300 mm
Scan Modes mfr 3
CT Data Acquisition Time mfr 10 sec
Main Unit mfr 200/220
External Leakage Dose mfr 0.5
Option mfr Operation Box. Exterior Camera Retraction Unit
Dimensions, Weight & Mechanical Design
Layout dimension (Layout and dimensional drawings) mfr 500 mm (Layout and dimensional drawings)
Shield Box Size and Mass mfr 2170
Electrical, Power & Drive Specifications
Cross-sectional image FOV (lithium-ion battery) (Lithium-ion rechargeable battery) mfr ø18.4 mm (Lithium-ion rechargeable battery)
Cross-sectional image FOV (lithium polymer battery) (Rectangular lithium polymer battery) mfr ø15 mm (Rectangular lithium polymer battery)
Rated Power mfr 135 W
Max. Tube Voltage mfr 225 kV
Max. Tube Current mfr 1000
Control Computer *6 *7 mfr 100 V
Ground mfr 100 ohm
Interfaces, Connections & Control
3D CT Scan Region Display Function mfr Yes
Standards, Safety & General Data
CE, KC mfr Yes

Description

Construction and positioning

Integration & Operating Notes

Electrical
Main unit requires tap-switched 200–240 V supply

The main unit must be connected to a 200/220/230/240 V AC supply with the correct tap selected; incorrect tap setting or voltage outside ±10% can cause malfunction or damage. The supply must be rated for 2 kVA.

Factory Specification Sheet
Mounting
Floor must support 3100 kg shield box

The shielded cabinet weighs approximately 3100 kg and measures 2170 × 1350 × 1858 mm. Ensure the floor and access route can accommodate this load and size; inadequate support may affect alignment and safety.

Factory Specification Sheet
Electrical
Type-D ground with 100 Ω max resistance required

A dedicated Type-D ground with resistance not exceeding 100 Ω is mandatory for safe operation and to meet leakage dose limits. Poor grounding can compromise safety and image quality.

Factory Specification Sheet
Control & I/O
Exterior camera aids positioning but is optional

An exterior camera is available for positioning the sample; an optional retraction unit and operation box are offered. These are not included in the base delivery and must be specified separately.

Factory Specification Sheet

Selection & Integration Q&A

What are the power and grounding requirements for installing the inspeXio 7000 alongside a control computer?

The main unit needs AC 200–240 V ±10%, 50/60 Hz, 2 kVA with tap switching; the control computer needs AC 100–240 V ±10%, 50/60 Hz, 1.5 kVA. A Type-D ground with ≤100 Ω resistance is required.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
c251-e046
Datasheet ·5,7 МБ ·20 pp.
View PDF
PDF
c251-e047
Datasheet ·1004,5 КБ ·6 pp.
View PDF